INFINITY EX CAME RA SY STEM MO DULE CALIBR ATIONS 17.5 4 Technical Reference Manual Chapter Issue 3 Jan 04 Associated Calibrations Figure 17-10 Associated Calibrations NO TE F or information on the T est Cycles function …

INFINITY
EX
CAMERA SYSTEM MODULE
CALIBRATIONS
17.52 Technical Reference Manual Chapter Issue 3 Jan 04
image) so that it is placed centrally about the box displayed.
18.Press the Step key, the camera moves to the second screen fiducial (rear left).
Position the fiducial centrally about the box as previously described.
19. Press the Step key, the camera moves to the third screen fiducial (front right).
Position the fiducial centrally about the box as previously described.
20.Press the Step key, the camera returns to the home position and the rising
table moves to the print height position.
21.Press the Step key. The print carriage moves to the rear position.
22.Using a piece of squeegee apply paste to all 25 positions on the screen.
23.Press the Step key twice.
24.Press Accept. The table drops and the camera moves to the first fiducial
(front left).
25.Press Brd Fid Setup and carry out Fiducial Setup procedure as described in
the User Manual, Machine Programming Chapter, Vision System.
26.Press Scrn Fid Setup and carry out Fiducial Setup procedure as described
in the User Manual, Machine Programming Chapter, Vision System.
27.Press the Step key. The Offset results are displayed.
28.Press the Single key. The camera moves to each of the 24 remaining fiducial
positions in turn.
Accept
Move
Carriage
Reject
Step
Brd Fid
Setup
Scrn Fid
Setup
Single Exit
Step
Brd Fid
Setup
Scrn Fid
Setup
Single Exit
Step
Brd Fid
Setup
Scrn Fid
Setup
Single Exit
Step Head Single Exit

INFINITY
EX
CAMERA SYSTEM MODULE
CALIBRATIONS
Chapter Issue 3 Jan 04 Technical Reference Manual 17.53
29.Select Auto Board. The board data file is saved.
This completes Offset Calibration.
NOTE
DEK recommends that the calibration screen is cleaned with Isopropyl Alcohol
(IPA) impregnated lint free wipe on completion of the offset calibration.
If for any reason the camera cannot locate the board or screen fiducial (except
for poor prints), press the fiducial setup key and relearn the fiducial, (refer to
the User Manual, Machine Programming Chapter, Vision System for details on
setting up the Fiducial Recognition Strategy).
Press Retry the sequence continues and does not have any detrimental effect to
the calibration.
Auto
Board
Manual
Board
Knead
Paste
Exit

INFINITY
EX
CAMERA SYSTEM MODULE
CALIBRATIONS
17.54 Technical Reference Manual Chapter Issue 3 Jan 04
Associated Calibrations
Figure 17-10 Associated Calibrations
NOTE
For information on the Test Cycles function see the Screen Alignment Chapter.
Camera Replacement
Motor Replacement
Offset Calibration
Belt Replacement
X Axis to Front Rail
Parallelism
X & Y Home Position
Vision Height
Y Belt Tension
X Belt Tension
X & Y Home Position
Camera Reference Position
Vision Calibration