753/754 Calibration Manual 24 5522A CALIBRATOR Fl u ke 75X Room Temperat u re La b Bath Fluke 1551A Ex Thermometer Fluke 5522A TYPE-K Thermocouple Wire Cu Cu gso12.eps Figure 10. Temperature Me asure (TC) Verification Co…

100%1 / 52
Documenting Process Calibrator
Performance Verification Tests
23
Thermocouple Measure
To verify the thermocouple measure function.
1. Use Type-K thermocouple wire and copper wire to connect the 5522A output to
the UUT-TC jack as shown in Figure 10. The Type K-to-Copper junctions must
be welded or made with tight screw terminals and submersed in the lag bath
(room temperature). Use the standard thermometer (0.1 °C accuracy) to measure
the temperature of the lag bath.
2. Set the 5522A to source dc millivolts and the UUT to the thermocouple measure
function, TC Type K; ITS-90 scale, internal reference, and °C.
3. Stop for a minimum of 1 minute for thermal “emfs” (caused by insertion of the
connectors) to dissipate, and let the lag bath become stable for a minimum of
15 minutes.
4. Use the 5522A to source the millivolt equivalents of the temperatures in
Table 13. At each point, correct the 5522A output voltage. To do this, subtract
the millivolt equivalent of the temperature at the lag bath junction (use reference
chart below).
5. Continue through the test points. See if the value shown on the UUT is in the
range shown in the applicable column of Table 13.
6. When you complete the test, set the 5522A to STANDBY.
753/754
Calibration Manual
24
5522A
CALIBRATOR
Fluke 75X
Room Temperature
Lab Bath
Fluke 1551A Ex
Thermometer
Fluke 5522A
TYPE-K
Thermocouple
Wire
CuCu
gso12.eps
Figure 10. Temperature Measure (TC) Verification Connections
Documenting Process Calibrator
Performance Verification Tests
25
Table 13. Temperature Measure Verification
Input dcmV (referenced to 0 °C)
Minimum
1-Year °C
Maximum
1-Year °C
Minimum
2-Year °C
Maximum
2-Year °C
-5.550 mV (-180 °C) -180.9 -179.1 -181.2 -178.8
0.000 mV (0 °C) -0.5 0.5 -0.6 0.6
52.410 mV (1300 °C) 1299.1 1300.9 1298.8 1301.2
Lag Bath Reference Table, Type K, ITS-90
Tem
p
. °C 18 19 20 21 22 23 24 25 26 27 28
mV 0.718 0.758 0.798 0.838 0.879 0.919 0.960 1.000 1.041 1.081 1.122
Thermocouple Source
To verify the thermocouple source function:
Note
This test uses a Type-K thermocouple setting on the UUT.
1. Use Type-K thermocouple wire and copper wire to connect the 8508A to the
UUT -TC jack as shown in Figure 10 (the 8508A is used in place of the 5522A).
The type K-to-copper junctions must be welded or made with tight screw
terminals and submersed in the lag bath (room temperature). Use the standard
thermometer (0.1 % accuracy) to measure the temperature of the lag bath.
2. Set the UUT to the thermocouple source function, linear mode, TC type K, ITS-
90 scale, internal reference, and °C.
3. Set the 8508A to measure mV dc.
4. Stop for a minimum of 1 minute for thermal “emfs” (caused by insertion of the
connectors) to dissipate, and let the lag bath become stable for a minimum of 15
minutes.
5. Source each of the temperatures in Table 14 from the UUT. At each point,
correct the DMM measured voltage. To do this, add the millivolt equivalent of
the Type-K junction at the lag bath temperature (use the Type-K ITS-90 chart).
6. Continue through the test points. See if the value shown on the UUT is in the
range shown in the applicable column in Table 14.
7. When you complete the test, push on the UUT two times to turn the source
function off. This conserves battery life.