Option Description
Reagent blank Set to on to add the reagent blank value from the measured values. Set to on to
subtract the reagent blank value from the measured values. The reagent blank sample
is used as an offset during calibration. The shape of the calibration curve stays the
same. Default: Off
Vial temperature
detection
Set to on to measure the temperature of the vial. A warning will show if the vial is too
hot or too cold for correct measurements. Default: On. Refer to Table 4 on page 15.
Temperature
compensation
Set to on to adjust for vial temperature effects on the measurement. Default: On. Refer
to Table 4 on page 15.
Turbidity detection Set to on to measure the turbidity. A warning will show if the turbidity is too high for
correct measurements. Default: On. Refer to Table 4 on page 15.
5. Push Save.
7.4.2 Select a Hach program
More than 200 pre-programmed Hach procedures are in Hach programs. All of the important
parameters and timers are included in the procedures.
1. Push , then select Sample analysis > Hach programs.
2. Select an option.
Option Description
Search Finds the procedures by name and program number.
Sort Sorts the list by program number (#), Parameter, Method# and favorites.
Star icon Adds a Hach program to a favorites list.
3. Select the applicable Hach program.
4. Push Start.
The procedure name shows on the left side of the display.
5. Push to change the settings.
Option Description
Chemical
form
Selects the chemical form and the applicable measurement range.
Reading unit Selects the absorbance (default), concentration or transmittance percentage.
Abs (Absorbance readings)—The light absorbed by the sample is measured in absorption
units.
%T (Transmittance reading)—The percentage of light that passes through the sample and is
measured by the detector.
Conc (Concentration readings)—When concentration is selected, the system changes the
absorbance value to a concentration value. The concentration is calibrated in a special
procedure based on absorption. The special procedure uses a linear or polynomial
calibration curve to get the best adjustment.
Dilution Set to on to enter a dilution factor (default: off). The number entered under Dilution is
multiplied by the result to compensate for the adjustment. For example, if the sample was
diluted by a factor of 2, enter 2. Default: Off
Note: When Dilution is set to on, the dilution icon will show on the display.
Standard
adjust
The instructions for the procedure parameters specify if a standard adjustment is necessary
and how to continue.
Reagent
blank
Set to on to add the reagent blank value from the measured values. Set to on to subtract the
reagent blank value from the measured values. The reagent blank sample is used as an
offset during calibration. The shape of the calibration curve stays the same. Default: Off
6. Push Save.
22
English
7.4.2.1 Start a Hach program
1. Push , then select Sample analysis > Hach programs.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.3 Configure the wavelength scan settings
1. Push , then select Sample analysis > Wavelength scan.
2. Push to change the settings.
3. Select an option.
Option Description
Integral Set to on to show the integral for the scan. The integral gives the area and the
derivative of the integral gives the original function. Default: Off
Wavelength range Enters the wavelength spectrum between 320 to 1100 nm. Default: 340 to 900 nm
Step Enters the wavelength interval for the scan: 1 nm (default), 2 nm or 5 nm
Reference scan Selects a saved scan to show with new scans as a reference. Default: Off
Reading unit Selects the absorbance or transmittance percentage: Abs (default) or %T
Scaling Selects automatic (default) or manual to adjust the Y-axis scale during the scan.
Automatic —The y-axis is automatically adjusted so that the total scan shows.
Manual —An area can be limited so that only a small section of the scan shows.
Transmittance range
or Absorbance range
Enters the transmittance or absorbance range for the y-axis: -0.300 minimum to
3.500 maximum. The function is disabled when Scaling is set to Automatic.
Default: -0.300 to 3.500 ABS
Grid Sets the grid lines in the graph to on. Default: On
4. Push Save.
7.4.3.1 Start a wavelength scan measurement
Note: Make sure that the compartment lid is closed during the measurement.
If the reference scan is selected, the reference scan will show in a different color than the scan
measured. Refer to Configure the wavelength scan settings on page 23.
1. Push , then select Sample analysis > Wavelength scan.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
The graph of the absorption or transmission values for the scanned wavelengths is continuously
shown.
6. The Wavelength scan is complete, if:
• The graph shows full-size.
• The scaling of the x-axis occurs automatically.
• The cursor functions in the vertical navigation bar are highlighted.
English
23
• A signal sounds.
The measured value shows on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.4 Configure the time course settings
1. Push , then select Sample analysis > Time course.
2. Push to change the settings. Refer to Table 1 on page 13.
3. Select an option.
Option Description
Reading unit Selects the absorbance or transmittance percentage: Abs (default) or %T
Scaling Selects automatic (default) or manual to adjust the Y-axis scale during the scan.
Automatic —The y-axis is automatically adjusted so that the total scan shows.
Manual —An area can be limited so that only a small section of the scan shows.
Transmittance range or
Absorbance range
Enters the transmittance or absorbance range for the y-axis: -0.300 minimum to
3.500 maximum. The function is disabled when Scaling is set to Automatic.
Default: -0.300 to 3.500 ABS
Wavelength Enters the wavelength. Options: 320 to 1100 nm (default: 560 nm)
Total time Enters the total time for the measurements (48 hours maximum). Default:
5 minutes
Reading interval Enters the time interval between measurements in seconds, minutes or hours.
Default: 12 seconds
Grid Sets the grid lines in the graph to on. Default: On
4. Push Save.
7.4.4.1 Start a time course measurement
Note: Make sure that the compartment lid is closed during the measurement.
1. Push , then select Sample analysis > Time course.
2. Put the sample cell that contains the zero solution into the cell compartment.
3. Push Zero.
The Read function is only enabled after the zero measurement is completed.
4. Put the sample cell into the cell compartment.
5. Push Read.
6. Select an option.
Option Description
Add mark Records the subsequent collected data point. The data point can show an important
measurement, such as the addition of a sample or other reagent. The data point is also
shown in the table.
Stop reading Select to stop the measurements.
The measured values show on the display. The measured values are automatically saved to the
data log if Auto save is enabled. Refer to Configure the instrument on page 15.
7.4.5 Configure the single wavelength settings
Configure the single wavelength settings as follows:
1. Push , then select Sample analysis > Single wavelength.
2. Push to change the settings.