Manual_BSA_II_英文操作说明书.pdf - 第46页

46 BSA II Operating Instructions T echnical Data 15.7 W arm-up T ime to Reach Operating T emperatur e Unit V alue Time period bet ween switching the device on and carrying out weighing tasks h 2 15.8 Metrological Data 15…

100%1 / 53
BSA II Operating Instructions 45
Technical Data
15.6 Electrical Data
15.6.1 Power Supply
Unit Value
Power supply only permitted via the power supply cable and power
supply unit supplied by Sartorius
Sartorius power supply unit, type YEPS01-15VOW, with coun-
try-specific power plug adapters YEPS01-PS8, YEPS01-PS9,
YEPS01-PS10
Power supply unit
AC voltage V 100 – 240 (±10%)
Frequency Hz 50–60
Current consumption, maximum A 0.2
Device
DC voltage V 15 (± 5%)
Current consumption, maximum A 0.53
Power supply unit fuses
Quantity 1
Type: Short-circuit protection, electronic
Protection class according to IEC 62368-1
Power supply unit II
Device II
Overvoltage category according to IEC61010-1
Power supply unit II
Device II
15.6.2 Safety of Electromagnetic Equipment
Safety provisions according to IEC 61010-1 Safety requirements for electrical equipment for measurement, con-
trol, and laboratory use – Part 1: General requirements
15.6.3 Electromagnetic Compatibility
Electromagnetic compatibility, according to DIN EN 61326-1 Electrical equipment for measurement, control, and
laboratory use – EMC requirements - Part 1: General requirements (IEC 61326-1:2013)
Interference immunity: Suitable for use in industrial areas (Table 2 of the standard)
Transient emissions: Class B; suitable for use in residential areas and areas that are directly connected to a low volt-
age network that (also) supplies residential buildings.
46 BSA II Operating Instructions
Technical Data
15.7 Warm-up Time to Reach Operating Temperature
Unit Value
Time period between switching the device on and carrying out weighing tasks h 2
15.8 Metrological Data
15.8.1 Devices with Analytical Draft Shield
Model
BSA224-1x |
BSA224I-1x
BSA124-1x |
BSA124I-1x
BSA623-1x |
BSA623I-1x
BSA423-1x |
BSA423I-1x
BSA323-1x |
BSA323I-1x
BSA223-1x |
BSA223I-1x
Value Value Value Value Value Value
Readability | Scale interval (d) mg 0.1 0.1 1 1 1 1
Maximum capacity (Max) g 220 120 620 420 320 220
Repeatability
At approx. maximum capacity, typical
value
mg 0.1 0.1 1 1 1 1
Linearity deviation
Limits ± mg 0.2 0.2 2 2 2 2
Sensitivity drift between +10°C and +30°C ± ppm/K 1.5 1.5 2 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
Only for models with approval:
Accuracy class I I II II II II
Type BC-EB BC-EB BC-ED BC-ED BC-ED BC-ED
Verification scale interval (e) mg 1 1 10 10 10 10
Minimum load (Min) mg 10 10 20 20 20 20
Typical measurement time s ≤2.0 ≤2.0 ≤1.5 ≤1.5 ≤1.5 ≤1.5
Recommended calibration weight
External calibrated test weight g 200 100 500 200 200 200
Accuracy class in accordance with OIML
R111-1
E2 E2 F1 F1 F1 F1
BSA II Operating Instructions 47
Technical Data
15.8.2 Devices with Frame Draft Shield
Models BSA6202 | BSA4202 | BSA3202 | BSA2202
Model
BSA6202-1x |
BSA6202I-1x
BSA4202-1x |
BSA4202I-1x
BSA3202-1x |
BSA3202I-1x
BSA2202-1x |
BSA2202I-1x
Unit Value Value Value Value
Readability | Scale interval (d) mg 10 10 10 10
Maximum capacity (Max) g 6200 4200 3200 2200
Repeatability
At approx. the maximum load, typical
value
mg 10 10 10 10
Linearity deviation
Limits ± mg 20 20 20 20
Sensitivity drift between +10°C and +30°C ± ppm/K 2 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
Only for models with approval:
Accuracy class II II II II
Type BC-EE BC-EE BC-EE BC-EE
Verification scale interval (e) mg 100 100 100 100
Minimum load (Min) mg 500 500 500 500
Typical measurement time s ≤1.0 ≤1.0 ≤1.0 ≤1.0
Recommended calibration weight
External calibrated test weight g 5000 2000 2000 2000
Accuracy class in accordance with
OIML R111-1
F1 F1 F1 F1