Manual_BSA_II_英文操作说明书.pdf - 第25页
BSA II Operating Instructions 25 Operating Design P arameters Settings Explanation PRT .INIT . OFF Deactiva tes the output of the application parameters. ALL * The print command prin ts all parameters. MAIN.P AR. The pri…

24 BSA II Operating Instructions
Operating Design
4.6.6 “DATA.OUT.“/“COM. SBI“ Menu
Parameters Settings Explanations
COM. OUTP. IND.NO* Activates the manual data output without stability.
IND.AFTR Activates the manual data output after stability.
AUTO.W/O Activates the automatic data output without stability.
AUTO W/ Activates the automatic data output after stability.
STOP.AUT. OFF* Deactivates the option to abort the automatic data output.
ON The automatic data output is aborted by pressing the [Print] button or a soft-
ware command.
AUTO.CYC. EACH.VAL* Starts the automatic data output with a cycle after each value.
2.WERT Starts the automatic data output with a cycle after every 2nd value.
INTERV. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.“
FORMAT 22CHARS* The data output provides 22 characters per line (16 characters for the mea-
sured value and 6 characters for identifiers).
16CHARS The data output provides 16 characters per line for the measured value.
EXTR.LIN. The data output provides an additional line with the date, time, and weight
value.
AUTO.TAR. OFF* Deactivates automatic taring after data output.
ON The device automatically tares after data output.
* Factory setting
4.6.7 “DATA.OUT.“/“PRNT.PAR.“ Menu
Parameters Settings Explanation
ACTIVAT. MAN. NO Manual without stability: The print process can be started manually at any
time.
MAN.AFTR* Manual after stability: After pressing the [Print] button, the print command is
only executed once stability is achieved.
INTERV.A. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.” if
stability is achieved.
INTERV.N. The output rate can be set from 0–9999 seconds under “INPUT/INTERV.”
without stability.
AUTO.LC Automatically at load change: The print process starts after every load
change.
FORMAT 22CHARS* The printer output prints 22 characters per line (16 characters for the mea-
sured values and 6 characters for identifiers)
EXTR.LIN. The printer output prints an additional line with the date, time, and weight
value.
* Factory setting

BSA II Operating Instructions 25
Operating Design
Parameters Settings Explanation
PRT.INIT. OFF Deactivates the output of the application parameters.
ALL* The print command prints all parameters.
MAIN.PAR. The print command only prints the main parameters.
GLP OFF* Deactivates the GLP printout.
CAL./ADJ. Activates the GLP printout for all calibration and adjustment processes.
ALWAYS The GLP printout is always switched on. All printouts contain a GLP header
and a GLP footer.
TAR./PRT. OFF* Deactivates automatic taring after printer output.
ON Automatically tares the device after every printout.
TIME 24H* Sets the time to 24-hour mode.
12H − Sets the time to 12-hour mode (AM/PM).
− Is blocked for ISO format.
DATE DD.MMM.YY* Sets the date display format to DD.MMM.YY
MMM.DD.YY Sets the date display format to MMM.DD.YY
YY.MM.DD − Sets the date display format to ISO FORMAT YYYY-MM-DD.
− Sets the time to 24-hour mode.
* Factory setting
4.6.8 “APPLIC.“/“WEIGH“ Menu
Parameters Settings Explanation
UNIT ON* Activates the “Toggle between weight units” touch function.
OFF Deactivates the “Toggle between weight units” touch function.
RECALL ON Activates the saving of the last stable weight value not equal to 0.
OFF* Deactivates saving.
* Factory setting
4.6.9 “APPLIC.“/“COUNT“ Menu
Parameters Settings Explanation
RESOLUT. DISP.ACC.* Sets the resolution to “display resolution“. The “Counting” application is
started with the displayed value.
10FACH Sets the resolution to 10-times more accurate than “display resolution“.
100FACH Sets the resolution to 100-times more accurate than “display resolution“.
REF.UPDT. OFF* Deactivates automatic reference sample updating.
AUTO Activates the automatic reference sample updating.
* Factory setting

26 BSA II Operating Instructions
Operating Design
4.6.10 “APPLIC.“/“PERCENT“ Menu
Parameters Settings Explanation
DEC.PLCS NONE The result of the “Weighing in Percent” application is displayed without deci-
mal places.
1 DEC.PL.* The result of the “Weighing in Percent” application is displayed to 1 decimal
place.
2 DEC.PL. The result of the “Weighing in Percent” application is displayed to 2 decimal
places.
3 DEC.PL. The result of the “Weighing in Percent” application is displayed to 3 decimal
places.
* Factory setting
4.6.11 “APPLIC.“/“NET.TOT.“ Menu
Parameters Settings Explanation
PRT.COMP. OFF Deactivates the component printout.
ON* Activates the component printout.
* Factory setting
4.6.12 “APPLIC.“/“TOTAL“ Menu
Parameters Settings Explanation
PRT.COMP. OFF Deactivates the component printout.
ON* Activates the component printout.
* Factory setting
4.6.13 “APPLIC” / “ANIMALW” Menu
Parameters Settings Explanation
ACTIVIT. CALM Sets the intensity of the “Animal activity“ to “calm“. Recommended for minor
movements of the sample, e.g. caused by placement on the weighing pan.
ACTIVE* Sets the intensity of the “Animal activity“ to “active“. Recommended for me-
dium movements of the sample, e.g. caused by placement on the weighing
pan.
V.ACTIVE Sets the intensity of the “Animal activity“ to “very active“. Recommended for
very active movements of the sample, e.g. caused by placement on the
weighing pan.
* Factory setting