MDO3000 Programmer Manual.pdf - 第419页

Commands Listed in Alphabetical Order Arguments ON turns on this feature, so that the instrument will gene rate a scre en hard copy to the default printer upon failure. OFF turns off t his feature. <NR1> = 0 turns …

100%1 / 1158
Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:DELay <NR3>
MASK:TESt:DELay?
Related Commands
MASK:TESt:STATE, MASK:TESt:TIME
Arguments
<NR3> is a floating point number that specifies the amount of time in seconds, to
delay the start of the mask pass/fail test.
Examples
MASK:TEST:DELAY 2.5 delays waveform evaluation by 2.5 seconds after the
start of pass/fail testing.
MASK:TEST:DELAY? might return 5, indicating that the instrument has been set
to wait 5 seconds after the start of a mask test before it evaluates the waveforms.
MASK:TESt:HARDCopy
This command will cause the ins
trument to generate a screen shot to the default
printer as soon as a pass/fail mask test fails, using the current instrumenthard
copy settings. See HARDCopy:ACTIVeprinter and other Hardcopy commands
for more information on accessing printer settings.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:HARDCopy {ON|OFF|<NR1>}
MASK:TESt:HARDCopy?
Related Commands
MASK:TESt:STATE,
HARDCopy:ACTIVeprinter
2-388 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Arguments
ON turns on this
feature, so that the instrument will generate a screen hard copy to
the default printer upon failure.
OFF turns off t
his feature.
<NR1> = 0 turns off this feature; any other value turns it on.
Examples
MASK:TESt:HARDCopy ON generates a screen shot to the default printer upon
test failure.
MASK:TESt:HARDCopy? might return 0, indicating that the instrument is not set
to generate a screen shot upon failure.
MASK:TESt:REPeat
This command causes the mask test cycle to continuously repeat upon the
completion of the previous test cycle.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:REPeat {ON|OFF|<NR1>}
MASK:TESt:REPeat?
Related Commands
MASK:TESt:STATE,
Arguments
ON turns on this feature, so that the mask test cycle repeats continuously uponthe
completion of the previous test cycle.
OFF turns off this feature.
<NR1> is an integer. 0 turns off this featu
re; any other integer turns it on.
Examples
MASK:TESt:REPeat ON
causes the mask test cycle to continuously repeat.
MASK:TESt:REPeat? might return 0, indicating that the mask test cycle is not
set to continuously
repeat.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-389
Commands Listed in Alphabetical Order
MASK:TESt:SA
VEIMAGE
This command causes the instrument to copy the screen image to a file on a USB
mass storage device or a mounted network drive if a pass/fail mask test fails. See
the command F
ILESystem:MKDir and other File System commands for more
information on saving to a file.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:SAVEIMAGE {ON|OFF|<NR1>}
MASK:TESt:SAVEIMAGE?
Related Commands
MASK:TESt:STATE,
FILESystem:MKDir
Arguments
ON turns on this feature, so that the screen image is copied to a speci
fied file
upon test failure.
OFF turns off this feature.
<NR1> is an integer. 0 turns off this feature; any other integer turns it on.
Examples
MASK:TESt:SAVEIMAGE On copies a screen image to a file upon test failure.
MASK:TESt:SAVEIMAGE? might return 0, indicating that the instrument is not set
to copy a screen image to a file upon failure.
MASK:TESt:SAVEWFM
This command copies the waveform data from all active channels to a file on a
flash memory device such as a USB stick or a network drive. See the command
FILESystem:MKDir and other File System commands for more information on
saving to a file.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
2-390 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual