DR4900 用户手册_Full version.pdf - 第24页
• A signal sounds. The measured value shows on the display. The measured values are automatically saved to the data log if Auto save is enabled. Refer to Configure the instrument on page 15. 7.4.4 Configure the time cour…

7.4.2.1 Start a Hach program
1. Push , then select Sample analysis > Hach programs.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.3 Configure the wavelength scan settings
1. Push , then select Sample analysis > Wavelength scan.
2. Push to change the settings.
3. Select an option.
Option Description
Integral Set to on to show the integral for the scan. The integral gives the area and the
derivative of the integral gives the original function. Default: Off
Wavelength range Enters the wavelength spectrum between 320 to 1100 nm. Default: 340 to 900 nm
Step Enters the wavelength interval for the scan: 1 nm (default), 2 nm or 5 nm
Reference scan Selects a saved scan to show with new scans as a reference. Default: Off
Reading unit Selects the absorbance or transmittance percentage: Abs (default) or %T
Scaling Selects automatic (default) or manual to adjust the Y-axis scale during the scan.
Automatic —The y-axis is automatically adjusted so that the total scan shows.
Manual —An area can be limited so that only a small section of the scan shows.
Transmittance range
or Absorbance range
Enters the transmittance or absorbance range for the y-axis: -0.300 minimum to
3.500 maximum. The function is disabled when Scaling is set to Automatic.
Default: -0.300 to 3.500 ABS
Grid Sets the grid lines in the graph to on. Default: On
4. Push Save.
7.4.3.1 Start a wavelength scan measurement
Note: Make sure that the compartment lid is closed during the measurement.
If the reference scan is selected, the reference scan will show in a different color than the scan
measured. Refer to Configure the wavelength scan settings on page 23.
1. Push , then select Sample analysis > Wavelength scan.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The graph of the absorption or transmission values for the scanned wavelengths is continuously
shown.
6. The Wavelength scan is complete, if:
• The graph shows full-size.
• The scaling of the x-axis occurs automatically.
• The cursor functions in the vertical navigation bar are highlighted.
English
23

• A signal sounds.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.4 Configure the time course settings
1. Push , then select Sample analysis > Time course.
2. Push to change the settings. Refer to Table 1 on page 13.
3. Select an option.
Option Description
Reading unit Selects the absorbance or transmittance percentage: Abs (default) or %T
Scaling Selects automatic (default) or manual to adjust the Y-axis scale during the scan.
Automatic —The y-axis is automatically adjusted so that the total scan shows.
Manual —An area can be limited so that only a small section of the scan shows.
Transmittance range or
Absorbance range
Enters the transmittance or absorbance range for the y-axis: -0.300 minimum to
3.500 maximum. The function is disabled when Scaling is set to Automatic.
Default: -0.300 to 3.500 ABS
Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm)
Total time Enters the total time for the measurements (48 hours maximum). Default:
5 minutes
Reading interval Enters the time interval between measurements in seconds, minutes or hours.
Default: 12 seconds
Grid Sets the grid lines in the graph to on. Default: On
4. Push Save.
7.4.4.1 Start a time course measurement
Note: Make sure that the compartment lid is closed during the measurement.
1. Push , then select Sample analysis > Time course.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
6. Select an option.
Option Description
Add mark Records the subsequent collected data point. The data point can show an important
measurement, such as the addition of a sample or other reagent. The data point is also
shown in the table.
Stop reading Select to stop the measurements.
The measured values show on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.5 Configure the single wavelength settings
Configure the single wavelength settings as follows:
1. Push , then select Sample analysis > Single wavelength.
2. Push to change the settings.
24
English

3. Select an option.
Option Description
Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm)
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system
changes the absorbance value to a concentration value. The concentration is
calibrated in a special procedure based on absorption. The special procedure uses a
linear or polynomial calibration curve to get the best adjustment.
Concentration
factor
Enters the multiplication factor that is used to change the absorbance to
concentration values (5 digits maximum). Default: Off
Concentration unit Selects the unit for concentration or adds a new unit. Options: mg/L, g/L, ng/L, ppm,
ppb. Default: 1000 mg/L
Concentration
resolution
Selects the number of decimal digits for the concentration. Options: 1, 0.1 (default),
0.01 or 0.001
4. Push Save.
7.4.5.1 Start a single wavelength measurement
1. Push , then select Sample analysis > Single wavelength.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.6 Configure the multi wavelength settings
1. Push , then select Sample analysis > Multi wavelength.
2. Push to change the settings.
3. Select an option:
Option Description
Absorbance
formula
Selects the absorbance formula to calculate the number of wavelengths and
coefficient keys that will show on the display. Defaults: 400 nm, 500 nm, 600 nm,
700 nm and K₁A₁+K₂A₂ (K1 to K4: 10000)
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system
changes the absorbance value to a concentration value. The concentration is
calibrated in a special procedure based on absorption. The special procedure uses a
linear or polynomial calibration curve to get the best adjustment.
Concentration
factor
Enters the multiplication factor that is used to change the absorbance to concentration
values (5 digits maximum). Default: Off
English 25