Option Description Backup & restore Backup & restore —Sends all of the data (e.g., sample-IDs, operatior-IDs and settings) to a connected instrument. Next service Next service —Shows the service due date. Hach se…

100%1 / 164
Option Description
Date & time Date & time —Sets the Date format, Date and Time format. Defaults: dd-mm-yyyy, date
set at factory and 24 hours. The Automatic time setting is only enabled when the
instrument is connected to Ethernet or Wi-Fi. Default: Off. Then, the Time zone and the
Time are automatically set. Default: Berlin, time set at factory
Display &
power
Display & power —Sets the Display brightness and the Standby timer.
Display brightness —Moves the slider to adjust the brightness of the display. Default: 90
Standby timer —Selects the time after which the instrument is set to standby: 30 minutes,
1 hour (default), 2 hours or 12 hours. When the device is in standby mode, touch the display
to wake up the instrument. Default: On
Sound Sound —Enables or disables the sound for All sounds. Default: On
Moves the slider to enable, disable or adjust the sound volume for the Touch screen
(default: 50%), Reading done (default: 50%), Timer (default: 50%), Startup (default: 30%),
Warning (default: 20%), Camera detection done (default: 20%) and Standby (default:
30%).
3. Push Save.
7.2.4.1 Set the instrument to standby mode
The standby mode decreases power consumption and keeps important functions and the latest
condition in memory so that the instrument can quickly start to operate again.
1. Push , then select Settings > General > Display & power.
2. Set the Standby timer to on.
3. Select the applicable standby time.
4. Push Save.
7.2.5 Configure external devices
1. Push , then select Settings > Connectivity.
2. Select an option.
Option Description
Ethernet Ethernet —Configures the Ethernet parameters to connect to a network or send data to
a PC. Connect an Ethernet cable to configure the network.
Wi-Fi Wi-Fi —Selects a Wi-Fi network for data transmission. Connect a Wi-Fi dongle to
configure the Wi-Fi network.
Network shared
folder
Network shared folder —Selects the destination folder for data transmission. Default:
Off
Printer Printer —Configures the network or USB printer settings for data transmission. Default:
Off
3. Push OK.
7.2.6 Service menu
1. Push , then select Settings > Service.
2. Select an option.
Option Description
Update Update —Updates the software of the instrument to the latest version. Refer to Update
the instrument software on page 19.
Service logs Service logs —Downloads the service logs for technical support.
18 English
Option Description
Backup & restore Backup & restore —Sends all of the data (e.g., sample-IDs, operatior-IDs and
settings) to a connected instrument.
Next service Next service —Shows the service due date.
Hach service menu Hach service menu —Shows the factory service menu. A password is necessary.
3. Push OK.
7.2.6.1 Update the instrument software
Install software updates regularly for best results.
1. Download the software from the manufacturer website to a USB flash drive.
2. Install the USB-flash drive in one of the USB ports of the instrument. Refer to Figure 1 on page 7.
3. Push , then select Settings > Service > Update.
A pop-up window shows the current and new software version.
4. Push OK to start the update process.
Wait until the software update is complete.
5. When the software was successfully updated, push OK.
The instrument will start again.
6. Remove the USB flash drive from the USB port.
7.2.7 Use a sample ID
Sample IDs are used to specify the sample location or to keep other special information about the
sample. The sample ID can have a maximum length of 26 characters.
To add a sample ID manually,
• scan the bottle label to import the sample ID information
3
or
• import a CSV table with a list of sample IDs.
1. Go to the measurement screen, select Sample ID list > .
2. Select an option.
Option Description
Customize list Adds the Sample ID name, Date of sampling, Time of sampling, Collected by, Comment
and identifies the sample ID as a favorite if necessary. Refer to Customize a sample ID
on page 19.
Import data Imports a sample ID list with a maximum of four columns in CSV format. Adds
information about the Sequence, Sample ID, Date time and Collected by. Add the file to
a folder called "SampleID". Save the folder to the root directory of the USB flash drive or
in the network shared folder if a device is connected.
Read bottle tag
(optional)
Scans the bottle tag to import the sample ID information if a camera is connected. Refer
to Use the camera (optional) on page 28.
7.2.7.1 Customize a sample ID
1. Go to the measurement screen, select Sample ID list > > Customize list.
2. Push Add to make a new sample ID.
3. Select a color identifier, if necessary.
4. Enter the sample ID name, the date and time, and the name of the operator.
5. Push to add a comment to the sample ID.
3
Only available for instruments with a connected camera
English 19
6. Select to add the sample ID as a favorite.
7. Push Done.
7.2.7.2 Add a sample ID to a measurement
1. Select the applicable measurement program.
2. Select the applicable sample ID from the Sample ID list.
3. Start the measurement.
The measurement result is recorded with the sample ID.
4. To remove the selected sample ID from measurements, push Deselect.
7.2.8 Set the timer
Use the timer to make sure that the individual steps of an analysis are correctly done (e.g., reaction
times, waiting times, etc.). An acoustic signal is heard when the timer has expired. The timer has no
influence on the measurement. The timer is not necessary for all procedures. Some Hach programs
use multiple timers.
1. Push to enter the minute(s) and/or seconds.
2. Enter a timer name if necessary.
3. Push OK.
4. Push to start the timer.
7.3 Accent light status
The accent light shows the status of the instrument and/or a measurement.
Table 5 Accent light status
LED status Description
Blue (constant) The static blue accent light shows when the instrument is in operational mode and when the
measurement is done.
Blue (slow on/off) The blue accent light pulses during startup, software updates and time course measurements.
Blue (fast on/off) The blue accent light blinks when the set timer is completed.
Red (constant) An internal error prevents the instrument to operate because of a system check error or a
hardware error. Refer to Troubleshooting on page 31.
7.4 Select the applicable sample analysis
To measure a sample, select the applicable sample analysis as follows:
1. Select , then select Sample analysis.
2. Select an option.
Option Description
Barcode
programs
Automatically starts a measurement when a sample cuvette with a barcode label is put in
cell compartment 1. The barcode label is scanned by the barcode reader. Refer to Use
the barcode program on page 21.
Hach programs Selects pre-programmed procedures that are used with Hach chemicals and Hach
pipette tests. Refer to Select a Hach program on page 22.
Wavelength
scan
Measures a sample at a set wavelength range to get a spectrum graph. Refer to
Configure the wavelength scan settings on page 23.
Time course Records the absorption or percentage transmittance at a wavelength in a specified period
of time. The data is shown in a table or as a graph. Refer to Configure the time course
settings on page 24.