MDO3000 Programmer Manual.pdf - 第81页
Command Groups When the two man ual markers are turned on, the Reference Marker is no longer automatically attached to the highe st amplitude peak. It can now be moved t oa n y desired location. This enables easy measure…

Command Groups
Spectrogram sl
ices are generated by taking each spectrum and flipping it on its
edge, so that it is one pixel row tall. Each new acquisition adds another slice at the
bottom of the spectrogram, and the previous acquisitions (slices) move up one
row; you can then navigate backwards through the history of the spectrogramby
selecting slice numbers to view. (The spectrogram slice trace is displayedas
the RF Normal trace.)
Spectrum Mo
de: Triggered
and Free Run
When only the frequency domain waveforms are displayed (no time domain
waveforms), you can choose whether the MDO4000/B/C should use Triggered
mode or Free Run mode. (MDO3000 only uses Free Run mode.) When Triggered
mode is se
lected, you can control all trigger settings, including Normal and
Auto triggering. When Free Run mode is selected, the oscilloscope generates
RF acquisitions as fast as possible. (To display only the frequency domain
waveforms, turn off all time domain waveforms, including channels 1- 4, digital
channels 0 - 15, buses, time domain math waveforms, time domain reference
waveforms, and any RF vs. Time traces.)
When the oscilloscope displays both time and frequency domain waveforms,
then the instrument’s trigger system is in control of the Triggered mode andthe
RF acq
uisitions.
Using Markers in the
Frequency Domain for
Measurement and Analysis
For frequency domain measurements, up to 11 automatic markers are available
to assist with quickly identifying the frequency and amplitude of peaks in the
spe
ctrum based upon user threshold and excursion settings. If more peaks meet
the criteria than the desired number of markers, then the highest amplitudepeaks
are shown. Two manual markers are also available for measuring non-peak areas
of interest, and to measure Noise Density and Phase Noise. If manual markersare
off, the reference marker is automatically placed on the highest amplitudepeak.
With manual markers on, the reference marker becomes the “A” manual marker.
A
utomatic peak markers are on by default.
Each automatic marker has a readout associated with it. These can be absolute
o
r delta readouts. An absolute marker readout shows the actual frequency and
amplitude of the associated marker. A delta marker readout shows the frequency
and amplitude of the automatic markers relative to the Reference Marker.
The Reference Marker’s readout indicates absolute frequency and amplitude,
regardless of the readout type. (It is marked on the display with a red R in a
triangle.) The marker measurement readouts are absolute in dBm or relativetothe
reference marker in dBc (dB below carrier amplitude).
The threshold and excursion settings define which peaks are marked automatically.
The threshold is a minimum amplitude that a signal must cross to be a valid peak.
If the threshold is lower, more peaks will tend to qualify for markers. If the
threshold is higher, fewer peaks tend to qualify for markers. The excursion is how
far a signal needs to fall in amplitude between marked peaks to be another valid
peak. If the excursion is low, more peaks will tend to qualify for markers. Ifthe
excursion is high, fewer peaks will tend to qualify for markers.
2-50 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Command Groups
When the two man
ual markers are turned on, the Reference Marker is no longer
automatically attached to the highest amplitude peak. It can now be moved toany
desired location. This enables easy measurement of any part of the spectrum,
as well as delta measurements to any part of the spectrum. This also lets you
measure non-peak spectral content of interest. The readouts for manual markers
indicate frequency, amplitude and noise (just like automatic marker readouts).
Taking Auto
matic
Measurements in the
Frequency Domain
You can take three automatic measurements in the frequency domain:
1. Channel Power (CP) — The total power within the bandwidth, defined by
the Channel Width.
2. Adjacent Channel Power Ratio (ACPR) — The power in the main channel
and the ratio of channel power to main power, for the upper and lower halves of
each adjacent channel.
3. Occupied Bandwidth (OBW) — The bandwidth that contains the specified
percentage of power within the analysis bandwidth.
Triggering using RF
Power L evel as a Source;
Searching
The ability to use the RF power level as the source for edge triggering and
searching is provided standard with the MDO4000/B and MDO4000C with
option SA3 or SA6 series models. With the MDO4TRIG (Advanced Trigger)
application module installed, the RF power level can also be used as the source
for p
ulse width, time-out, runt, logic, and sequence triggering. (See page 2-76,
Trigger Command Group.) A number of search commands are also available.
