Commands Listed in A lphabetical Order Examples SEARCH:SEARCH 1:TRIGger:A:LOGi c:INPut:RF_AMPl itude HIGH specifies to set a ma rk when the wavefor m goes abov e the logic threshold spec ified by SEARCH:SEARCH<x>:TRI…

Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:LOGIc:INPut:REF<x>
This command specifies the logic condition of the reference waveform to be
used in a logic search.
SEARCH<x> is the search number, which is always 1, and
REF<x> is the
reference channel number, which can be 1–4.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:REF<x> {HIGH|LOW|X}
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:REF<x>?
Arguments
HIGH specifies a high logic level.
LOW specifies a low logic level.
X specifies a “don't care” condition.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_AMPlitude
This command specifies the logic criteria to search for when searching using
the RF Amplitude vs. Time trace. This command is used together with
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_AMPlitude to
search for an instance when the waveform goes higher or lower than the specified
threshold. “LOW” specifies to only set a mark when the RF_AMPlitude
waveform drops below the logic threshold; likewise, “HIGH
” specifies to set a
mark when the waveform goes above the logic threshold. “X” specifies “don’t
care”.
SEARCH<x> is the search number, which is always 1.
Conditions
MDO4000/B/C models only.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_AMPlitude
{HIGH|LOW|X}
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_AMPlitude?
Related Commands
TRIGger:A:LOGIc:INPut:RF
Arguments
HIGH specifies a high logic level.
LOW specifies a low logic level.
X specifies a “don't care” condition.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-723

Commands Listed in Alphabetical Order
Examples
SEARCH:SEARCH
1:TRIGger:A:LOGic:INPut:RF_AMPlitude HIGH
specifies to set a mark when the waveform goes above the logic threshold specified
by SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_AMPlitude.
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:RF_AMPlitude? might
return HIGH, indicating that the search engine has been set to set a
mark when a waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_AMPlitude.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_FREQuency
This comm
and specifies the logic criteria to search for when searching using
the RF Frequency vs. Time trace. This command is used together with
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency to
search for an instance when the waveform goes higher or lower than the specified
threshold. “LOW” specifies to only set a mark when the RF_FREQuency
waveform drops below the logic threshold; likewise, “HIGH” specifies to seta
mark wh
en the waveform goes above the logic threshold. “X” specifies “don’t
care”.
SEARCH<x> is the search number, which is always 1.
Conditions
MDO4000/B/C models only.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_FREQuency
{HIGH|LOW|X}
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_FREQuency?
Related Commands
TRIGger:A:LOGIc:INPut:RF
Arguments
HIGH specifies a high logic level.
LOW specifies a low logic level.
X specifies a “don't care” condition.
Examples
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:RF_FREQuency HIGH
specifies to set a mark when the waveform goes above the logic threshold specified
by SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency.
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:RF_FREQuency? might
return HIGH, indicating that the search engine has been set to set a
mark when a waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency.
2-724 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:LOGIc:INPut:RF_PHASe
This command specifies the logic criteria to search for when searching
using the RF Phase vs. Time trace. This command is used together with
SEARCH:SEAR
CH<x>:TRIGger:A:LOGIc:THReshold:RF_PHASe to search for
an instance when the waveform goes higher or lower than the specified threshold.
“LOW” specifies to only set a mark when the RF_PHASe waveform drops below
the logic threshold; likewise, “HIGH” specifies to set a mark when the waveform
goes above the logic threshold. “X” specifies “don’t care”.
SEARCH<x> is the
search number, which is always 1.
Conditions
MDO4000/B/C models only.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_PHASe {HIGH|LOW|X}
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_PHASe?
Related Commands
TRIGger:A:LOGIc:INPut:RF
Arguments
HIGH specifies a high logic level.
LOW specifies a low logic level.
X specifies a “don't care” condition.
Examples
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:RF_PHASe HIGH specifies
to set a mark when the waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_PHASe.
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:RF_PHASe? might
return HIGH, indicating that the search engine has been set to set a
mark when a waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_PHASe.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:PATtern:INPut:CH<x>
This command specifies the logic condition to be used in a logic search when the
input is an analog channel.
SEARCH<x> is the search number, which is always 1,
and
CH<x> is the analog channel number, which can be 1– 4.
Group
Search
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-725