753/754 Calibration Manual 16 Frequency Measurement To verify the frequency measurement function: 1. Connect the UUT as shown in Figure 4 . 2. Set the UUT to the frequency measurement function. 3. Select the <20 Hz ra…

Documenting Process Calibrator
Performance Verification Tests
15
5522A
CALIBRATOR
Fluke 75X
Fluke 5522A
2-WIRE COMP ON
gso05.eps
Figure 3. Resistance Measurement Verification Connections
Table 6. Resistance Measurement Verification Points
UUT
Range
Input
Minimum
1-Year
Maximum
1-Year
Minimum
2-Year
Maximum
2-Year
10.000 Ω 0 Ω -0.050 0.050 -0.070 0.070
10.000 Ω 10 Ω 9.945 10.055 9.923 10.077
100.00 Ω 0 Ω -0.05 0.05 -0.07 0.07
100.00 Ω 100 Ω 99.90 100.10 99.86 100.14
1000.0 Ω 0 Ω -0.5 0.5 -0.7 0.7
1000.0 Ω 1 kΩ 999.0 1001.0 998.6 1001.4
10.000 kΩ 0 Ω -0.010 0.010 -0.015 0.015
10.000 kΩ 10 kΩ 9.980 10.020 9.970 10.030

753/754
Calibration Manual
16
Frequency Measurement
To verify the frequency measurement function:
1. Connect the UUT as shown in Figure 4 .
2. Set the UUT to the frequency measurement function.
3. Select the <20 Hz range for the first step. Use the ≥20 Hz range thereafter.
4. Set the 5522A to the first test point in Table 7.
5. See if the frequency value shown on the UUT is in the range shown in the
applicable column.
6. Continue through the test points.
7. When you complete the test, set the 5522A to STANDBY.
5522A
CALIBRATOR
Fluke 75X
Fluke 5522A
gso06.eps
Figure 4. Frequency Measurement Verification Connections
Table 7. Frequency Measurement Verification Points
UUT Range Frequency Input V RMS
Minimum
1- & 2-Year
Maximum
1- & 2-Year
<20 Hz 10 Hz 300 mV 9.95 10.05
>20 Hz 150 Hz 300 mV 149.5 150.5
>20 Hz 1.2 kHz 1.0 V 1.195 1.205
>20 Hz 12 kHz 1.0 V 11.95 12.05
>20 Hz 49 kHz 2.0 V 48.95 49.05

Documenting Process Calibrator
Performance Verification Tests
17
DC Volts Source
To verify the dc volts source function:
1. Connect the UUT to the 8508A as shown in Figure 5.
2. Set the 8508A to measure dc volts.
3. Set the UUT to the dc volts source function at -10 mV. Let the UUT warm up for
a minimum of 5 minutes before you read the first indication.
4. See if the value shown on the 8508A is in the range shown in the applicable
column in Table 8.
5. Continue through the test points. See if the value shown on the UUT is in the
range shown in the applicable column.
6. When you complete the test, push on the UUT two times to turn the source
function off. This conserves battery life.
Fluke 75X
Fluke 8508A
gso07.eps
Figure 5. DC Volts Source Verification Connections
Table 8. DC Volts Source Verification Points
UUT
Range
UUT Output
Minimum
1-Year
Maximum
1-Year
Minimum
2-Year
Maximum
2-Year
100.000 10 mV 9.9940 10.0060 9.9935 10.0065
100.000 0.1 V 99.9850 100.0150 99.9800 100.0200
1.00000 V 0.15 V 0.14994 0.15007 0.14993 0.15007
1.00000 V 1 V 0.99985 1.00015 0.99980 1.00020
15.0000 V 1.5 V 1.49935 1.50065 1.49928 1.50073
15.0000 V 10 V 9.99850 10.00150 9.99800 10.00200