Manual_BSA_II_英文操作说明书.pdf - 第14页

14 BSA II Operating Instructions Operating Design 4.2.1 Buttons Symbol Designation Description [On/Off] button − When the butt on is pressed: Switches the opera ting display on. − If the button is held down: Switches the…

100%1 / 53
BSA II Operating Instructions 13
Operating Design
4 Operating Design
4.1 Operating Display in Weighing
Mode
3
4
5
6
78
9
1
2
Fig. 7: Operating display in weighing mode (example)
Pos. Designation Description
1 Metrological
Data
2 Menu
3 Application
overview
Displays the selected ap-
plication program during
operation
4 Toolbar
5 Weighing unit Displays the selected unit,
e.g., grams, [g]
6 Navigation bar For navigation in the
menu and system settings
7 Measurement
display
8 Visual touch
feedback
9 Toolbar
4.2 Menu and System Settings
Display
3
1
2
Fig. 8: Menu and system settings display (example)
Pos. Designation Description
1 Menu level Shows the position of the
displayed menu or param-
eter in up to 4 levels
2 Menu or param-
eter entry
3 [Selection] dis-
play
14 BSA II Operating Instructions
Operating Design
4.2.1 Buttons
Symbol Designation Description
[On/Off] button When the button is pressed: Switches the operating display on.
If the button is held down: Switches the operating display off.
[Menu] button When the button is pressed: The settings menu opens.
If the button is held down: It switches to version display.
[Zero] button Zeroes the device.
[Tare] button Starts taring.
[Print] button Exports the readouts to the integrated data interfaces.
[Adjust] button Starts the set calibration and adjustment function.
[GLP] button Exits the GLP printout and starts printing the GLP footer.
If the “Net total”, “Totalizing”, or “Statistics” application is active: Prints
and deletes the saved values and exits the application.
[Unit change] button If the “Unit change” function is active:
If the button is held down: Accesses the “Unit change” function menu.
When the button is pressed: Switches between the basic unit display
and up to 4 other units.
[Back] button In the menu:
When the button is pressed: Returns to the previous display.
If the button is held down: Saves the menu settings.
When entering digits: Selects the previous digit position.
For an active application: Exits the application and deletes the set ref-
erence value.
[Up] button In the menu: Scrolls through the menu levels or the available parame-
ters.
When entering digits: Increases the displayed value.
In the main display of an active application: Switches to the display of
the current weight value/parameter.
[Down] button In the menu: Scrolls through the menu levels or the available parame-
ters.
When entering digits: Decreases the displayed value.
In the main display of an application that is not active: Opens the dis-
play to set the reference values.
In the main display of an active application: Switches to the display of
the current weight value/parameter.
[Confirm] button In the menu: Opens the displayed menu level or confirms the dis-
played parameter.
When entering digits: Selects the next digit position.
In the main display of an application that is not active: Starts the appli-
cation process and saves the set reference value.
In the main display of an active application: Adopts the next compo-
nent or the next parameter.
BSA II Operating Instructions 15
Operating Design
4.3 Displays in the Operating Display
Symbol Designation Description
[Counting] display Indicates that the “Counting” application is selected.
[Percent weighing] dis-
play
Indicates that the “Percent weighing” application is selected.
[Calculation] display Indicates that the “Calculation” application is selected.
[Animal weighing] display Indicates that the “Animal weighing” application is selected.
[Net total] display Indicates that the “Net total” application is selected.
[Totalizing] display Indicates that the “Totalizing” application is selected.
[Density determination]
display
Indicates that the “Density determination” application is selected.
[Statistics] display Indicates that the “Statistics“ application is selected.
[Busy] display Indicates that the device is processing a command.
[Sign] display Indicates whether the value being displayed is positive or negative.
[Zero] display For some conformity-assessed devices: Indicates that the device has
been zeroed.
[AUTO] display Indicates that the “Animal weighing” application starts automatically.
[Application help] display Indicates the number of components for “Totalizing“, “Net total“, and
“Statistics“.
[R12] display Indicates the active range for multi-range balances.
[Printer] display Indicates that a printer has been detected at the USB port.
Flashes if the data output is active.
[PC-Connect] display Indicates that a PC or a second display has been detected at the USB
port.
Flashes if the data connection is active.
[Percent] display Indicates that a percentage value is being displayed.
[Net] display Indicates that a net value is being displayed.
[Gross] display Indicates that a gross value is being displayed.