Manual_BSA_II_英文操作说明书.pdf - 第17页
BSA II Operating Instructions 17 Operating Design 4.5 Menu Structure 4.5.1 “Main Menu“ Menu Structure t Navigating the menus (see Chapter 4.4, page 16). Level 1: Level 2: Description SETUP BALANCE Set the functions of …

16 BSA II Operating Instructions
Operating Design
Symbol Designation Description
[Selection] display In the menu: Identifies the selected parameter.
If the “Calculation“ or “Density determination“ application is active: Indi-
cates that a calculated value is being displayed.
[Unit symbol] display Indicates the set weight unit, e.g., [g] for “grams”.
[Quantity] display Indicates that a quantity is being displayed.
[Invalid weight value] dis-
play
− Indicates that the display does not contain a weight value, but is in-
stead the calculated result of an application, e.g., for the “Totalizing”
application.
− For conformity-assessed devices: Indicates a fault. The cause of this
fault is displayed after pressing the [Unit change] key.
4.4 Navigating the Menus
t To call up the main menu: Press the [Menu] button.
t To display menu items or parameters of a level: Press the [Up] or [Down] button.
t To return to the next higher menu level or exit the menu: Press the [Back] button.
t To open a displayed menu level or a displayed parameter: Press the [Confirm] button.

BSA II Operating Instructions 17
Operating Design
4.5 Menu Structure
4.5.1 “Main Menu“ Menu Structure
t Navigating the menus (see Chapter 4.4, page 16).
Level 1: Level 2: Description
SETUP BALANCE Set the functions of the device.
GEN.SERV.
“General services”
Reset the menu to factory settings.
DEVICE RS-232
“RS232, 9-pin”
Define the parameters for the COM interface.
USB
“USB-C”
Define the parameters for the USB interface.
EXTRAS Define the functions of the operating display.
DATA.OUTP.
“Data output”
COM.SBI
“SBI communication”
Configure the automatic data output.
PRNT.PR.
“Printout settings”
Perform the settings for the printout.
APPLIC.
“Applications”
WEIGH − Determine the weight value of a sample.
− Activate the functions for all applications.
COUNT Determine the number of parts that have approximately equal weight.
PERCENT
“Percent weighing”
Determine the percentage share of a
sample based on a reference weight.
NET.TOT.
“Net total”
Carry out the weighing of components for a mixture.
TOTAL
“Totalizing”
Add weight values of independent weighing processes in a memory.
ANIMALW.
“Animal weighing”
Weigh unstable samples, e.g., animals. This program calculates the av-
erage of several measurement cycles.
CALC.
“Calculation”
Calculate the weight value using a multiplier or divisor, e.g., for deter-
mining the weight per unit area of paper.
DENSITY
“Density determination”
Determine the density of solid samples based on the buoyancy meth-
od.
STATIST.
“Statistics”
Save and statistically analyze weight values and calculated values.

18 BSA II Operating Instructions
Operating Design
Level 1: Level 2: Description
INPUT DEV.ID. Save the entered ID number for the device.
DATE Set the date.
TIME Set the time.
SERV. PW. Activate the service mode.
INTERV. The SBI output rate can be set from 0 - 9999 seconds.
INFO
“Device infor-
mation”
VERSION
“Version number”
Display the software version number.
SER. NO.
“Serial number”
Display the device’s serial number.
MODEL Display the device’s model designation.
BAC VER. Display the version of the BAC processor.
LANGUAG. Set the menu language of the operating display.
4.5.2 “Toggle Between Weight Units“ Menu Structure
t Navigating the menus (see Chapter 4.4, page 16).
Level 1: Description
Unit 1 – unit 4 Define the displayed weight unit and the resolution for the 1st to 4th convertible unit.