The Ethernet port sends data to local networks or LIMS systems. Use a shielded cable (e.g., STP, FTP, S/FTP) that is 20 m (65.6 ft) or less. 4.4 Cell compartments and adapters The instrument has two cell compartments. Re…

• In a location that is not dusty, humid or wet. Always keep the surface of the instrument, the cell
compartment and all accessories clean and dry.
4.2 Startup
W A R N I N G
Fire hazard. Only use the supplied benchtop power supply LZV844.
C A U T I O N
Electrical shock and fire hazards. Make sure that the supplied cord and non‐locking plug meet the
applicable country code requirements.
N O T I C E
Always set the power switch to off before the power supply is connected to a electrical outlet or damage to the
instrument can occur.
N O T I C E
To prevent damage to the electronics and mechanics of the instrument, wait approximately 20 seconds after the
power switch is set to off before the power switch is set to on again.
Connect the power supply to the instrument. Refer to Figure 3.
Figure 3 AC power connection
4.3 Interfaces
The instrument has four USB ports and one Ethernet port. Refer to Figure 1 on page 7.
A USB hub connected to power can be used to connect multiple USB accessories at a time.
Note: Make sure that the USB cable is 3 m (9.8 ft) or less.
Send the saved data through the USB port to a printer. Use the USB-serial converter cable to
connect the instrument to a PC. Use the USB port to update the instrument software, export data and
certificates with a USB flash drive.
English
9

The Ethernet port sends data to local networks or LIMS systems. Use a shielded cable (e.g., STP,
FTP, S/FTP) that is 20 m (65.6 ft) or less.
4.4 Cell compartments and adapters
The instrument has two cell compartments. Refer to Figure 4. Only one sample cell can be used at a
time.
Cell compartment 1—Barcode, temperature and turbidity detection for 13-mm round sample cells
Cell compartment 2—50-mm and 1-inch rectangular sample cells and 1-inch flow sample cells, cell
adapter A and cell adapter B
Figure 4 Cell compartment
1 Cell compartment 1: round sample cells 3 Cell adapter A: 10-mm rectangular sample cells, 1-
inch round sample cells and AccuVac
®
vials
2 Cell compartment 2: 50-mm and 1-inch rectangular
sample cells and 1-inch flow sample cells
4 Cell adapter B: 30-mm vials (China only)
4.4.1 Install the cell adapter
Install the applicable cell adapter for the sample cell type as follows:
1. Open the cell compartment.
2. Install the cell adapter in cell compartment 2 so that the arrow on the cell adapter points to the
left. Refer to Figure 5.
Note: The arrow on top of the cell adapter shows the direction of the light beam path.
10
English

Figure 5 Cell adapter installation
4.4.2 Install the light shield
The light shield keeps light out of cell compartment 2.
Install the light shield in the cell compartment 2 before a measurement is started in cell compartment
1. Make sure that the arrow on the light shield points to the left. Refer to Figure 6.
The instrument is shipped with the light shield installed. Remove the light shield before cell
compartment 2 is used.
Figure 6 Light shield installation
4.5 Install the camera (optional)
1. Install the supplied camera on top of the camera holder. Tighten the screw. Refer to step 1 of
Figure 7.
2. Connect the supplied USB cable to the camera and the DR4900.
The shows in the menu bar. Refer to Figure 8 on page 13.
English
11