Manual_BSA_II_英文操作说明书.pdf - 第19页
BSA II Operating Instructions 19 Operating Design 4.6 P aramet er List 4.6.1 “SETUP“/“BALANCE“ Menu P arameters Settings Explanation AMBIENT V .ST ABLE Sets the ambien t conditions to “very st able”: Activates a fast cha…

18 BSA II Operating Instructions
Operating Design
Level 1: Level 2: Description
INPUT DEV.ID. Save the entered ID number for the device.
DATE Set the date.
TIME Set the time.
SERV. PW. Activate the service mode.
INTERV. The SBI output rate can be set from 0 - 9999 seconds.
INFO
“Device infor-
mation”
VERSION
“Version number”
Display the software version number.
SER. NO.
“Serial number”
Display the device’s serial number.
MODEL Display the device’s model designation.
BAC VER. Display the version of the BAC processor.
LANGUAG. Set the menu language of the operating display.
4.5.2 “Toggle Between Weight Units“ Menu Structure
t Navigating the menus (see Chapter 4.4, page 16).
Level 1: Description
Unit 1 – unit 4 Define the displayed weight unit and the resolution for the 1st to 4th convertible unit.

BSA II Operating Instructions 19
Operating Design
4.6 Parameter List
4.6.1 “SETUP“/“BALANCE“ Menu
Parameters Settings Explanation
AMBIENT V.STABLE Sets the ambient conditions to “very stable”: Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
STABLE* Sets the ambient conditions to “stable”. Recommended for the following
work environment:
− Stable table
− Slight movement in the room
− Slight draft
UNSTABL. Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate. Recommended for the following
work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
V.UNSTBL. Sets the ambient conditions to “very unstable”: Activates a significantly
delayed change in the weight values and long wait for stability with a further
reduction in the output rate.
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
APP.FILT. FINAL.RD.* Activates a filter that enables a fast change in the display for very fast load
changes. Display changes with minimal load changes (in the digit range) oc-
cur more slowly.
FILLING Activates a filter that enables a very fast change in the display with minimal
load changes, e.g., when dosing or filling vessels.
STB.RNG. V. ACC. Sets the stability to “very accurate“.
ACC.* Sets the stability to “accurate“.
FAST Sets the stability to “fast“.
V. FAST Sets the stability to “very fast“.
AUTOZER. ON* Activates automatic zeroing. The display is automatically set to zero in case
of a deviation of 0 less than (X).
OFF Deactivates automatic zeroing. Zeroing must be triggered with the [Zero]
button.
* Factory setting

20 BSA II Operating Instructions
Operating Design
Parameters Settings Explanation
WT.UNIT GRAMS*, KI-
LOGR., CARATS,
POUNDS, OUNC-
ES, TROY.OZ., HK-
TAEL, SNGTAEL,
TWNTAEL,
GRAINS, PENY-
WT., MILLIGR.,
PTS./LB, CHN-
TAEL., MOMMES,
AUSTR.CT, TOLA,
BAHT, MESGHAL,
NEWTON
The device displays the weight in the selected unit.
The availability of units depends on national legislation and is therefore
country-specific.
DISP.DIG. ALL* “Show all decimal places”: All decimal places are shown in the display. The
setting change is not available on conformity-assessed devices.
MINUS 1 “Last decimal place off”: The last decimal place is switched off.
CAL./ADJ. EXT.CAL. The [Adjust] button starts an external calibration and adjustment process
with the preset adjustment weight.
E.CAL.USR. The [Adjust] button starts an external calibration and adjustment process
with the user-defined adjustment weight value.
INT.CAL.* The [Adjust] button starts an internal calibration and adjustment process.
CAL.SEQ. ADJUST* The adjustment starts automatically after the calibration.
CAL.-ADJ. The adjustment must be started or exited manually after calibration with the
[Confirm] button.
* Factory setting
4.6.2 “SETUP“/“GEN.SERV.“ Menu
Parameters Settings Explanation
MENU.RES. DEFAULT Resets the system settings to the factory default settings.
NO* Deactivates the option of resetting the device menu.
* Factory setting