2 Functions for the Adjustment of the Machine User Manual Test Program SITEST 2.1 Axis Functions Software Version 405.xx Issue 01/99 2 - 10 ● Change th e value as requir ed and c onfir m your entr y b y cli cking on Acce…

100%1 / 168
User Manual Test Program SITEST 2 Functions for the Adjustment of the Machine
Software Version 405.xx Issue 01/99 2.1 Axis Functions
2 - 9
Adjusting the P gain for the z-axis:
Click on the
Adjust P-gain...
button. The following setting box opens.
Select the lines containing the parameters you wish to edit by double-clicking, and edit the value or the
positioning type as required
.
Once you have completed all desired entries or changes, click on the
Start
button in the setting box.
The continuous run for the adjustment of the P gain is started, and a dialog box displaying the start time
opens.
Click on the
Cancel
button in the dialog box to terminate the continuous run.
The completion time of the continuous run is displayed in the dialog box. The label of the "Cancel"
button changes to "OK".
Close the dialog box by clicking on
OK
.
Adjusting the P gain for the dp-axis:
To adjust the P gain of the dp-axis, click on the
Adjust P-gain...
button.
The setting box for the parameter group opens.
All further steps are the same as described in section 2.1.1, page 2 - 6.
If you wish to index the star axis in the continuous run mode, click on the
Star continuous run
button in
the "Star axis" display.
The input box below opens.
2 Functions for the Adjustment of the Machine User Manual Test Program SITEST
2.1 Axis Functions Software Version 405.xx Issue 01/99
2 - 10
Change the value as required and confirm your entry by clicking on
Accept
.
The continuous run is started and a dialog box displaying the start time opens.
Click on the
Cancel
button in the dialog box to terminate the continuous run.
The completion time of the continuous run is displayed in the dialog box. The label of the "Cancel"
button changes to "OK".
Close the dialog box by clicking on
OK
.
If the calibration tool position is to be approached, click on the
Calibration tool pos.
button.
To index the star, continue to click on the
Step star axis
button until the number of the desired
(center) sleeve is displayed in the bottom position (see Fig. 2.1.2).
User Manual Test Program SITEST 2 Functions for the Adjustment of the Machine
Software Version 405.xx Issue 01/99 2.1 Axis Functions
2 - 11
2.1.3 IC Head Axis Functions (SIPLACE 80F
4
/80F
5
)
In the "IC head" display (see Fig. 0.3.6) click on the icon to switch to the "IC head, axis func-
tions" display.
Fig. 2.1.3 "IC head, axis functions" Display
To perform the individual functions, proceed as follows:
NOTE
Since the layout of some of the setting boxes is identical or similar to that for the gantry axis functions and RV
head axis functions (see section 2.1.1 and section 2.1.2), these boxes will not be described and displayed
again in this section. 2
In the "Axis" field activate the radio button for the axis to be adjusted ("z-axis" or "dp-axis").
Select the Axis reference run function if the axis has not been referenced yet.
To move the axis to a given position in one operation, click on the Position axis.. button. The setting
box for the "Target position" and "Positioning type" opens.
All further steps are the same as described in section 2.1.1, page 2 - 4.