MDO3000 Programmer Manual.pdf - 第652页
Commands Listed in A lphabetical Order SEARCH:SEARC H<x>:TRIGger:A:BUS:B<x>:AUDio:DA T a:QUALi fi er This command sets the qualifier (<, >, =, <=, >=, not =, in range, out of range) to be used to se…

Commands Listed in Alphabetical Order
search on the lo
wer word value.) The search condition must be set to DATA using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:CONDition.
SEARCH<x> is t
he search number, which is always 1, and
B<x>
Conditions
This command requires a DPO4AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:HIVALue
<String>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:HIVALue?
Arguments
<String> specifies the upper word value.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:OFFSet
This commands sets the data offset value to be used to search on
audio bus data. The search condition must be set to DATA using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:CONDition.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:OFFSet <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:OFFSet?
Arguments
<NR1> is the data offset value.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-621

Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifier
This command sets the qualifier (<, >, =, <=, >=, not =, in range, out of range) to
be used to search on audio bus data. The search condition must be set to DATA
using SEARCH
:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:CONDition.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4
AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSEQual|MOREEQual
|INrange|OUTrange}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifier?
Arguments
LESSthan sets the qualifier to less than.
MOREthan sets the quali
fier to greater than.
EQual sets the qualifier to equal.
UNEQual sets the qualifier to not equal.
LESSEQual sets the qualifier to less than or equal.
MOREEQual sets the qualifier to greater than or equal.
INrange sets the qualifier to in range.
OUTrange sets the qualifier to out of range.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:VALue
This command sets the lower word value to be used to search on audio bus data.
(Use SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:HIVALue
to set the upper word value.) The search condition must be set to DATA using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:CONDition.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
2-622 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:VALue
<String>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:VALue?
Argum
ents
<String> is the lower word value.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:WORD
This command sets the alignment of the data (left, right or either) to be used
to search on audio bus data. The search condition must be set to DATA using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:CONDition.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:WORD
{EITher|LEFt|RIGht}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:AUDio:DATa:WORD?
Arguments
EITher aligns the data to either left or right.
LEFt aligns the data to the left.
RIGht aligns the data to the right.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
This command sets the condition (start of frame, frame type, identifier, matching
data, end of frame, missing ACK field, bit-stuffing error, form error, any error,
CAN FD BRS bit, or CAN FD ESI bit) to be used to search on CAN bus data.
SEARCH<x> is the search number, which is always 1, and B<x> is the bus number
(1-4). 1-3 for the MDO4000C and 1-2 for the MDO3000.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-623