BSA II Operating Instructions 47 T echnical Data 15.8.2 Devices with F rame Draft Shield Models BSA6202 | BSA4202 | BSA3202 | BSA2202 Model BSA6202-1x | BSA6202I-1x BSA4202-1x | BSA4202I-1x BSA3202-1x | BSA3202I-1x BSA22…

46 BSA II Operating Instructions
Technical Data
15.7 Warm-up Time to Reach Operating Temperature
Unit Value
Time period between switching the device on and carrying out weighing tasks h 2
15.8 Metrological Data
15.8.1 Devices with Analytical Draft Shield
Model
BSA224-1x |
BSA224I-1x
BSA124-1x |
BSA124I-1x
BSA623-1x |
BSA623I-1x
BSA423-1x |
BSA423I-1x
BSA323-1x |
BSA323I-1x
BSA223-1x |
BSA223I-1x
Value Value Value Value Value Value
Readability | Scale interval (d) mg 0.1 0.1 1 1 1 1
Maximum capacity (Max) g 220 120 620 420 320 220
Repeatability
At approx. maximum capacity, typical
value
mg 0.1 0.1 1 1 1 1
Linearity deviation
Limits ± mg 0.2 0.2 2 2 2 2
Sensitivity drift between +10°C and +30°C ± ppm/K 1.5 1.5 2 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
Only for models with approval:
Accuracy class I I II II II II
Type BC-EB BC-EB BC-ED BC-ED BC-ED BC-ED
Verification scale interval (e) mg 1 1 10 10 10 10
Minimum load (Min) mg 10 10 20 20 20 20
Typical measurement time s ≤2.0 ≤2.0 ≤1.5 ≤1.5 ≤1.5 ≤1.5
Recommended calibration weight
External calibrated test weight g 200 100 500 200 200 200
Accuracy class in accordance with OIML
R111-1
E2 E2 F1 F1 F1 F1

BSA II Operating Instructions 47
Technical Data
15.8.2 Devices with Frame Draft Shield
Models BSA6202 | BSA4202 | BSA3202 | BSA2202
Model
BSA6202-1x |
BSA6202I-1x
BSA4202-1x |
BSA4202I-1x
BSA3202-1x |
BSA3202I-1x
BSA2202-1x |
BSA2202I-1x
Unit Value Value Value Value
Readability | Scale interval (d) mg 10 10 10 10
Maximum capacity (Max) g 6200 4200 3200 2200
Repeatability
At approx. the maximum load, typical
value
mg 10 10 10 10
Linearity deviation
Limits ± mg 20 20 20 20
Sensitivity drift between +10°C and +30°C ± ppm/K 2 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
Only for models with approval:
Accuracy class II II II II
Type BC-EE BC-EE BC-EE BC-EE
Verification scale interval (e) mg 100 100 100 100
Minimum load (Min) mg 500 500 500 500
Typical measurement time s ≤1.0 ≤1.0 ≤1.0 ≤1.0
Recommended calibration weight
External calibrated test weight g 5000 2000 2000 2000
Accuracy class in accordance with
OIML R111-1
F1 F1 F1 F1

48 BSA II Operating Instructions
Technical Data
Models BSA822 | BSA8201 | BSA5201 | BSA2201
Model
BSA822-1x |
BSA822I-1x
BSA8201-1x
BSA8201I-1x
BSA5201-1x |
BSA5201I-1x
BSA2201-1x |
BSA2201I-1x
Unit Value Value Value Value Value
Readability | Scale interval (d) mg 10 100 100 100 100
Maximum capacity (Max) g 820 8200 8200 5200 2200
Repeatability
At approx. the maximum load, typical
value
mg 10 100 100 100 100
Linearity deviation
Limits ± mg 20 300 100 100 100
Sensitivity drift between +10°C and +30°C ± ppm/K 2 7 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
Only for models with approval:
Accuracy class II II II II II
Type BC-EE BC-EI BC-EG BC-EE BC-EE
Verification scale interval (e) mg 100 1000 1000 100 100
Minimum load (Min) mg 500 5000 5000 5000 5000
Typical measurement time s ≤1.0 ≤1.0 ≤1.0 ≤1.0 ≤1.0
Recommended calibration weight
External calibrated test weight g 500 5000 5000 5000 2000
Accuracy class in accordance with
OIML R111-1
F2 F2 F2 F2 F2