(See page 2-61, Search Command Group.)
Transfe rring and Saving
RF Trace Information
Yo
u can perform waveform transfer commands and queries using RF traces. (See
page 2-95, Waveform Transfer Command Group.)
RF
traces can be saved to an .ISF or .CSV file for subsequent recall to any of the
4 internal reference memory locations. The oscilloscope can also save, butnot
recall, RF acquisitions as .TIQ files. You can import .TIQ files into Tektronix
SignalVu-PC software (PC based), SignalVu software (oscilloscope based),
RSAVu software (PC based) or into a Tektronix real-time spectrum analyzer for
pulse analysis and demodulation analysis. (See page 2-58, Save and Recall
Command Group.)
NOTE. The RF input replaces the Aux Input connector on the front panel,
therefore aux-in commands and arguments are not supported on the MDO4000/B,
MDO4000C with option SA3 or SA6, and MDO3000 models with 4-channels.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-51

Command Groups
Table 2-37: RF C
ommands
Item Description
MARKER:M<x>:AMPLitude:
ABSolute?
This query ret
urns the actual amplitude (vertical) value of the either of the two manual markers
that are available for frequency domain traces, in dBm.
MARKER:M<x>
:AMPLitude:
DELTa?
This query returns the delta amplitude (vertica l) value of either of the two manual markers that
are available for frequency domain traces, in relation to the Reference Marker.
MARKER:M<x>:FREQuency:
ABSolute
This command specifies the actual frequency (horizontal) value of either of the two manual
markers tha
t a re available for frequency domain traces.
MARKER:M<
x>:FREQuency:
DELTa?
This query
returns the delta frequency (horizontal) value of either of the two manual markers
that are available for frequency domain traces, in relation to the Reference Marker.
MARKER:M<x>:NOISEDensity?, This c ommand returns the noise density of the RF_NORMal trace at the specified marker
position in <RF Units>/Hz units, where <RF Units> are the units specified by the command
RF:UNIts
.
MARKER:
M<x>:PHASENoise?
This com
mand returns the phase noise of the RF_NORMal trace at the specified marker
position in dBc/Hz units.
MARKER:
MANual
This command switches on or off the manual markers a and b that are available for frequency
domain traces.
MARKER:PEAK:EXCURsion This command specifies the excursion value, in user-selected units, for the frequency domain
trace a
utomatic peak markers. You can select the units with the command RF:UNIts.
MARKER:PEAK:MAXimum
This c
ommand specifies the maximum number of frequency domain trace peaks that should
have automatic markers associated with them. This can be a number between 1 and 11.
MARKER:PEAK:STATE This command switches on or off the automatic peak markers that are available for frequency
domain traces. The default is 1 (on). There are up to 11 automatic markers.
MARKER:PEAK:THReshold
This command specifies the threshold value, in the same vertical units as the source waveform,
of t
he automatic peak markers available f or frequency domain traces. (Use the RF:UNIts
to specify the units.)
MAR
KER:REFERence
This command changes the Center Frequency to the frequency indicated by the Reference
Marker, in effect moving the Reference Marker to the center of the screen.
MARKER:REFERence:
AMPlitude?
This query returns the amplitude (vertical) value of the Reference Marker in dBm when markers
ar
e turned on (using the command MARKER:PEAK :STATE or MARKER:MANual).
MARKER:REFERence:
F
REQuency?
T
his query returns the frequency of the Reference Marker when the frequency domain
trace markers have been turned on (using either the command MARKER:P EAK:STATE or
MARKER:MANual).
MARKER:TYPe
This command specifies the marker type (either DELTa or ABS olute) to use when the automatic
markers for the frequency domain traces are turned on. To turn on the automatic markers,
use the command MARKER:PE AK:STATE.
RF:CLIPPing? Returns a boolean indicating whether the RF input is “clipping” due to input over/under-range
detection.
RF:DETECTionmethod:MODe This command specifies whether the R F detection within the oscilloscope occurs automatically
or manually. T he default is AUTO. If you set the detection method mode to MANual, you
are then able to use the related
RF:DETECTionmethod commands to specify detection
method options for the frequency dom ain traces (the options are
MINUSpeak, SAMple,
PLUSpeak and AVErage).
2-52 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual