Figure 1 Product overview 1 Touch screen 6 Power switch 2 Cell compartment lid 7 Power supply connection 3 USB port type C 8 Ethernet port 4 USB dust cover 9 USB port type A 5 USB port type A 10 USB port type C 3.2.1 The…

3.1.3 Chemical and biological safety
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or
chemical feed system for which there are regulatory limits and monitoring requirements related to
public health, public safety, food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to have sufficient and
appropriate mechanisms in place for compliance with applicable regulations in the event of
malfunction of the instrument.
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
• Do not use the instrument in samples that could cause a biological hazard.
• Observe all cautionary information printed on the original solution containers and safety data
sheets prior to their use.
• Dispose of all consumed solutions in accordance with the local and national regulations and laws.
• Select the type of protective equipment suitable to the concentration and quantity of the dangerous
material being used.
3.1.4 Korean certification
Company name: Hach Lange GmbH
Name of the product: Measurement and Test equipment
Model name: DR4900
Time of manufacture : > 2026-05
Manufacturer : Hach Lange GmbH / Germany
Management number : hle-DR4900
https://www.rra.go.kr/selform/hle-DR4900
3.2 Product overview
N O T I C E
Perchlorate Material—Special handling may apply. Refer to www.dtsc.ca.gov/perchlorate. This perchlorate
warning applies only to primary batteries (provided singly or installed on this equipment) when sold or distributed
in California, USA.
The DR4900 is a VIS spectrophotometer that measures in the wavelength range of 320 to 1100 nm.
The instrument is used to measure various parameters in drinking water, wastewater and industrial
applications. The instrument comes with a complete set of application programs: Hach programs
(pre-installed methods), LCK or TNTplus methods (barcode programs), user programs, single
wavelength, multi-wavelength, wavelength scan and time course modes. Refer to Figure 1.
6
English

Figure 2 DR4900 light beam path
1 Halogen lamp 6 Lens 11 Lens
2 Lamp opening 7 Filter wheel 12 Measurement element
3 Grating 8 Beam splitter mirror 13 Cell compartment 1 for round
cells
4 Grating angle indicator 9 Reference element
5 Exit slit 10 Cell compartment 2 for
rectangular cells
Section 4 Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
4.1 Operating environment
Install the instrument:
• On a level, rigid surface with sufficient load bearing capacity. Do not put objects below the
instrument.
• In a location without direct sunlight.
• Away from very high temperatures caused by sunlight, heaters or other sources.
• On a surface so that the power cord is not bent and the instrument can be easily set to off with the
power switch.
• In an ambient temperature of 10–40 °C (50–104 °F).
• In relative humidity of less than 80%. Do not let moisture collect on the instrument.
• Away from high humidity or chemically contaminated air or instrument damage will occur.
• With a minimum clearance of 15 cm around the instrument. Damage will occur if the instrument
becomes too hot.
8
English