DR4900 用户手册_Full version.pdf - 第25页
3. Select an option. Option Description Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm) Reading unit Selects the absorbance (default), concentration or transmittance percentage. Abs (Absorbanc…

• A signal sounds.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.4 Configure the time course settings
1. Push , then select Sample analysis > Time course.
2. Push to change the settings. Refer to Table 1 on page 13.
3. Select an option.
Option Description
Reading unit Selects the absorbance or transmittance percentage: Abs (default) or %T
Scaling Selects automatic (default) or manual to adjust the Y-axis scale during the scan.
Automatic —The y-axis is automatically adjusted so that the total scan shows.
Manual —An area can be limited so that only a small section of the scan shows.
Transmittance range or
Absorbance range
Enters the transmittance or absorbance range for the y-axis: -0.300 minimum to
3.500 maximum. The function is disabled when Scaling is set to Automatic.
Default: -0.300 to 3.500 ABS
Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm)
Total time Enters the total time for the measurements (48 hours maximum). Default:
5 minutes
Reading interval Enters the time interval between measurements in seconds, minutes or hours.
Default: 12 seconds
Grid Sets the grid lines in the graph to on. Default: On
4. Push Save.
7.4.4.1 Start a time course measurement
Note: Make sure that the compartment lid is closed during the measurement.
1. Push , then select Sample analysis > Time course.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
6. Select an option.
Option Description
Add mark Records the subsequent collected data point. The data point can show an important
measurement, such as the addition of a sample or other reagent. The data point is also
shown in the table.
Stop reading Select to stop the measurements.
The measured values show on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.5 Configure the single wavelength settings
Configure the single wavelength settings as follows:
1. Push , then select Sample analysis > Single wavelength.
2. Push to change the settings.
24
English

3. Select an option.
Option Description
Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm)
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system
changes the absorbance value to a concentration value. The concentration is
calibrated in a special procedure based on absorption. The special procedure uses a
linear or polynomial calibration curve to get the best adjustment.
Concentration
factor
Enters the multiplication factor that is used to change the absorbance to
concentration values (5 digits maximum). Default: Off
Concentration unit Selects the unit for concentration or adds a new unit. Options: mg/L, g/L, ng/L, ppm,
ppb. Default: 1000 mg/L
Concentration
resolution
Selects the number of decimal digits for the concentration. Options: 1, 0.1 (default),
0.01 or 0.001
4. Push Save.
7.4.5.1 Start a single wavelength measurement
1. Push , then select Sample analysis > Single wavelength.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.6 Configure the multi wavelength settings
1. Push , then select Sample analysis > Multi wavelength.
2. Push to change the settings.
3. Select an option:
Option Description
Absorbance
formula
Selects the absorbance formula to calculate the number of wavelengths and
coefficient keys that will show on the display. Defaults: 400 nm, 500 nm, 600 nm,
700 nm and K₁A₁+K₂A₂ (K1 to K4: 10000)
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system
changes the absorbance value to a concentration value. The concentration is
calibrated in a special procedure based on absorption. The special procedure uses a
linear or polynomial calibration curve to get the best adjustment.
Concentration
factor
Enters the multiplication factor that is used to change the absorbance to concentration
values (5 digits maximum). Default: Off
English 25

Option Description
Concentration unit Selects the unit for concentration or adds a new unit. Options: mg/L, g/L, ng/L, ppm,
ppb. Default: 1000 mg/L
Concentration
resolution
Selects the number of decimal digits for the concentration. Options: 1, 0.1 (default),
0.01 or 0.001
4. Push Save.
7.4.6.1 Start a multi wavelength measurement
1. Push , then select Sample analysis > Multi wavelength.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured values show on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.7 Add a user program
User programs give the option to do customized analyses. The User programs database is used to
keep programs made by users for special requirements. Specify or select the program sequences,
absorption formulas, measurement wavelengths, factors and measurement range limits. Change
procedures. Add user procedures to the favorites list. Collect a special selection of procedures and
tests.
Note: User programs with 13-mm round cuvettes can only be completed with the empty LCW906 cuvettes.
1. Push , then select Sample analysis > User programs.
2. Select an option.
Option Description
Add Adds a new program type to the list.
Single wavelength—Select for measurements at one specified wavelength. Refer to Configure a
program with a single wavelength on page 26.
Multi wavelength—Select for measurements with a maximum of four wavelengths and the results
can be mathematically calculated to get sums, differences and relationships. Refer to Configure a
program with multi wavelengths on page 27.
Free programming—Select to customize a procedure. Refer to Configure a freely programmed test
on page 27.
Edit Edits a selected program from the list.
Delete Removes a selected program from the list of user programs. Push Delete and select the applicable
program from the list. Push Done.
Send Sends the selected program to the connected hardware. Push Send and select the applicable
program from the list. Push Done.
Cancel Interrupts the ongoing transaction.
Done Accepts the selected option.
Start Starts the selected user program.
7.4.7.1 Configure a program with a single wavelength
1. Push , then select Sample analysis > User programs > Add > Single wavelength.
2. Push OK.
26
English