BSA II Operating Instructions 39 Malfunctions 10 Malfunctions 10.1 W arning Messages W arning mes- sage Malfunction Cause Solution APP .ERR. The device has measur ed an invalid weight value. The applied weight is t oo lo…

100%1 / 53
38 BSA II Operating Instructions
Cleaning and Maintenance
9.1.4 Cleaning the Device, Draft Shield, and
Associated Components
Procedure
t Only use suitable cleaning agents and cleaning pro-
cedures and observe the product information for the
cleaning agent used (see Chapter “15.12 Cleaning
Agents and Cleaning Procedures”, page 50).
t Remove dust and powdery sample residue with a
brush or hand-held vacuum cleaner.
t Only slightly moisten cleaning cloths.
t Wipe the housing and the dismounted components
with a slightly damp cloth.
9.2 Mounting the Analytical Draft
Shield
Procedure
t Slowly slide the side
panels and the upper
panel into the frame
rails.
t Hold the handles on
the side panels or the
upper panel from the
outside and fasten to
the inner side with the
screws provided.
9.3 Mounting the Removed
Device Components
Procedure
t Mount the removed device components, e.g. mount
the weighing pan and associated components (see
Chapter 5.4, page29 and Chapter 5.5,
page30).
9.4 Recommissioning
Procedure
t Connect the device to the power supply (see Chap-
ter 6.2, page31).
9.5 Maintenance Schedule
Interval Component Activity
12 to 24months,
depending on op-
erating conditions
Device Contact Sarto-
rius Service.
9.6 Software Update
For a software update, contact Sartorius Service.
BSA II Operating Instructions 39
Malfunctions
10 Malfunctions
10.1 Warning Messages
Warning mes-
sage
Malfunction Cause Solution
APP.ERR. The device has measured an
invalid weight value.
The applied weight is too low. Increase the applied weight
to more than the minimum
load.
The weight value is negative.
No sample has been placed
on the balance.
Place the sample on the bal-
ance.
DIS.ERR. The value to be output can-
not be shown in the operating
display.
The data to be displayed is
not compatible with the set
display format.
Adjust the display settings in
the menu, e.g. resolution,
unit, decimal places.
HIGH or
ERR55
The device is overloaded. The device’s maximum weigh-
ing capacity has been ex-
ceeded
Reduce the applied weight
to below the device’s maxi-
mum weighing capacity.
LOW or
ERR54
The modulation of the weigh-
ing converter inside the de-
vice is too low.
No weighing pan has been
placed on the balance.
Insert the weighing pan into
the device and switch the
device off and on again.
A previously forgotten weight
was removed after starting
the device.
An error exists in the weighing
system or in the device elec-
tronics.
Contact Sartorius Service.
COMM.ERR. The device is not receiving
any weight values.
No communication exists be-
tween the control unit and
the load cell.
Wait until the control unit re-
stores the communication
with the load cell.
If the problem occurs again:
Contact Sartorius Service.
PRT.ERR. The [Print] key is locked. The data interface for printout
is set to xBPI mode.
Reset the menu to the facto-
ry settings.
If the problem occurs again:
Contact Sartorius Service.
SYS.ERR. The system data is faulty. A memory error exists in the
control unit.
Switch the device off and on
again.
If the problem occurs again:
Contact Sartorius Service.
ERR 02 The device cannot be adjust-
ed when starting the adjust-
ment function due to a zero
point error.
The device was not zeroed
before the adjustment pro-
cess.
Zero the device.
Check the preload and set if
necessary.
The device is loaded. Remove the sample from the
weighing pan.
40 BSA II Operating Instructions
Malfunctions
Warning mes-
sage
Malfunction Cause Solution
ERR 10 Taring is not possible. The device cannot be manu-
ally tared because an applica-
tion program has the tare
memory reserved.
To release the tare memory:
Exit the application program
with the [Back] button.
ERR 11 The weight value cannot be
saved in the tare memory.
The weight value is negative
or “zero”.
Check the sample being
weighed.
Zero the device before plac-
ing the sample on the bal-
ance if necessary.
10.2 Troubleshooting
Malfunction Cause Solution Chapter,
Page
The operating display is
blank.
The device is disconnected. Check the connection to
the power supply.
The power supply unit is not connect-
ed.
Connect the power supply
cable to the power supply.
6.2, 31
The displayed weight
value changes con-
stantly.
The device installation site is unstable. Adjust the parameter for the
ambient conditions.
7.1, 32
Change the installation site.
A foreign object is positioned between
the weighing pan and the housing.
Remove the foreign object.
The weight readout dis-
played by the device is
obviously wrong.
The device has not been calibrated and
adjusted.
Calibrate and adjust the de-
vice.
7.2, 32
The device was not tared before weigh-
ing.
Tare the device.
For a conformity-as-
sessed device: The [In-
valid weight value] dis-
play is displayed.
The cause of this fault is displayed after
pressing the [Unit change] key. At the
same time, the “Toggle between
weight units” function is locked.
WARMU.xx.: The device is in the warm-
up phase and has not yet reached its
operating temperature. xx = remaining
time in minutes
Comply with the warm-up
time after switching the de-
vice on.
15.7, 46
VALUE.ERR.: The displayed value is in-
valid.
Set the device to zero.