BA-UI-EN-D33363.pdf - 第27页
D33363 | Date 01/2020 27 Operation and control TIMER End 14 : 45 44 h :44 m 180 .4 °C TEMP Set 180 .4 °C 20 % FLAP FAN % 0 FLAP 40 % 5 3 4 12 TIMER 30 m 04 h End 13:30 23.11. 11 12 67 8 9 1 0 Fig. 15 ControlCOCKPIT for U…

26 D33363 | Date 01/2020
Operation and control
5.3 Loading the appliance
Warning!
When loading the appliance with an unsuitable load, poisonous
or explosive vapours or gases may be produced. This could cause
the appliance to explode, and persons could be severely injured or
poisoned. The appliance may only be loaded with materials which do
not form any toxic or explosive vapours when heated up and cannot
ignite (see also page 11). If there is any doubt as to the composi-
tion of materials, they must not be loaded into the appliance.
Caution:
Check the chamber load for chemical compatibility with the materials of the appliance
(see page 12).
Caution:
If the chamber load is wet or very humid, water may accumulate on the floor and dam-
age the heater. If wet, use a drip tray on the bottom of the tube.
Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the
load capacity are specified in the technical data overview on page 14.
The chamber must not be
loaded too tightly, so that
proper air circulation in the
working chamber is guaran-
teed. Do not place any
chamber load on the floor,
touching the side walls or
right below the ceiling of the
working chamber (Fig. 14 , see
also the "correct loading"
sticker on the appliance).
In case of improper loading
(chamber loaded too tightly),
the set temperature may
be exceeded or it may take
longer until it is reached.
To achieve the correct heating capacity, the type of slide-in unit used – Grid or Shelf –
must be set in the menu under
SETUP (see page 41).
5.4 Operating the appliance
5.4.1 ControlCOCKPIT
In manual operation, the desired parameters are entered at the ControlCOCKPIT on the front
of the appliance (Fig. 15 and Fig. 16 ). You can also make basic settings here (menu). Addition-
ally, warning messages are displayed, e.g. if the temperature is exceeded.
Fig. 14 Correct placement of the chamber load

D33363 | Date 01/2020 27
Operation and control
TIMER
End 14 : 45
44h:44m
180.4 °C
TEMP
Set 180
.4 °C
20%
FLAP
FAN
%
0
FLAP
40 %
53 412
TIMER
30m04h
End
13:30 23.11.
11 1267 8 9 10
Fig. 15 ControlCOCKPIT for UF.../IF... appliances in operating mode
180.0 °C
Set 180.9 °C
TEMP
TIMER
End 14 : 45
44h:44m
FLAP
FLAP
40 %
12 5
TIMER
30m04h
End
13:30 23.11.
10 11 1267 8 9
Fig. 16 ControlCOCKPIT for UN.../IN... appliances in operating mode
1 Activation key for temperature setpoint
adjustment
2 Setpoint and actual temperature display
3 Fan speed display
4 Activation key for fan speed setting
5 Switch to menu mode (see page 36)
6 Activation key digital backwards counter
with target time setting, adjustable from 1
minute to 99 days
7 On/Off switch
8 Display digital backwards counter with
target time setting, adjustable from 1 min-
ute to 99 days
9 Air flap position display
10 Activation key for air flap position adjust-
ment
11 Turn control for setpoint adjustment
12 Confirmation key (accepts setting made
with the turn control)

28 D33363 | Date 01/2020
Operation and control
5.4.2 Basic operation
In general, all settings are made according to the following pattern:
1. Activate the desired parameter (e.g. tem-
perature). To do so, press the correspond-
ing activation key on the left or right of
the respective display. The activated dis-
play is lined in colour, the other displays
are dimmed. The set value is highlighted
in colour.
. 5°C100
TEMP
TIMER
TIMER
-
-
-
-
m
h
22.4 °C
TEMP
37.0 °CSet
2. By turning the turn control to the left
or right, adjust the set value (e.g. to
180.0 ºC).
3. Save the set value by pressing the confir-
mation key.
The display returns to normal and the
appliance begins adjusting to the defined
set value.
23.2 °C
TEMP
Set 180
.0 °C
Additional parameters (air flap position etc.) can be set accordingly.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati-
cally returns to the main menu and restores the former values.
If you want to cancel the setting procedure, press the
activation key on the left or right of the display that you
want to exit. The appliance restores the former values.
Only the settings that you have confirmed by pressing
the confirmation key before cancelling the setting
procedure are accepted.
5.4.3 Operating modes
The appliance can be operated in different modes:
► Manual mode: The appliance runs in permanent operation at the values set on the
ControlCOCKPIT. Operation in this mode is described in chapter 5.4.4.
► Operation with digital backwards counter with target time setting, adjustable from 1 min-
ute to 99 days (Timer): The appliance runs at the values set until the timer has elapsed.
Operation in this mode is described in chapter 5.4.5 .
► via remote control (see page 41)
22.4 °C
TEMP
180.0 °CSet
T
T
T
-
T