Message Possible cause Solution The chemistry has expired. The chemical has expired and could cause incorrect measurements. Use new chemicals and measure again. The zero solution is not correct! The zero cuvette and the …

Figure 10 Replace the cell compartment
Section 9 Troubleshooting
Message Possible cause Solution
Critical hardware error A hardware error has occurred. Set the instrument to off, then to on again. If
the problem continues, contact technical
support.
There is too much
ambient light.
There is too much ambient light. The
instrument is in full sunlight or below a
different light source.
Decrease the ambient light. Remove the
instrument from full sunlight. Close the lid.
Lighting conditions are
unstable.
There is too much ambient light. The
instrument is in full sunlight or below a
different light source.
Make sure that the ambient light is constant
during the measurement and is not too
bright.
A test result is not
possible.
The settings are not correctly set for
the procedure. The instrument
software is outdated.
Examine the settings for the user program
or update the instrument.
The concentration is too
high.
The concentration is above the
maximum limit of the measurement
range.
Dilute the sample and measure again.
The absorbance is more
than 3.5.
The measured absorbance is more
than 3.5.
Dilute the sample and measure again.
The result is negative. The calculated result is negative. Examine the concentration of the sample.
English 31

Message Possible cause Solution
The chemistry has
expired.
The chemical has expired and could
cause incorrect measurements.
Use new chemicals and measure again.
The zero solution is not
correct!
The zero cuvette and the sample
cuvette are not compatible.
Make sure that the zero and sample
cuvettes are from the same batch.
The measured value is
over the measurement
range.
The measured value is above the set
limit.
Dilute the sample and do the test again.
Select a test with the applicable
measurement range.
The measured value is
under the measurement
range.
The measured value is below the set
limit.
Select a test with the applicable
measurement range.
The measured value is
outside of the control
limits.
The value is higher or lower than the
control limits set in Trends.
Change the control limits in Trends.
The cuvette is dirty. The cuvette is dirty or there is too
much turbidity in the sample.
Clean the cuvette. Wait until the particles
settle. Then measure again.
The barcode was not
read.
The label of the barcode cuvette was
not read.
Use a different cuvette. Make sure that the
cuvette has a HACH barcode.
Select the correct and applicable procedure
from the list.
Turbidity too high The turbidity of the liquid in the cuvette
is too high.
Clean the cuvette. If necessary, filter the
sample. Then measure again.
For COD: After the particles fall to the
bottom of the cuvette, measure again.
Cuvette temperature too
high
The cuvette temperature is too high. Do the measurement again with a
temperature-controlled cuvette, reagent and
sample.
For COD: Decrease the temperature of the
cuvette to room temperature. Then measure
again.
Cuvette temperature too
low
The cuvette temperature is too low. Do the measurement again with a
temperature-controlled cuvette, reagent and
sample.
For COD: Increase the temperature of the
cuvette to room temperature. Then measure
again.
Temperature
compensation not
possible
The temperature is not within the
correction range. For Ammonia: 25 to
38 °C (77.0 to 100.4 °F). The
measurement correction cannot be
completed.
Do the measurement again at a
temperature within the range where
compensation is possible or at room
temperature.
Section 10 Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
32
English

Table 6 Replacement parts
Description Item no.
Dust cover LPZ451.99.00001
Cell adapter A LPZ451.99.00002
Cell adapter B LPZ451.99.00003
Cell compartment, 50 mm LPZ451.99.00021
Light shield LPZ451.99.00047
DR4900 camera LPZ451.99.00004
USB cable for AP3900 procedures LPZ451.99.00045
Bench top power supply LZV844
Power cord for Europe YAA080
Power cord for Switzerland XLH051
Power cord for United Kingdom XLH057
Power cord for Unites States XLH055
Power cord for China XLH069
Power cord for South Africa LPZ440.99.00010
Power cord for Australia XLH099
Power cord for Korea LPZ440.99.00008
Power cord for Brazil LPZ440.99.00009
Table 7 Accessories
Description Item no.
SIP 10 sipper set with 1-inch round cell, Europe LQV157.99.10001
DR/check gel test standard 2763900
Verification kit LZV537
Test solution LZV810
Hand barcode scanner LZV566
USB flash drive LZV791
Ethernet cable, shielded, 2 m LZV873
Rectangular cells, 1 inch, matched pair, glass (10 mL) 2495402
Rectangular cells, 1 cm, matched pair, glass (3.5 mL) 2095100
Rectangular cells, 1 cm, quartz 2624410
Round cell, 1 inch, glass with cap (10 mL) 2122800
Round cell, 13 mm with rubber stopper LCW906
Rectangular micro-cell, 1 cm single use (1.5 mL) 2629500
Rectangular plastic cells, 1 cm 2743400
Rectangular cell, 5 cm, glass (17.5 mL) 2629250
Rectangular cell, 5 cm, quartz (17.5 mL) 2624450
English 33