1 61 9 6A Parallel el ec trode S MD test f ix ture T erminal connector : 7 mm DUT connec tion: 2 -Te r m i n a l Elec trical leng th: 26.2 mm Dimensions ( appr o x.): 1 4 0 ( W) x 4 8 (H) x 78 (D) [ mm] Wei gh t (ap pr o…

100%1 / 64
Up to 3 GHz (7 mm): SMD continued
Exchanging the device holder
1. Remove the ground plate
2. When measuring SMD, attach the
knob on the device holder.
3. Select the device holder suitable for
the device type. Loosen its knob and
insert into the arm.
4. Set the ground plate.
Placing the device
16194A High temperature component test fixture continued
Options:
16194A-010: Add EIA/EIAJ industrial standard sizedshorting bar set
16194A-701: Add general sized shorting bar set
Compensation and measurement: Before beginning the measurement, the appropriate
device holder (for a SMD or lead component) must be prepared with the text fixture. The
following figure shows how the device holder is exchanged to match the device type.
The next step is to perform open and short compensations in combination with the
electrical length compensation. When measuring above 500 MHz, load compensation
is also recommended. The fixture’s electrical length must be entered into the electrical
length compensation function of the measurement instrument first. Then open com-
pensation is performed by separating the high and the low electrodes from each other.
The separation should be equivalent in size to the DUTs width. Short compensation is
performed by using the option 16194A-010/701 shorting bar set. Load compensation is
performed by using the furnished 50 Ω SMD chip resistor. After performing open, short,
and load compensations in combination with the electrical length compensation, the
DUT is inserted into the test fixture. The following figures show how measurement is
performed.
33 | Keysight | Accessories Catalog for Impedance Measurements - Catalog
16196A Parallel electrode SMD
test fixture
Terminal connector: 7 mm
DUT connection: 2-Terminal
Electrical length: 26.2 mm
Dimensions (approx.):
140 (W) x 48 (H) x 78 (D) [mm]
Weight (approx.): 250 g
Additional error:
f: frequency [GHz]
Fixture overview
Up to 3 GHz (7 mm): SMD continued
Description: This test fixture is designed for impedance evaluations of parallel electrode
SMDs. It achieves stable frequency characteristics up to 3 GHz and provides highly
repeatable measurements. The applicable SMD size code is 0603 (inch)/1608 (mm).
Applicable instrument: E4982A, E4990A + 42942A*, E4991B, E5061B-3L3/3L4/3L5
with Opt. 005 + 16201A
* Option E4990A-120 is required
Frequency: DC to 3 GHz
Maximum voltage: ±42 V peak max. (AC +DC)
Operating temperature: –55 to +85°C
DUT size: The applicable SMD size is 0603 (inch)/1608 (mm). For details, see the figure
below.
The 16196A is furnished with three different insulator assemblies, since any gaps
between the DUT and the cylindrical insulator will result in improper positioning and
subsequent measurement errors. Select an insulator assembly that reduces the gap the
most. See the table below for dimensions of the insulator assemblies.
1. Opt. 16196A-710 only
Type of error Impedance
Proportional error 1.0 x f
2
[%]
Open repeatability 5 + 40 x f [
µ
S]
Short repeatability 30 + 125 x f [mΩ]
Description P/N Qty.
Operation and service manual 16196-90040 1
Insulator assembly ϕ1.34 mm 16196-60112 1
Insulator assembly ϕ1.14 mm 16196-60113 1
Insulator assembly ϕ1.08 mm 16196-60114 1
Open plate 16196-29002 1
Short plate 16196-29026 1
Push ring 16196-24004 1
Magnifying Lens
1
16193-60002 1
Tweezers 8710-2081 1
Wrench 8710-0909 1
Cleaning rod 5182-7586 1
Carrying case 16196-60150 1
Hole diameter of insulator
assembly (mm)
SMD case size examples
Length, width, height (mm)
16196A ϕ1.34 1.6 x 0.8 x 0.8
ϕ1.14 1.6 x 0.8 x 0.6
ϕ1.08 1.6 x 0.8 x 0.5
E4982A with 16196A
Furnished accessories:
34 | Keysight | Accessories Catalog for Impedance Measurements - Catalog
Up to 3 GHz (7 mm): SMD continued
16196A Parallel electrode SMD test fixture continued
Options:
16196A-710 : Add the magnifying lens and tweezers
To maintain adequate measurement performance, keep the electrodes and the short
plate in good condition. Contaminants and abrasion on these parts considerably affect
measurement results, especially for low value measurements. Periodic fixture cleaning
and part replacement is recommended to avoid deterioration of measurement perfor-
mance. The 16196x fixtures are designed with simplicity in mind, so that an operator
can easily replace parts. Spare parts, which are likely to be abraded, are supplied with
the 16196U Maintenance Kit.
16196U Maintenance kit
Opt. 16196U-010: Upper electrode, 5 piece set (common to 16196A/B/C models)
Opt. 16196U-100: Short plate for 0603 (inch)/1608 (mm) size, 5 piece set (for 16196A)
Opt. 16196U-110: Lower electrode, 5 piece set (for 16196A)
Compensation and measurement: First of all, install the appropriate insulator as-
sembly into the fixture. Then, perform compensation. Open and short compensations
are recommended in combination with the electrical length compensation before
measurement. The fixture’s electrical length must be entered into the electrical length
compensation function of the measurement instrument first. Next, open compensation
is performed by placing the furnished open plate on top of the insulator assembly.
Short compensation is performed by placing the furnished shorting plate on top of the
insulator assembly. After performing open and short compensations in combination
with the electrical length compensation, the DUT is inserted into the test fixture. Once
the measurement of the DUT is complete, remove the DUT from the fixture, by using the
furnished push ring. The following figures show how compensation and measurement is
performed.
35 | Keysight | Accessories Catalog for Impedance Measurements - Catalog