Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:MIL1553B:STATus:BIT:TF
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word terminal flag bit value (bit 19) to be used in the trigger.
Returned val
ues are 0, 1, or X (don’t care, which is the default).
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRI
Gger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TR
IGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:BIT:TF NOCARE sets the status word
terminal flag bit value to be used in the trigger to X.
TRIG:A:BUS:B1:MIL1553B:STAT:BIT:TF? might return 1.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status parity bit value to be used in the trigger. Returned values are 0,
1, or X (don’t care, which is the default).
B<x>
2-850 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:PAR ONE sets the parity bit value to be
used in the trigger to 1.
TRIG:A:BUS:B1:MIL1553B:STATUS:PARity? might return X, indicating
that the parity doesn’t matter.
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies either the minimum remote terminal response time (RT) limit
for the amount of time the terminal has to transmit, or it specifies the
minimum inter-message gap (IMG). (You can specify RT or IMG using the
TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe command.)
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-851
Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit <NR3>
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
Arguments
<NR3> is a floating point number that specifies either the minimum remote
terminal response time (RT) or the inter-message gap (IMG) in seconds.
Examples
TRIGger:A:BUS:B1:MIL1553B:TIMe:LESSLimit 3.0000e-6 would set
either the remote terminal response time (RT) or the inter-message gap (IMG) to
3.00 μS.
TRIGger:A:BUS:B1:MIL1553B:TIMe:LESSLimit? might return 2.0000e-6,
indicating that the RT or IMG has been set to 2.00 μS.
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies either the maximum remote terminal response time (RT) limit for
the amount of time the terminal has to transmit, or it specifies the maximum
inter-message gap (IMG). (You can specify the RT and IMG using the
TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe command.)
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit <NR3>
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
2-852 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual