Commands Listed in A lphabetical Order SEARCH:SEARC H<x>:TRIGger:A:BUS:B<x>:CAN:DA T a:V ALue This command sets the binary data value to be used to search on CAN bus data. This only applies if the search cond…

Commands Listed in Alphabetical Order
Syntax
SEARCH:SEARCH
<x>:TRIGger:A:BUS:B<x>:CAN:DATa:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSEQual|EQual}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:QUALifier?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue
Arguments
LESSthan searches for bus data less than the value specified by
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
MOREthan searches for bus data greater than the value specified by
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
EQual searches for bus data equal to the value specified by SEARCH:
SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
UNEQual searches for bus data not equal to the value specified by
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
LESSEQual searches for bus data less equal to the value specified by
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
EQual searches for bus data equal to the value specified by SEARCH:
SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:SIZe
This command sets the length of the data string, in bytes, to be used to search
on CAN bus data. This only applies if the search condition has been set to
IDANDDATA or DATA (using SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:
CAN:CONDition).
SEARCH<x> is the search number, which is always 1, and B<x> is the bus number
(1-4). 1-3 for the MDO4000C and 1-2 for the MDO3000.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX applic
ation module
on DPO/MSO4000B and MDO4000/B/C models and a MDO3AUTO application
module on MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:SIZe <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:SIZe?
Arguments
<NR1> is the data string length in bytes.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-627

Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue
This command sets the binary data value to be used to search on CAN bus data.
This only applies if the search condition has been set to IDANDDATA or DATA
(using SEARC
H:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition).
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application module
on DPO/MSO4000B and MDO4000/B/C models and a MDO3AUTO application
module on MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue <bin>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DAT
a:QUALifier
Arguments
<bin> is the data in binary format.
SEARCH:SEARCH<x>
:TRIGger:A:BUS:B<x>:CAN:FD:BRSBIT
This command specifies the binary data value to be used to search on CAN FD
BRS bits. This only applies if the search condition has been set to FDBRSBIT
(using
SEARCH:SE
ARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
).
SEARCH:<x> is the search number, which is always 1, and B<x> is the bus
number (1-3), ex
cept for the MDO3000 which only has two buses.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application module
on MDO4000C models and a MDO3AUTO application module on MDO3000
models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:
BRSBIT{ZERo|OFF|0|ONE|ON|1|NOCARE|X|x}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:BRSBIT?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
2-628 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
ZERo, OFF,and0
specify that the BRS bit must have value 0. If queried, the
command will always return 0 if set with these arguments.
ONE, ON,and1 s
pecify that the BRS bit must have value 1. If queried, the
command will always return 1 if set with these arguments.
NOCARE, X,a
nd
x specify that the BRS bit may be either 1 or 0. If queried, the
command will always return X if set with these arguments.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:ESIBIT
This command specifies the binary data value to be used to search on CAN FD
ESI bits. This only applies if the search condition has been set to FDESIBIT
(using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition).
SEARCH:<x> is the search number, which is always 1, and B<x> is the bus
number (1-3), except for the MDO3000 which only has two buses.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application module
on MDO4000C models and a MDO3AUTO application module on MDO3000
models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:
ESIBIT{ZERo|OFF|0|ONE|ON|1|NOCARE|X|x}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:ESIBIT?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
Arguments
ZERo, OFF,and0 specify that the BRS bit must have value 0. If queried, the
command will always return 0 if set with these arguments.
ONE, ON,and1 specify that the BRS bit must have value 1. If queried, the
command will always return 1 if set with these arguments.
NOCARE, X,andx specify that the BRS bit may be either 1 or 0. If queried, the
command will always return X if set with these arguments.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FRAMEtype
This command sets the frame type (data, remote, e
rror or overload) to be used
to search on CAN bus data. This only applies if the search condition has been
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-629