32 BSA II Operating Instructions
System Settings
7 System Settings
7.1 Performing System Settings
Default settings can be adjusted for the device and the
applications in order to align with the ambient condi-
tions and individual operating requirements.
The following settings are necessary to operate the de-
vice together with connected components:
Set up the communication of the connected devices
Set up additional components
The following settings are recommended to set up the
device:
Set the menu language
Set date and time
Set the calibration and adjustment
Procedure
t Press the [Menu] button.
t To adjust settings: Open the desired menu.
t Select and confirm the desired parameters (parame-
ters, see Chapter “4.6 Parameter List”, page 19).
t Exit the menu.
7.2 Setting the Calibration and
Adjustment
7.2.1 Setting Internal Calibration and
Adjustment (Only I-1x Models)
The following functions can be set for the internal cali-
bration and adjustment:
Internal calibration with automatic start of the ad-
justment.
Internal calibration with manual start of the adjust-
ment.
Procedure
t Open the “SETUP” / “BALANCE” menu. Call up the
“CAL./ADJ.” parameter and select the value “INT.
CAL.”.
t If the calibration function needs to be set with subse-
quent automatic adjustment: In the “SETUP” / “BAL-
ANCE” menu, for the “CAL.SEQ.“ parameter, select
the “ADJUST“ setting value.
t If the calibration function needs to be set without
subsequent automatic adjustment: In the “SET-
UP“ / “BALANCE“ menu, for the “CAL.SEQ.” parame-
ter, select the “CAL.-ADJ.setting value.
7.2.2 Setting the External Calibration and
Adjustment
The following functions can be set for the external cali-
bration and adjustment:
External calibration with manual start of the adjust-
ment.
If this relates to a conformity-assessed de-
vice in legal metrology: External calibration
and adjustment is not possible.
Procedure
t Open the “SETUP“ / “BALANCE“ menu.
t If the calibration function needs to be set with subse-
quent automatic adjustment: For the “CAL.SEQ.“ pa-
rameter, select the “ADJUST“ setting value.
t If the calibration function needs to be set without
subsequent automatic adjustment: For the “CAL.
SEQ.“ parameter, select the “CAL./ADJ.setting val-
ue.
Setting the Weight Value for the External Weight
A preset weight value can be set for the external weight.
Procedure
t In the “SETUP“ / “BALANCE“ menu, for the “CAL./
ADJ.“ parameter, select the “EXT.CAL.setting value.
BSA II Operating Instructions 33
Operation
8 Operation
8.1 Waiting for the Warm-up Time
In order to provide accurate weighing results, the device
must have reached the necessary operating tempera-
ture. The operating temperature is reached after the
warm-up time. The warm-up time starts when the device
is switched on.
Procedure
t Switch on the device.
t Wait until the operating temperature is reached
(warm-up time, see Chapter 15.7, page 46).
If a weighing process is carried out on a
conformity-assessed device during the
warm-up time: The weight value is marked
as invalid.
8.2 Leveling the Device with a
Level
Unevenness at the installation site of the device may re-
sult in incorrect weighing results. Leveling balances out
unevenness at the installation site.
Procedure
t To level the device: Bring the air bubble into the mid-
dle of the circular marking. To do this, turn the level-
ing feet to the left or right.
8.3 Overview of Calibration and
Adjustment
During calibration, a calibration weight is used to deter-
mine the deviation of the displayed value from the actu-
al value. The subsequent adjustment eliminates this
deviation.
We recommend regular calibration and adjustment:
Daily, every time the device is switched on
After every leveling
After changing the ambient conditions (temperature,
humidity, or air pressure)
After setting the device up at a new installation site
8.4 Calibrating and Adjusting the
Device Internally (Only I-1x
Models)
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The internal calibration and adjustment function is
set (see Chapter 7.2.1, page 32).
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
t The “INT.CAL.“ display is opened
t If the calibration with subsequent automatic adjust-
ment function is selected:
y The “CAL.RUN.“ display is opened in the operat-
ing display during the process.
y The “CAL.END.“ display indicates the end of the
calibration process.
t If the calibration without subsequent automatic ad-
justment function is selected:
t Press the [Confirm] button.
y The calibration process is started.
y The “CAL.RUN.“ display is opened in the operat-
ing display during the process.
y The calibration error is displayed.
t Press the [Confirm] button
y The adjustment process is performed.
y The “CAL.END.“ display indicates the end of the
calibration process.
34 BSA II Operating Instructions
Operation
8.5 Externally Calibrating and
Adjusting the Device
Material Calibration and adjustment weight
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The external calibration and adjustment function is
set (see Chapter 7.2.2, page 32).
The preset weight value for the adjustment weight or
the user-defined weight value for the adjustment
weight is selected (see Chapter 7.2.2, page 32).
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
t Place the indicated calibration and adjustment
weight on the balance.
y The weight value is displayed.
t If the calibration with subsequent automatic adjust-
ment function is selected and the adjustment weight
placed on the device is within the specified limits:
y The adjustment process starts.
y If the “+“ prefix is displayed: The applied weight is
too high.
y If the “–“ prefix is displayed: The applied weight is
too low.
y The “CAL.END.“ message indicates the end of the
calibration process.
t Remove the adjustment weight.
t If the calibration without subsequent automatic ad-
justment function is selected and the adjustment
weight placed on the device is within the specified
limits:
t Press the [Confirm] button.
y The adjustment process starts.
y If the “+“ prefix is displayed: The applied weight is
too high.
y If the “–“ prefix is displayed: The applied weight is
too low.
y The “CAL.END.“ message indicates the end of the
calibration process.
t Remove the adjustment weight.
8.6 Printing Results of the Calibra-
tion and Adjustment Process
The results of the calibration and adjustment process
can be printed in a GLP printout.
Procedure
t Set the GLP printout in the menu (setting parame-
ters, see Chapter 4.6.7, page 24).
t Adjust the device.
y Once the calibration is complete: The printing pro-
cess starts.
8.7 Weighing
Chemicals can attack the device or the connected ac-
cessories internally and externally. This may damage the
device and accessories.
We recommend using appropriate vessels when weigh-
ing chemicals.
Requirements
The device has been leveled.
The device is calibrated and adjusted.
Procedure
t Zero the device. To do this, press the [Zero] button.
t If using a vessel for the sample:
t Tare the device. To do this, press the [Tare] but-
ton.
t Place the sample in the vessel.
t If no vessel is used for the sample: Place the sample
on the weighing pan.
y The weight value is displayed depending on the
selected application program.