AppX.A_spec_mx3600_sys_10.11.pdf - 第10页
MX - 3600 / CXI3600M X / Specification confidential 10/ 39 spec_mx3600_sys_10.11.doc 2.2. MX - 3600 / X -RAY Imaging Tra y 2.2.1. X – RAY Trans mission Imaging The X2 Imaging Tray is based on X -RA Y transm ission t echn…

MX-3600 / CXI3600MX / Specification
confidential 9/39 spec_mx3600_sys_10.11.doc
2.1.2. Basic System Features / Specification Items
System Cabinet Conditions:
MX-3600 System Dimension: w=3.300 x d=1530 mm x h=1.400mm
w=130’’ x d=60’’ x h=55’’
Ground floor clearance: > 80mm
Transportation Height SMEMA: 950…1000mm
Number of feet: 4
MX-3600 weight: ca. 1.800 kg
Max. weight/foot-point: < 500kg/point
Lead shielding thickness: 3-5mm
Transportation Height SMEMA: 95…100mm
System Colour white/grey RAL 9002
Operator & Safety Labelling: English or German
Power, Environment & Interface Conditions:
Power Connection: 230V(208…240)/25 A (2 phase)
Max. Power Consumption: 4 kW
Main Power Switch 1
External Service Power Point: 4 x 110V
Air pressure: 6 bar / up to 140l/h max
Network Connection: 2 internal connection points
(Control PC)
1 external service connection points
Barcode Connection Point: RS232 Connection point
SMEMA Interface Standard for In and Out Conveyor
Safety, ESD & Lighting Operations
Light-Tower 3 lights in 3 colors
Optional: LED Setup Alarm Siren
Emergency Switches 3 (Input & Output & XRAY control
panel)
Noise < 68 dB in 1 m distance
ESD Ground Connection Point: 2 x Front Side

MX-3600 / CXI3600MX / Specification
confidential 10/39 spec_mx3600_sys_10.11.doc
2.2. MX-3600 / X-RAY Imaging Tray
2.2.1. X – RAY Transmission Imaging
The X2 Imaging Tray is based on X-RAY transmission technique therefore
providing a maximum on resolution in conjunction with variable field of view
performance for an optimized application setup.
Fig. 6: X-Ray Transmission Technology
2.2.2. X – Ray tube and control system
The MX-3600 X-Ray source is specifically selected for high performance
transmission mode characteristics under high volume production conditions.
Key parameters are in this respect:
X-Ray Flux and Power
X-Ray Power Stability
High Resolution
Fast Ramp-up
Max. life time.

MX-3600 / CXI3600MX / Specification
confidential 11/39 spec_mx3600_sys_10.11.doc
For the X-Ray tube supply – two alternative closed X-Ray tube types are
proven for the X2 / X- Ray source performance:
(Please note this part definition may change - depends on the up to date X-
Ray tube quality performances)
125kV / Micro-Focus Kevex
Fig. 7: Closed (Sealed) X-Ray tube principle
2.2.3. X-RAY tube features / Specification Items
Kevex 125kV
Tube Type
closed µfocus
Operating Voltage
40….130 kV
Target Current
up to 356µA
Max. Power
8W
Target Material:
Tungsten
Spot-Sizing:
Min.
Max.
10-15µ @ 4W
20µ @ 8W
Cooling:
Air-Cooling