6. Select to add the sample ID as a favorite. 7. Push Done . 7.2.7.2 Add a sample ID to a measurement 1. Select the applicable measurement program. 2. Select the applicable sample ID from the Sample ID list. 3. Start the…

Option Description
Backup & restore Backup & restore —Sends all of the data (e.g., sample-IDs, operatior-IDs and
settings) to a connected instrument.
Next service Next service —Shows the service due date.
Hach service menu Hach service menu —Shows the factory service menu. A password is necessary.
3. Push OK.
7.2.6.1 Update the instrument software
Install software updates regularly for best results.
1. Download the software from the manufacturer website to a USB flash drive.
2. Install the USB-flash drive in one of the USB ports of the instrument. Refer to Figure 1 on page 7.
3. Push , then select Settings > Service > Update.
A pop-up window shows the current and new software version.
4. Push OK to start the update process.
Wait until the software update is complete.
5. When the software was successfully updated, push OK.
The instrument will start again.
6. Remove the USB flash drive from the USB port.
7.2.7 Use a sample ID
Sample IDs are used to specify the sample location or to keep other special information about the
sample. The sample ID can have a maximum length of 26 characters.
To add a sample ID manually,
• scan the bottle label to import the sample ID information
3
or
• import a CSV table with a list of sample IDs.
1. Go to the measurement screen, select Sample ID list > .
2. Select an option.
Option Description
Customize list Adds the Sample ID name, Date of sampling, Time of sampling, Collected by, Comment
and identifies the sample ID as a favorite if necessary. Refer to Customize a sample ID
on page 19.
Import data Imports a sample ID list with a maximum of four columns in CSV format. Adds
information about the Sequence, Sample ID, Date time and Collected by. Add the file to
a folder called "SampleID". Save the folder to the root directory of the USB flash drive or
in the network shared folder if a device is connected.
Read bottle tag
(optional)
Scans the bottle tag to import the sample ID information if a camera is connected. Refer
to Use the camera (optional) on page 28.
7.2.7.1 Customize a sample ID
1. Go to the measurement screen, select Sample ID list > > Customize list.
2. Push Add to make a new sample ID.
3. Select a color identifier, if necessary.
4. Enter the sample ID name, the date and time, and the name of the operator.
5. Push to add a comment to the sample ID.
3
Only available for instruments with a connected camera
English 19

6. Select to add the sample ID as a favorite.
7. Push Done.
7.2.7.2 Add a sample ID to a measurement
1. Select the applicable measurement program.
2. Select the applicable sample ID from the Sample ID list.
3. Start the measurement.
The measurement result is recorded with the sample ID.
4. To remove the selected sample ID from measurements, push Deselect.
7.2.8 Set the timer
Use the timer to make sure that the individual steps of an analysis are correctly done (e.g., reaction
times, waiting times, etc.). An acoustic signal is heard when the timer has expired. The timer has no
influence on the measurement. The timer is not necessary for all procedures. Some Hach programs
use multiple timers.
1. Push to enter the minute(s) and/or seconds.
2. Enter a timer name if necessary.
3. Push OK.
4. Push to start the timer.
7.3 Accent light status
The accent light shows the status of the instrument and/or a measurement.
Table 5 Accent light status
LED status Description
Blue (constant) The static blue accent light shows when the instrument is in operational mode and when the
measurement is done.
Blue (slow on/off) The blue accent light pulses during startup, software updates and time course measurements.
Blue (fast on/off) The blue accent light blinks when the set timer is completed.
Red (constant) An internal error prevents the instrument to operate because of a system check error or a
hardware error. Refer to Troubleshooting on page 31.
7.4 Select the applicable sample analysis
To measure a sample, select the applicable sample analysis as follows:
1. Select , then select Sample analysis.
2. Select an option.
Option Description
Barcode
programs
Automatically starts a measurement when a sample cuvette with a barcode label is put in
cell compartment 1. The barcode label is scanned by the barcode reader. Refer to Use
the barcode program on page 21.
Hach programs Selects pre-programmed procedures that are used with Hach chemicals and Hach
pipette tests. Refer to Select a Hach program on page 22.
Wavelength
scan
Measures a sample at a set wavelength range to get a spectrum graph. Refer to
Configure the wavelength scan settings on page 23.
Time course Records the absorption or percentage transmittance at a wavelength in a specified period
of time. The data is shown in a table or as a graph. Refer to Configure the time course
settings on page 24.
20 English

Option Description
Single
wavelength
Shows the absorbance, concentration or transmittance results at a single wavelength
measurement.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes
the absorbance value to a concentration value. The concentration is calibrated in a
special procedure based on absorption. The special procedure uses a linear or
polynomial calibration curve to get the best adjustment.
Refer to Configure the single wavelength settings on page 24.
Multi wavelength Shows the absorbance, concentration or transmittance results at a maximum of four
wavelength measurements. The absorption differences and absorption relationships are
calculated. Refer to Configure the multi wavelength settings on page 25.
User programs Lets users program their own procedures or adjust pre-programmed procedures to other
requirements. Refer to Add a user program on page 26.
7.4.1 Use the barcode program
A barcode reader in cell compartment 1 automatically reads the barcode on the 13-mm sample
cuvettes when the sample cuvette has completed one turn. The instrument uses the barcode
identification to automatically set the correct wavelength for the analysis and immediately calculates
the result with pre-set factors.
In addition, measured values are recorded at 10 different positions during one turn. A program is
started to remove outliers and the average of the measured values is calculated. All of the effects on
the measurement that occur in the cell compartment are identified to make sure that the result is
accurate.
1. Put the light shield adapter in cell compartment 2.
2. Prepare the barcode test with the procedure specifications.
3. Put the sample cuvette of the selected barcode procedure into cell compartment 1.
The measurement starts automatically and the result value shows.
4. Push to change the settings.
Option Description
Chemical form Selects the chemical form and the applicable measurement range.
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in
absorption units.
%T (Transmittance reading)—The percentage of light that passes through the sample
and is measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes
the absorbance value to a concentration value. The concentration is calibrated in a
special procedure based on absorption. The special procedure uses a linear or
polynomial calibration curve to get the best adjustment.
Dilution Set to on to enter a dilution factor (default: off). The number entered under Dilution is
multiplied by the result to compensate for the adjustment. For example, if the sample
was diluted by a factor of 2, enter 2. Default: Off
Note: When Dilution is set to on, the dilution icon will show on the display.
Multiple
determination
Enters a maximum of 10 measurements. Default: Off
English 21