MDO3000 Programmer Manual.pdf - 第236页

Commands Listed in A lphabetical Order Syntax BUS:LOWerthre shold:REF<x> {<NR3>|ECL|T TL} BUS:LOWerthre shold:REF<x>? Related Commands BUS:UPPerthreshold:REF<x> Arguments <NR3> is a floating …

100%1 / 1158
Commands Listed in Alphabetical Order
Arguments
<NR3> isafloati
ng point number that specifies the threshold, in volts.
ECL specifies a preset ECL high level of -1.3V.
TTL specifies a preset TTL high level of 1.4V.
Examples
BUS:LOWERTHRESHOLD:CH1 TTL sets the CH1 lower threshold to 800mV.
BUS:LOWERTHRESHOLD:CH1? might return :BUS:LOWERTHRESHOLD:CH1
-800.0000E-3
indicatingtheCH1lowerthresholdis-800mV.
BUS:LOWerthreshold{:MATH|:MATH1}
This command specifies the lower threshold for the math waveform. This will
apply to all search and trigger types that use the math waveform.
Group
Bus
Syntax
BUS:LOWerthreshold{:MATH|:MATH1} {<NR3>|ECL|TTL}
BUS:LOWerthreshold{:MATH|:MA
TH1}?
Related Commands
BUS:UPPerthreshold{:MATH|:MATH1}
Arguments
<NR3> is a floating point number that specifies the lower threshold for the
reference waveform, in volts.
ECL ECL (-1.3 volts). Note that this setting is constrained, depending upon the
vertical scale of the specified math waveform.
TTL TTL (1.4 volts). Note that this setting is constrained, depending upon the
vertical scale of the specified math waveform.
Examples
BUS:LOWerthreshold:MATH TTL sets the lower threshold for the math
waveform to 1.4 volts.
BUS:LOWerthreshold:MATH? might return 0.0E+0, which is the default.
BUS:LOWerthreshold:REF<x>
This command sets the lower threshold for each reference waveform. This applies
to all search and trigger types that use that reference waveform.
Group
Bus
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-205
Commands Listed in Alphabetical Order
Syntax
BUS:LOWerthre
shold:REF<x> {<NR3>|ECL|T TL}
BUS:LOWerthreshold:REF<x>?
Related Commands
BUS:UPPerthreshold:REF<x>
Arguments
<NR3> is a floating point number that specifies the lower threshold for the
reference waveform, in volts
ECL ECL (-1.3 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified reference waveform.
TTL TTL (1.4 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified reference waveform.
Examples
BUS:LOWerthreshold:REF2 TTL sets the lower threshold for reference
waveform 2 to 1.4 volts. Note that this setting is constrained, depending upon the
vertical scale for the specified reference waveform.
BUS:LOWerthreshold:REF3? might return 0.0E+0, which is the default.
BUS:THReshold:CH<x>
This command specifies the threshold for analog channel <x>, where x is the
channel number (1–4). This setting applies to all trigger types that use the channel.
Conditions
This command requires a serial bus analysis module on DPO models.
Group
Bus
Syntax
BUS:THReshold:CH<x> {ECL|TTL|<NR3>}
BUS:THReshold:CH<x>?
Arguments
ECL specifies a preset ECL high level of –1.3V.
TTL specifies a TTL preset high level of 1.4V.
<NR3> is a floating point number that specifies the threshold level, in volts.
BUS:THReshold:D<x>
This command specifies the threshold for digital channel <x>, where x is the
digital channel number (0–15). This will apply to all Search and Trigger Types
that use the channel.
2-206 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
NOTE. MDO3000 m
odels only have two digital threshold controls: one for digital
channels D0 - D7, and the other for D8 - D15. Changing one digital channel’s
threshold l evel changes all of the channels’ levels in the same set.
Group
Bus
Syntax
BUS:THReshold:D<x> {<NR3>|ECL|TTL}
BUS:THReshold:D<x>?
Arguments
ECL specifies a preset ECL high level of -1.3V.
TTL specifies a preset TTL high level of 1.4V.
<NR3> is a floating point number that specifies the threshold level, in volts.
BUS:UPPerthreshold:CH<x>
Sets the upper threshold for each analog channel (1–4). This applies to all search
and trigger types that use the channel.
Group
Bus
Syntax
BUS:UPPerthreshold:CH<x> {<NR3>|ECL|TTL}
BUS:UPPerthreshold:CH<x>?
Arguments
<NR3> is a floating point number that specifies the threshold, in volts.
ECL specifies a preset ECL high level of -1.3V.
TTL specifies a preset TTL high level of 1.4V.
Examples
BUS:UPPERTHRESHOLD:CH1 800.0000E-3 sets the CH1 upper threshold to
800 mV.
BUS:UPPERTHRESHOLD:CH1? might return :BUS:UPPERTHRESHOLD:CH1
-800.0000E-3
indicating that the CH1 upper threshold is set to -800 mV.
BUS:UPPerthreshold{:MATH|:MATH1}
This command specifies the upper threshold of the math waveform. This will
apply to all sea
rch and trigger types that use the math waveform.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-207