7.2.1 Configure the measurement settings
1. Push , then select Settings > Measurement settings.
2. Select an option.
Option Description
Data export Selects the settings for data storage, data management, scan log, file types, send data
and configure file export.
Auto save—Set to on to automatically keep all of the measured data in the data log.
Default: On
Send data as—Selects CSV (default) or XML format for the data exports.
Auto send—Set to on to automatically send all measured data to all of the connected
devices. Default: Off
Auto send as—Selects a New file (default) or Continue file.
New file—Makes a new table file for each measurement.
Continue file—Adds continuously all of the measured values to one table file.
Temperature
detection
Global vial temperature detection —Set to on to show the vial temperature for all
barcode programs. Default: On
Temperature warning: Cold —Set to on to show a warning when the vial is too cold for
correct measurements. Default: On
Temperature warning: Hot —Set to on to show a warning when the vial is too hot for
correct measurements. Default: On
Global temperature compensation —Set to on to adjust the measured values based on
the vial temperature. Default: On
Turbidity
detection
Global turbidity detection —Set to on to show a warning when the turbidity is too high
for correct measurements for all of the barcode programs that have the function enabled.
Default: On
High turbidity warning —Set to on to show a warning when the turbidity is too high for
correct measurements. Default: On
Truecal Truecal —Set to on to decrease the variation in results caused by batch-to-batch
variations in chemical raw materials. The barcode contains the calibration data for each
reagent batch and automatically updates the calibration curve. Default: On
Temperature
unit
Temperature unit —Selects the temperature unit shown. Options: Celsius °C (default) or
Fahrenheit °F
Interference
check
Interference check —Set to on to automatically search for interference ions. Typical
interferences, such as chloride for nitrate, are recognized from calculated concentrations
and a warning is shown. The interference check is only enabled for barcode tests that are
shown with the same sample IDs. Default: On
Certificate of
analysis
Certificate of analysis —Set to on to show if a new certificate of analysis is available.
Default: On
Sipper Sipper —Set to on when a sipper module is connected. Default: On
Refer to the SIP 10 user manual for more information.
3. Push Save.
7.2.2 Add an operator ID
1. Push , then select Settings > Operator ID.
2. Select an option.
Option Description
Add Adds a new operator ID. The operator ID can have a maximum length of 30 characters.
Edit Edits an operator ID from the list.
Delete Removes an operator ID from the list.
16 English
3. Push Add to add a new operator ID.
4. Select an option.
Option Description
Color Selects a color to easily distinguish between the different operators.
Name Enters a unique name or ID.
Password Sets a password for the new operator.
Confirm password Accepts the set password for the operator.
Security level Selects the security level: Basic, Intermediate or Advanced. The change must be done
with the administrator password.
5. Push Save.
7.2.2.1 Use the operator ID to log in
Log in to the instrument to add the operator ID to the measurements.
1. Push in the menu bar.
2. Select the applicable Operator ID.
3. Enter the password.
4. Push Log in.
7.2.3 Configure security management
Do the instructions that follow to configure different access permissions.
1. Push , then select Settings > Security.
2. Push Security setup.
3. Set Security to on.
4. Set a password for administrator rights.
Note: The password must contain a minimum of 10 characters with a combination of numbers, symbols,
uppercase and lowercase letters.
5. Enter the password again.
A pass phrase shows. Record the pass phrase to change the password if the password is
forgotten.
6. Push Save.
A default security list for Basic, Intermediate, Advanced and Admin rights shows.
7. Enter the admin password again, then select an option.
Option Description
Security setup Changes the administrator password.
Restore defaults Sets all of the security rights to the factory defaults.
Edit operator rights Edits the operator rights. Select or unselect different functions.
8. Push Done.
9. Push to go back to the home screen.
7.2.4 Configure the general settings
1. Push , then select Settings > General.
2. Select an option.
Option Description
Language Language —Selects the applicable language from the list. Default: English
English 17
Option Description
Date & time Date & time —Sets the Date format, Date and Time format. Defaults: dd-mm-yyyy, date
set at factory and 24 hours. The Automatic time setting is only enabled when the
instrument is connected to Ethernet or Wi-Fi. Default: Off. Then, the Time zone and the
Time are automatically set. Default: Berlin, time set at factory
Display &
power
Display & power —Sets the Display brightness and the Standby timer.
Display brightness —Moves the slider to adjust the brightness of the display. Default: 90
Standby timer —Selects the time after which the instrument is set to standby: 30 minutes,
1 hour (default), 2 hours or 12 hours. When the device is in standby mode, touch the display
to wake up the instrument. Default: On
Sound Sound —Enables or disables the sound for All sounds. Default: On
Moves the slider to enable, disable or adjust the sound volume for the Touch screen
(default: 50%), Reading done (default: 50%), Timer (default: 50%), Startup (default: 30%),
Warning (default: 20%), Camera detection done (default: 20%) and Standby (default:
30%).
3. Push Save.
7.2.4.1 Set the instrument to standby mode
The standby mode decreases power consumption and keeps important functions and the latest
condition in memory so that the instrument can quickly start to operate again.
1. Push , then select Settings > General > Display & power.
2. Set the Standby timer to on.
3. Select the applicable standby time.
4. Push Save.
7.2.5 Configure external devices
1. Push , then select Settings > Connectivity.
2. Select an option.
Option Description
Ethernet Ethernet —Configures the Ethernet parameters to connect to a network or send data to
a PC. Connect an Ethernet cable to configure the network.
Wi-Fi Wi-Fi —Selects a Wi-Fi network for data transmission. Connect a Wi-Fi dongle to
configure the Wi-Fi network.
Network shared
folder
Network shared folder —Selects the destination folder for data transmission. Default:
Off
Printer Printer —Configures the network or USB printer settings for data transmission. Default:
Off
3. Push OK.
7.2.6 Service menu
1. Push , then select Settings > Service.
2. Select an option.
Option Description
Update Update —Updates the software of the instrument to the latest version. Refer to Update
the instrument software on page 19.
Service logs Service logs —Downloads the service logs for technical support.
18 English