3. Select an option. Option Description Name Enters a name for the new single wavelength program. Units Selects the unit for concentration or adds a new unit: mg/L (default), g/L, ng/L, ppm, ppb Wavelength Enters the wav…

100%1 / 164
Option Description
Concentration unit Selects the unit for concentration or adds a new unit. Options: mg/L, g/L, ng/L, ppm,
ppb. Default: 1000 mg/L
Concentration
resolution
Selects the number of decimal digits for the concentration. Options: 1, 0.1 (default),
0.01 or 0.001
4. Push Save.
7.4.6.1 Start a multi wavelength measurement
1. Push , then select Sample analysis > Multi wavelength.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured values show on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.7 Add a user program
User programs give the option to do customized analyses. The User programs database is used to
keep programs made by users for special requirements. Specify or select the program sequences,
absorption formulas, measurement wavelengths, factors and measurement range limits. Change
procedures. Add user procedures to the favorites list. Collect a special selection of procedures and
tests.
Note: User programs with 13-mm round cuvettes can only be completed with the empty LCW906 cuvettes.
1. Push , then select Sample analysis > User programs.
2. Select an option.
Option Description
Add Adds a new program type to the list.
Single wavelength—Select for measurements at one specified wavelength. Refer to Configure a
program with a single wavelength on page 26.
Multi wavelength—Select for measurements with a maximum of four wavelengths and the results
can be mathematically calculated to get sums, differences and relationships. Refer to Configure a
program with multi wavelengths on page 27.
Free programming—Select to customize a procedure. Refer to Configure a freely programmed test
on page 27.
Edit Edits a selected program from the list.
Delete Removes a selected program from the list of user programs. Push Delete and select the applicable
program from the list. Push Done.
Send Sends the selected program to the connected hardware. Push Send and select the applicable
program from the list. Push Done.
Cancel Interrupts the ongoing transaction.
Done Accepts the selected option.
Start Starts the selected user program.
7.4.7.1 Configure a program with a single wavelength
1. Push , then select Sample analysis > User programs > Add > Single wavelength.
2. Push OK.
26
English
3. Select an option.
Option Description
Name Enters a name for the new single wavelength program.
Units Selects the unit for concentration or adds a new unit: mg/L (default), g/L, ng/L,
ppm, ppb
Wavelength Enters the wavelength: 320 to 1100 nm (default: 560 nm)
Resolution Selects the number of decimal digits for the concentration: 1 (default), 0.1,
0.01 or 0.0001
Chemical form 1 Enters the chemical formula that will show the analysis parameter.
Calibration curve values Shows a calibration curve. Enter values, measuring standards or formulas.
Push Start assistant to get instructions for the setup.
Upper limit Enters the maximum measured value if enabled.
Lower limit Enters the minimum measured value if enabled.
Timers Adds a maximum of four timers to the program.
Chemical forms Enters a maximum of four chemical forms and four factors for the calculation.
4. Push Save.
7.4.7.2 Configure a program with multi wavelengths
1. Push , then select Sample analysis > User programs > Add > Multi wavelength.
2. Push OK.
3. Select an option.
Option Description
Name Enters a name for the new multi wavelength program.
Units Selects the unit for concentration or adds a new unit: mg/L (default), g/L, ng/L,
ppm, ppb
Absorbance formula Selects the absorbance formula to calculate the number of wavelengths and
coefficient keys that will show below. Default: 400 nm, 500 nm, 600 nm, 800 nm
and K₁A₁+K₂A₂
Wavelength Enters the wavelength: 320 to 1100 nm (default: 560 nm)
Resolution Selects the number of decimal digits for the concentration: 1 (default), 0.1,
0.01 or 0.0001
Chemical form 1 Enters the chemical formula that will show the analysis parameter.
Calibration curve values Makes a calibration curve. Enter values, measuring standards or formulas.
Push Start assistant to get instructions for the setup.
Upper limit Enters the maximum measured value if enabled.
Lower limit Enters the minimum measured value if enabled.
Timers Adds a maximum of four timers to the program.
Chemical forms Enters a maximum of four chemical forms and four factors for the calculation.
4. Push Save.
7.4.7.3 Configure a freely programmed test
1. Push , then select Sample analysis > User programs > Add > Free programming.
2. Push OK.
English
27
3. Select an option.
Option Description
Name Enters a name for the new program.
Version Enters an abbreviation or a number for the version.
Measurement process Sets the measurement formula.
Formulas Sets the formula with which the test result is calculated.
Formula variables Enters the wavelengths, factors, constants and more.
Timers Adds a maximum of four timers to the program.
Control number Calculates based on the selected measurement procedure.
4. Push OK.
7.5 Use the camera (optional)
Use the external camera to:
• Show, keep and send the certificate of analysis for each chemical consumable package.
4
Refer to
step a.
• Use the QR code to update the chemistry database with the latest calibration information. Refer to
step b.
• Read the sample information. Special labels are used to record the information about where, by
whom and when the sample was collected. The handwritten text (in uppercase letters) is imported
by the external camera. Refer to step c.
1. Make sure that the camera is correctly installed and shows on the display. Refer to Install the
camera (optional) on page 11.
2. Push . The camera is set on and a window with a live video stream shows.
a. Hold the package for chemical consumables 5 to 10 cm (2 to 4 inches) above the camera.
A video stream is shown on the screen to help the user navigate through the package so that
the data matrix code can be read by the camera.
b. Put the barcode cuvette in cell compartment 1.
The system examines if necessary updates are available. If yes, the camera is set to on and a
window with a live video stream shows.
c. Hold the bottle with the sample ID label in front of the camera. Refer to Figure 9.
Note: Use only capital letters written with a black pen with a line width of 0.6 to 0.7 mm (0.023 to
0.028 inches)
The information is read, decoded and shown on the screen. Correct the data if necessary.
3. Select an option.
4. Push OK to go back to the main menu.
4
Each package for chemical consumables has a data matrix code on the bottom with the most
important information.