DR4900 用户手册_Full version.pdf - 第21页

Option Description Single wavelength Shows the absorbance, concentration or transmittance results at a single wavelength measurement. Abs (Absorbance readings)—The light absorbed by the sample is measured in absorption u…

100%1 / 164
6. Select to add the sample ID as a favorite.
7. Push Done.
7.2.7.2 Add a sample ID to a measurement
1. Select the applicable measurement program.
2. Select the applicable sample ID from the Sample ID list.
3. Start the measurement.
The measurement result is recorded with the sample ID.
4. To remove the selected sample ID from measurements, push Deselect.
7.2.8 Set the timer
Use the timer to make sure that the individual steps of an analysis are correctly done (e.g., reaction
times, waiting times, etc.). An acoustic signal is heard when the timer has expired. The timer has no
influence on the measurement. The timer is not necessary for all procedures. Some Hach programs
use multiple timers.
1. Push to enter the minute(s) and/or seconds.
2. Enter a timer name if necessary.
3. Push OK.
4. Push to start the timer.
7.3 Accent light status
The accent light shows the status of the instrument and/or a measurement.
Table 5 Accent light status
LED status Description
Blue (constant) The static blue accent light shows when the instrument is in operational mode and when the
measurement is done.
Blue (slow on/off) The blue accent light pulses during startup, software updates and time course measurements.
Blue (fast on/off) The blue accent light blinks when the set timer is completed.
Red (constant) An internal error prevents the instrument to operate because of a system check error or a
hardware error. Refer to Troubleshooting on page 31.
7.4 Select the applicable sample analysis
To measure a sample, select the applicable sample analysis as follows:
1. Select , then select Sample analysis.
2. Select an option.
Option Description
Barcode
programs
Automatically starts a measurement when a sample cuvette with a barcode label is put in
cell compartment 1. The barcode label is scanned by the barcode reader. Refer to Use
the barcode program on page 21.
Hach programs Selects pre-programmed procedures that are used with Hach chemicals and Hach
pipette tests. Refer to Select a Hach program on page 22.
Wavelength
scan
Measures a sample at a set wavelength range to get a spectrum graph. Refer to
Configure the wavelength scan settings on page 23.
Time course Records the absorption or percentage transmittance at a wavelength in a specified period
of time. The data is shown in a table or as a graph. Refer to Configure the time course
settings on page 24.
20 English
Option Description
Single
wavelength
Shows the absorbance, concentration or transmittance results at a single wavelength
measurement.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes
the absorbance value to a concentration value. The concentration is calibrated in a
special procedure based on absorption. The special procedure uses a linear or
polynomial calibration curve to get the best adjustment.
Refer to Configure the single wavelength settings on page 24.
Multi wavelength Shows the absorbance, concentration or transmittance results at a maximum of four
wavelength measurements. The absorption differences and absorption relationships are
calculated. Refer to Configure the multi wavelength settings on page 25.
User programs Lets users program their own procedures or adjust pre-programmed procedures to other
requirements. Refer to Add a user program on page 26.
7.4.1 Use the barcode program
A barcode reader in cell compartment 1 automatically reads the barcode on the 13-mm sample
cuvettes when the sample cuvette has completed one turn. The instrument uses the barcode
identification to automatically set the correct wavelength for the analysis and immediately calculates
the result with pre-set factors.
In addition, measured values are recorded at 10 different positions during one turn. A program is
started to remove outliers and the average of the measured values is calculated. All of the effects on
the measurement that occur in the cell compartment are identified to make sure that the result is
accurate.
1. Put the light shield adapter in cell compartment 2.
2. Prepare the barcode test with the procedure specifications.
3. Put the sample cuvette of the selected barcode procedure into cell compartment 1.
The measurement starts automatically and the result value shows.
4. Push to change the settings.
Option Description
Chemical form Selects the chemical form and the applicable measurement range.
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes
the absorbance value to a concentration value. The concentration is calibrated in a
special procedure based on absorption. The special procedure uses a linear or
polynomial calibration curve to get the best adjustment.
Dilution Set to on to enter a dilution factor (default: off). The number entered under Dilution is
multiplied by the result to compensate for the adjustment. For example, if the sample
was diluted by a factor of 2, enter 2. Default: Off
Note: When Dilution is set to on, the dilution icon will show on the display.
Multiple
determination
Enters a maximum of 10 measurements. Default: Off
English 21
Option Description
Reagent blank Set to on to add the reagent blank value from the measured values. Set to on to
subtract the reagent blank value from the measured values. The reagent blank sample
is used as an offset during calibration. The shape of the calibration curve stays the
same. Default: Off
Vial temperature
detection
Set to on to measure the temperature of the vial. A warning will show if the vial is too
hot or too cold for correct measurements. Default: On. Refer to Table 4 on page 15.
Temperature
compensation
Set to on to adjust for vial temperature effects on the measurement. Default: On. Refer
to Table 4 on page 15.
Turbidity detection Set to on to measure the turbidity. A warning will show if the turbidity is too high for
correct measurements. Default: On. Refer to Table 4 on page 15.
5. Push Save.
7.4.2 Select a Hach program
More than 200 pre-programmed Hach procedures are in Hach programs. All of the important
parameters and timers are included in the procedures.
1. Push , then select Sample analysis > Hach programs.
2. Select an option.
Option Description
Search Finds the procedures by name and program number.
Sort Sorts the list by program number (#), Parameter, Method# and favorites.
Star icon Adds a Hach program to a favorites list.
3. Select the applicable Hach program.
4. Push Start.
The procedure name shows on the left side of the display.
5. Push to change the settings.
Option Description
Chemical
form
Selects the chemical form and the applicable measurement range.
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in absorption
units.
%T (Transmittance reading)—The percentage of light that passes through the sample and is
measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes the
absorbance value to a concentration value. The concentration is calibrated in a special
procedure based on absorption. The special procedure uses a linear or polynomial
calibration curve to get the best adjustment.
Dilution Set to on to enter a dilution factor (default: off). The number entered under Dilution is
multiplied by the result to compensate for the adjustment. For example, if the sample was
diluted by a factor of 2, enter 2. Default: Off
Note: When Dilution is set to on, the dilution icon will show on the display.
Standard
adjust
The instructions for the procedure parameters specify if a standard adjustment is necessary
and how to continue.
Reagent
blank
Set to on to add the reagent blank value from the measured values. Set to on to subtract the
reagent blank value from the measured values. The reagent blank sample is used as an
offset during calibration. The shape of the calibration curve stays the same. Default: Off
6. Push Save.
22
English