MDO3000 Programmer Manual.pdf - 第546页
Commands Listed in A lphabetical Order Group Power Syntax POWer:SOA:RES ult:FAILures:QTY ? Examples POWER:SOA:RES ULT:FAILURES:QTY ? might return :POWER:SOA:RE SULT:FAILURES 0 indicating that the number of failures is 0.…

Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:STOPOnviol {OFF|ON|0|1}
POWer:SOA:MASK:STOPOnviol?
Arguments
OFF or 0 enables mask stop on violations.
ON or 1 disables mask stop on violations.
Examples
POWER:SOA:MASK:STOPONVIOL 1 disables the state of the mask stop on
violation condition.
POWER:SOA:MASK:STOPONVIOL ? might return
:POWER:SOA:MASK:STOPONVIOL 0 indicating that the mask stop is enabled.
POWer:SOA:PLOTTYPe
This command specifies the Safe Operating Area (SOA) plot type.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:PLOTTYPe {LOG|LINear}
POWer:SOA:PLOTTYPe?
Arguments
LOG for logarithmic SOA plot type.
LINear for linear SOA plot type.
Examples
POWER:SOA:PLOTTYPE LINear sets the SOA plot type to Linear.
POWER:SOA:PLOTTYPE ? might return :POWER:SOA :PLOTTYPE LO G indicating
that the SOA plot type is set to Log.
POWer:SOA:RES
ult:FAILures:QTY? (Query Only)
Returns the number of failures in the test.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-515

Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:SOA:RESult:FAILures:QTY?
Examples
POWER:SOA:RESULT:FAILURES:QTY? might return
:POWER:SOA:RESULT:FAILURES 0 indicating that the number of failures is 0.
POWer:SOA:RESult:NUMACq? (Query Only)
Returns the number of acquisitions in the test.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:RESult:NUMACq?
Examples
POWER:S
OA:RESULT:NUMACQ?
might return :POWER:SOA:RESULT:NUMACQ
10
indicating that the number of acquisitions is 10.
POWer:SOA:RESult:STATE? (Query Only)
Returns the pass/fail state of the SOA test.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:RESult:STATE?
Returns
PASS or FAIL.
Examples
POWER:SOA:RESULT:STATE? might return :POWER:SOA:MASK:STATE PASS
indicating that the SOA test state is PASS.
2-516 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
POWer:STATIs
tics (No Query Form)
Clears all the accumulated statistics of all measurements. Performs the same
function as the MEASUrement:STATIstics command.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:STATIstics {RESET}
Arguments
RESET clears the measurement statistics.
POWer:STATIstics:MODe
Enables or disables the display of the measurement statistics. Performs the same
function as the MEASUrement:STATIstics:MODe command.
Conditions
This command requires a DPO4PWR application modul
e for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:STATIstics:MODe {OFF|ALL}
POWer:STATIstics:MODe?
Arguments
ALL turns on measurement statistics display.
OFF turns all measurements statistics off.
Examples
POWER:STATISTICS:MODE OFF turns measurements statistics display off.
POW
ER:STATISTICS:MODE?
might return :MEASUREMENT:STATISTICS:MODE
ALL
indicating that measurement statistics are turned on and all statistics are
being displayed for each measurement.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-517