Manual_BSA_II_英文操作说明书.pdf - 第23页
BSA II Operating Instructions 23 Operating Design 4.6.5 “DEVICE“/“EXTRAS“ Menu P arameters Settings Explanation MENU EDIT ABL.* Activat es writ e access. T he menu parameters can be changed. RD.ONL Y Activates read acces…

22 BSA II Operating Instructions
Operating Design
4.6.4 “DEVICE“/“USB“ Menu
Parameters Settings Explanations
DAT.REC.** XBPI Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the device.
SBI* Enables SBI communication. The data is output to a PC or control unit. En-
ables the use of ESC commands from a PC to control the basic balance
functions with ASCII protocol.
REM.DISP. Enables data output on another display.
YDP20 Sets the standard settings of YDP20 printers.
YDP30 Sets the standard settings of YDP30 printers.
OFF Deactivates the data output.
BAUD** 600, 1200, 2400,
4800, 9600*,
19200, 38400,
57600, 115200
Sets the baud rate to the selected value.
PARITY** ODD* Applies an odd parity.
EVEN Applies an even parity.
NONE Does not apply a parity.
STOPBIT ** 1 BIT*
2 BITS
HANDSHK.** SFTWARE Sets the handshake protocol to software handshake.
HRDWARE Sets the handshake protocol to hardware handshake.
NONE* Does not set a handshake protocol.
DATABIT** 7 BITS Sets the number of data bits to 7.
8 BITS* Sets the number of data bits to 8.
DEV.USED NONE*, PRINT-
ER, VIRT.COM,
PC HOST, REM.
DISP.
Indicates which connection is detected at the USB port.
* Factory setting
** Are blocked if “PRINTER“ or “REM.DISP.“ are displayed under “DEV.USED“

BSA II Operating Instructions 23
Operating Design
4.6.5 “DEVICE“/“EXTRAS“ Menu
Parameters Settings Explanation
MENU EDITABL.* Activates write access. The menu parameters can be changed.
RD.ONLY Activates read access. The menu parameters cannot be changed.
SIGNAL OFF Switches the acoustic signal off.
ON* Switches the acoustic signal on.
KEYS UNLOCK.* Deactivates the button lock.
LOCKED Activates the button lock.
EXT.KEY PRINT Assigns the print function to the external key.
CAL. Assigns the calibration and adjustment function selected under [CAL.-ADJ.]
to the external key.
CF Assigns the [Back] function to the external key.
ENTER Assigns the [Confirm] function to the external key.
ZERO Assigns the [Zero] function to the external key.
TARE Assigns the [Tare] function to the external key.
GLP END Assigns the [GLP] function to the external key.
NO* No function is assigned to the external key.
ON MODE ON/STDB.* The [On/Off] button switches between on/standby with clock display.
ON/OFF The [On/Off] button switches between on/standby without clock display.
AUTO ON Changes the function of the [On/Off] button: The device no longer switches
off or to standby mode, instead it starts a boot process.
LIGHT OFF Deactivates the lighting on the operating display.
ON* Activates the lighting on the operating display.
* Factory setting

24 BSA II Operating Instructions
Operating Design
4.6.6 “DATA.OUT.“/“COM. SBI“ Menu
Parameters Settings Explanations
COM. OUTP. IND.NO* Activates the manual data output without stability.
IND.AFTR Activates the manual data output after stability.
AUTO.W/O Activates the automatic data output without stability.
AUTO W/ Activates the automatic data output after stability.
STOP.AUT. OFF* Deactivates the option to abort the automatic data output.
ON The automatic data output is aborted by pressing the [Print] button or a soft-
ware command.
AUTO.CYC. EACH.VAL* Starts the automatic data output with a cycle after each value.
2.WERT Starts the automatic data output with a cycle after every 2nd value.
INTERV. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.“
FORMAT 22CHARS* The data output provides 22 characters per line (16 characters for the mea-
sured value and 6 characters for identifiers).
16CHARS The data output provides 16 characters per line for the measured value.
EXTR.LIN. The data output provides an additional line with the date, time, and weight
value.
AUTO.TAR. OFF* Deactivates automatic taring after data output.
ON The device automatically tares after data output.
* Factory setting
4.6.7 “DATA.OUT.“/“PRNT.PAR.“ Menu
Parameters Settings Explanation
ACTIVAT. MAN. NO Manual without stability: The print process can be started manually at any
time.
MAN.AFTR* Manual after stability: After pressing the [Print] button, the print command is
only executed once stability is achieved.
INTERV.A. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.” if
stability is achieved.
INTERV.N. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.”
without stability.
AUTO.LC Automatically at load change: The print process starts after every load
change.
FORMAT 22CHARS* The printer output prints 22 characters per line (16 characters for the mea-
sured values and 6 characters for identifiers)
EXTR.LIN. The printer output prints an additional line with the date, time, and weight
value.