SUSS / PA300 / User Manual / M10-121841-00 / February 2002
Indicator Part
SESC
Criteria
Actual Value Notes
Sitting work place
computer: sitting
work place
table to be
supplied by
customer
Location of frequently
used controls: should
not be located below
533 mm or above 1378
mm; interpolate for
immediate values:
1397 mm
1270 mm
1169 mm
1067 mm
965 mm
864 mm
762 mm
660 mm
533 mm
distance from
operator to
controls:
356 mm
432 mm
470 mm
483 mm
483 mm
470 mm
445 mm
381 mm
254 mm
N/A
Location of infrequently
used controls: should
not be located below
762 mm or above 1067
mm; interpolkate for
immediate values:
1067 mm
1016 mm
995 mm
914 mm
864 mm
813 mm
762 mm
distance from
operator to
controls:
330 mm
368 mm
394 mm
406 mm
419 mm
419 mm
419 mm
table to be
supplied by
customer
see
section
7.2.2
Workstation Design
Standing work place Prober
Work surface edge
radius where the
operator can assume a
static posture in contact
with the edge
radius min. 6.4
mm
work surface
should not be
leant against
see
section
2.9.1
Height of keyboard,
trackball or mouse
max. 1020 mm
min. 970 mm
location to be
determined by
customer
see
section
2.9.1
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
55
Indicator Part
SESC
Criteria
Actual Value Notes
Height of microscope
eyepieces. Must be
adjustable.
max. 1730 mm
min. 1270 mm
1430 mm Yes
Foot clearance:
Height:
Depth:
Width:
min. 150 mm
min. 150 mm
min. 490 mm
in front:
550 mm
270 mm
790 mm
Yes
Sitting work place
computer: sitting
work place
table to be
supplied by
customer
Height of keyboard,
trackball or mouse
max. 760 mm
min. 710 mm
location to be
determined by
customer
see
section
2.9.1
Vertical leg clearance min. 623 mm
location to be
determined by
customer
see
section
2.9.1
Horizontal leg
clearance, depth at
knee level
min. 508 mm
location to be
decided by
customer
see
section
2.9.1
Horizontal leg
clearance, depth at foot
level
min. 660 mm
(254 mm for
vertical foot
clearance)
location to be
detremined by
customer
see
section
2.9.1
Horizontal leg
clearance, width
min. 610 mm
location to be
determined by
customer
see
section
2.9.1
Height of microscope
eyepiece. Must be
adjustable
max. 1370 mm
min. 1220 mm
N/A
Thickness of work
surface
max. 51 mm
location to be
determined by
customer
see
section
2.9.1
Work surface edge
radius where the
operator can assume a
static posture in contact
with the edge
radius min. 6.4
mm
location to be
determined by
customer
see
section
2.9.1
56
SUSS / PA300 / User Manual / M10-121841-00 / February 2002
10.4 Installation Checklist
This is the Installation Checklist for the SUSS PA300 Semiautomatic Prober. Review
and complete the check points to ensure proper installation and operation.
All transport locks have been removed, including microscope
movement, platen and chuck stage
All components and accessories have been installed
All electrical interconnects are made and secured
Non-electrical utilities are connected, including vacuum and air
All components requiring AC power are plugged into outlets with
the required voltage and VA or wattage
Apply power and ensure the EMO system operates (including the
loading lid interlock)
Check functionality of all components
Check operation of chuck stage, microscope movement and
platen, ensuring free and smooth movement throughout their
entire range of travel
Check (and adjust if required) the planarity of the chuck stage
and microscope movement
Ensure all safety covers, panels and enclosures are installed or
secured
Ensure that the customer is trained in the proper operation of the
system