WARNING
High voltage.— Do not remove any panels or
covers on the
FLX500
. There are no customer-ser-
viceable parts behind the covers. High voltage
may be present and exposed if covers are
removed. High voltage can cause injury or death.
2 … Periodic
Maintenance
Data I/O
FLX500 Chptr 2
—10—
NOTE: Once a year,
FLX500
Programmer
Module performance can be verified or
returned to Data I/O for verification. See
Annual Performance Verification on page
23.
Gantry Movement and
Alignment
The FLX500 gantry scans fiducials on the Tray
Modules and Programmer Modules using an
optical sensor on the PNP Head. This occurs
when programming is started after any module
receives power for the first time.
If you experience any alignment problems, such
as pick and place failures, you can:
rescan the fiducials at the Diagnostics –Mod-
ules Window
verify system calibrations at the Diagnostics
–Alignment Window. If the verification fails,
see Calibration Fails in the on-screen Help.
For more information, see the on-screen Help in
the diagnostics area of the application.
Daily Maintenance
Clean the Sockets
Tools Required:
Shop air, clean and dry, .48-.55 MPa
(70-80 PSI).
4 mm Hex Key (Allen Wrench)
Remove dust and debris from the sockets with
clean, dry, compressed air. Actuate the sockets to
ensure dust and debris are cleared from beneath
the contacts. If the Socket Adapter is installed in
the FLX500 programmer, the sockets can be actu-
2 … Periodic Maintenance Weekly Maintenance
—11—
ated by pushing the Socket Actuator button on
the monitor.
CAUTION: Pinch Point. The Actu-
ator Plate lowers onto the sockets to open
them. Keep fingers and objects away from the
Socket Adapters when actuating them.
Check the sockets for wear; replace the Socket
Adapter if necessary.
If further cleaning is needed for best throughput,
remove the Socket Adapter and clean it in an
ultrasonic bath with approved cleaning fluid.
Isopropyl alcohol (isopropanol) can be used to
clean the Socket Adapter.
Weekly Maintenance
Running the Self-test
Checking Probe Tips
Cleaning the Fiducials and Sensors
Running the Self-test
Tools Required:
•Vacuum tweezers
Perform the system Self-test once each week to
verify proper operation of all internal electronic
systems.
NOTE: The Self-test runs automatically
when power is applied to the
FLX500
unit.
To run the Self-test:
1. End a job if one is running.