DR4900 用户手册_Full version.pdf - 第30页

C A U T I O N Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS…

100%1 / 164
Figure 9 Read sample ID label
7.6 Show reports and data
1. Push , then select Reports & data.
2. Select an option.
Option Description
Data log Keeps a data log with a maximum of 10,000 measurements. A complete report of the analysis is
kept, which includes the date, time, parameter, value, sample ID and operator.
Trends Use the trend option to show a time curve or history curve for the saved measurements for each
parameter and location. The concentration of the applicable parameter shows over time.
Ratios Monitors the parameter relationships at a special location and shows the parameter in a chart with
the ratio function.
Scan log Records 200 time-course measurements and multi wavelength scans.
AQA log Examines the analytical quality. Configures the control of individual procedures with standard
solutions or the testing of cross-sensitivity of analyses with standard addition solutions, multiple
determinations and dilutions. A maximum of 1000 records are recorded.
7.7 Do a system check
The System checks menu contains some optical checks for the system.
7.7.1 Do an optical check
At regular intervals, do an optical verification to look at stray light, photometric and wavelength
precision.
An optional test filter set (verification kit) with six precision glass filters, target values and instructions
is available as an aid to do a comprehensive instrument verification. Refer to Table 7 on page 33.
Section 8 Maintenance
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
English 29
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
8.1 Clean the instrument
N O T I C E
Do not use cleaning agents such as turpentine, acetone or similar products to clean the instrument including the
display and accessories.
Clean the enclosure, the cell compartment and all of the accessories with a moist cloth and a mild
soap solution.
8.2 Clean spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
1. Obey all facility safety protocols for spill control.
2. Discard the waste according to applicable regulations.
8.3 Replace the cell compartment
Replace the cell compartment if the cell compartment is very dirty and cannot be cleaned with a mild
soap solution.
1. Set the power switch to off. Refer to Startup on page 9.
2. Replace the cell compartment. Make sure that the surface of the cell compartment is aligned flush
with the surface of the enclosure. Refer to Figure 10.
30
English
Figure 10 Replace the cell compartment
Section 9 Troubleshooting
Message Possible cause Solution
Critical hardware error A hardware error has occurred. Set the instrument to off, then to on again. If
the problem continues, contact technical
support.
There is too much
ambient light.
There is too much ambient light. The
instrument is in full sunlight or below a
different light source.
Decrease the ambient light. Remove the
instrument from full sunlight. Close the lid.
Lighting conditions are
unstable.
There is too much ambient light. The
instrument is in full sunlight or below a
different light source.
Make sure that the ambient light is constant
during the measurement and is not too
bright.
A test result is not
possible.
The settings are not correctly set for
the procedure. The instrument
software is outdated.
Examine the settings for the user program
or update the instrument.
The concentration is too
high.
The concentration is above the
maximum limit of the measurement
range.
Dilute the sample and measure again.
The absorbance is more
than 3.5.
The measured absorbance is more
than 3.5.
Dilute the sample and measure again.
The result is negative. The calculated result is negative. Examine the concentration of the sample.
English 31