• Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation. • Measurement Category III. For measurements performed in the building installation. • Measurement Cat…

100%1 / 62
Compliance information
This section lists the safety and environmental standards with which the instrument complies. This product is intended for use by
professionals and trained personnel only; it is not designed for use in households or by children.
Compliance questions may be directed to the following address:
Tektronix, Inc.
PO Box 500, MS 19-045
Beaverton, OR 97077, USA
tek.com
Safety compliance
This section lists the safety standards with which the product complies and other safety compliance information.
EU declaration of conformity – low voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment
U.S. nationally recognized testing laboratory listing
UL 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment
Canadian certification
CAN/CSA-C22.2 No. 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test
equipment
Additional compliances
IEC 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment
Equipment type
Test and measuring equipment.
Pollution degree rating
Pollution degree 2 (as defined in IEC 61010-1). Rated for indoor, dry location use only.
IP rating
IP20 (as defined in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see
specific ratings marked on the product and in the manual).
Overvoltage Category I. For equipment intended to be connected to a mains supply in which means have been taken to substantially
and reliably reduce transient overvoltages to a level where they cannot cause a hazard.
Compliance information
High Voltage Differential Probes THDP0100/0200 and TMDP0200 User Manual 9
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category IV. For measurements performed at the source of low-voltage installation.
Note: Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a measurement
category rating. Other circuits within the product do not have either rating.
Mains overvoltage category rating
Overvoltage category II (as defined in IEC 61010-1).
Environmental compliance
This section provides information about the environmental impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling Production of this equipment required the extraction and use of natural resources. The equipment may
contain substances that could be harmful to the environment or human health if improperly handled at the
product’
s end of life. To avoid release of such substances into the environment and to reduce the use of
natural resources, we encourage you to recycle this product in an appropriate system that will ensure that
most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives
2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about
recycling options, check the T
ektronix Web site (www.tek.com/productrecycling).
Compliance information
10
Overview
This document provides operating information and specifications for the Tektronix THDP0100/0200 and TMDP0200 high voltage differential
probes. The probes share similar functions, properties, and operating procedures, and are discussed first, followed by the specifications,
which differ by probe model. Service procedures include performance verification and adjustments.
WARNING: Only use the accessories that are designed for your probe and that are rated at or above the voltages you are
measuring.
Name Bandwidth Peak voltage range Oscilloscope interface
THDP0100 100 MHz 600 V/6000 V TekVPI
THDP0200 200 MHz 150 V/1500 V TekVPI
TMDP0200 200 MHz 75 V/750 V TekVPI
Figure 1: THDP0100 probe with accessories
Overview
High Voltage Differential Probes THDP0100/0200 and TMDP0200 User Manual 11