7.2.1 Configure the measurement settings 1. Push , then select Settings > Measurement settings . 2. Select an option. Option Description Data export Selects the settings for data storage, data management, scan log, fi…

Section 7 Operation
7.1 New program functions
Refer to Table 4 for an overview of new program functions.
Table 4 DR4900 functions
New function Description
Temperature
detection
The temperature of the vial is measured with an infrared sensor. The vial temperature
shows on the screen.
Temperature warning The vial temperature is examined based on the chemical and physical properties for the
specified procedure and measurement range. If the vial temperature has a sufficiently
large effect on the measured value, a temperature warning shows.
The temperature warning is only enabled for selected procedures.
Temperature
compensation
The vial temperature is used to correct the measured values when the vial temperature
has a sufficiently large effect on the measured values. The correction is dependant on the
procedure, measured value and temperature. The temperature compensation is
applicable to a procedure-specified temperature range.
The temperature compensation is only enabled for selected procedures.
Turbidity warning The turbidity in the vial is examined by nephelometric measurement based on the
procedure and measurement range. If the turbidity has a sufficiently large effect on the
measured values, a turbidity warning shows.
The turbidity warning is only enabled for selected procedures.
7.2 Configure the instrument
1. Push , then select Settings.
2. Select an option.
Option Description
Measurement
settings
Selects the settings for data export, temperature detection, turbidity detection,
TrueCal, temperature unit, interference check, certificate of analysis and sipper. Refer
to Configure the measurement settings on page 16.
Operator ID Adds an operator ID to the system. Refer to Add an operator ID on page 16.
Security Sets a password for administrator access. Security management can be divided into
three different user groups: basic, intermediate and advanced.
Note: A basic user must not have more permissions than an intermediate user. An
intermediate user must not have more permissions than an advanced user.
Refer to Configure security management on page 17.
General Sets the language, date and time, display and power, and sound settings. Refer to
Configure the general settings on page 17.
Connectivity Connects the instrument to external devices. Refer to Configure external devices
on page 18.
Service Gives information about updates, service logs, backups, the next service and Hach
service menu. Refer to Service menu on page 18.
English 15

7.2.1 Configure the measurement settings
1. Push , then select Settings > Measurement settings.
2. Select an option.
Option Description
Data export Selects the settings for data storage, data management, scan log, file types, send data
and configure file export.
Auto save—Set to on to automatically keep all of the measured data in the data log.
Default: On
Send data as—Selects CSV (default) or XML format for the data exports.
Auto send—Set to on to automatically send all measured data to all of the connected
devices. Default: Off
Auto send as—Selects a New file (default) or Continue file.
New file—Makes a new table file for each measurement.
Continue file—Adds continuously all of the measured values to one table file.
Temperature
detection
Global vial temperature detection —Set to on to show the vial temperature for all
barcode programs. Default: On
Temperature warning: Cold —Set to on to show a warning when the vial is too cold for
correct measurements. Default: On
Temperature warning: Hot —Set to on to show a warning when the vial is too hot for
correct measurements. Default: On
Global temperature compensation —Set to on to adjust the measured values based on
the vial temperature. Default: On
Turbidity
detection
Global turbidity detection —Set to on to show a warning when the turbidity is too high
for correct measurements for all of the barcode programs that have the function enabled.
Default: On
High turbidity warning —Set to on to show a warning when the turbidity is too high for
correct measurements. Default: On
Truecal Truecal —Set to on to decrease the variation in results caused by batch-to-batch
variations in chemical raw materials. The barcode contains the calibration data for each
reagent batch and automatically updates the calibration curve. Default: On
Temperature
unit
Temperature unit —Selects the temperature unit shown. Options: Celsius °C (default) or
Fahrenheit °F
Interference
check
Interference check —Set to on to automatically search for interference ions. Typical
interferences, such as chloride for nitrate, are recognized from calculated concentrations
and a warning is shown. The interference check is only enabled for barcode tests that are
shown with the same sample IDs. Default: On
Certificate of
analysis
Certificate of analysis —Set to on to show if a new certificate of analysis is available.
Default: On
Sipper Sipper —Set to on when a sipper module is connected. Default: On
Refer to the SIP 10 user manual for more information.
3. Push Save.
7.2.2 Add an operator ID
1. Push , then select Settings > Operator ID.
2. Select an option.
Option Description
Add Adds a new operator ID. The operator ID can have a maximum length of 30 characters.
Edit Edits an operator ID from the list.
Delete Removes an operator ID from the list.
16 English

3. Push Add to add a new operator ID.
4. Select an option.
Option Description
Color Selects a color to easily distinguish between the different operators.
Name Enters a unique name or ID.
Password Sets a password for the new operator.
Confirm password Accepts the set password for the operator.
Security level Selects the security level: Basic, Intermediate or Advanced. The change must be done
with the administrator password.
5. Push Save.
7.2.2.1 Use the operator ID to log in
Log in to the instrument to add the operator ID to the measurements.
1. Push in the menu bar.
2. Select the applicable Operator ID.
3. Enter the password.
4. Push Log in.
7.2.3 Configure security management
Do the instructions that follow to configure different access permissions.
1. Push , then select Settings > Security.
2. Push Security setup.
3. Set Security to on.
4. Set a password for administrator rights.
Note: The password must contain a minimum of 10 characters with a combination of numbers, symbols,
uppercase and lowercase letters.
5. Enter the password again.
A pass phrase shows. Record the pass phrase to change the password if the password is
forgotten.
6. Push Save.
A default security list for Basic, Intermediate, Advanced and Admin rights shows.
7. Enter the admin password again, then select an option.
Option Description
Security setup Changes the administrator password.
Restore defaults Sets all of the security rights to the factory defaults.
Edit operator rights Edits the operator rights. Select or unselect different functions.
8. Push Done.
9. Push to go back to the home screen.
7.2.4 Configure the general settings
1. Push , then select Settings > General.
2. Select an option.
Option Description
Language Language —Selects the applicable language from the list. Default: English
English 17