SMW200A_specs_en_3606-8037-22_v2700.pdf - 第42页

Version 27.00, October 2024 42 Rohde & Schwar z R&S ® SMW200A Vec tor Signal Generator Differential analog I/Q output s ( R&S ® SMW-K16 option) This option can be installed once if the instrument is equipped …

100%1 / 88
Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 41
Standard baseband characteristics
Standard baseband is not available for instruments equipped with R&S
®
SMW-B1044O, R&S
®
SMW-B2044O, R&S
®
SMW-B1056O or
R&S
®
SMW-B1067O frequency options.
Internal baseband characteristics
(R&S
®
SMW-B13 or R&S
®
SMW-B13T option)
The R&S
®
SMW-B13 option provides one I/Q path to the RF section (to RF path A) as well as one analog I/Q output (i.e. one I and one
Q output connector). The R&S
®
SMW-B13T option provides two I/Q paths to the RF section (if two RF paths are installed) as well as
two analog I/Q outputs. With two RF paths, R&S
®
SMW-B13T is required.
Either R&S
®
SMW-B13 or R&S
®
SMW-B13T must be installed on the instrument.
R&S
®
SMW-B13 and R&S
®
SMW-B13T cannot be installed in instruments equipped with R&S
®
SMW-B1044O, R&S
®
SMW-B2044O,
R&S
®
SMW-B1056O or R&S
®
SMW-B1067O frequency options.
D/A converter
Data rate
200 MHz
Resolution
16 bit
Sample rate
800 MHz (internal interpolation · 4)
Aliasing filter
with amplitude, group delay and S
i
correction
Bandwidth, rolloff to 0.1 dB
80 MHz
SFDR (excluding harmonics)
up to 10 MHz
< 80 dBc
up to 80 MHz
< 73 dBc
I/Q impairments (digital baseband)
These impairments are set in the digital baseband section of the R&S
®
SMW200A. They
act on the I/Q signal sent to the I/Q modulator/RF section, as well as on the I/Q signals
at the analog or digital I/Q outputs (of the respective path).
Carrier leakage
Setting range
10 % to +10 %
Setting resolution
0.01 %
I Q (imbalance)
Setting range
1 dB to +1 dB
Setting resolution
0.001 dB
Quadrature offset
Setting range
10° to +10°
Setting resolution
0.01°
Analog I/Q outputs (R&S
®
SMW-B13 or R&S
®
SMW-B13T option)
Number of I/Q outputs
with R&S
®
SMW-B13 option
1
with R&S
®
SMW-B13T option
2
Output impedance
50 Ω
Output voltage
EMF (output voltage depends on set
modulation signal)
1 V (V
p
)
Offset
EMF
< 1 mV
Frequency response
9
at R
L
= 50 Ω
Magnitude
up to 10 MHz
0.02 dB (meas.)
up to 80 MHz
0.03 dB (meas.)
I/Q balance
10
at R
L
= 50 Ω
Magnitude
up to 10 MHz
0.01 dB (meas.)
up to 80 MHz
0.02 dB (meas.)
Spectral purity
at R
L
= 50 Ω
SFDR (sine wave)
up to 2 MHz
> 70 dB
up to 20 MHz
60 dB (meas.)
Wideband noise
10 MHz sine wave at 1 MHz offset
155 dBc (typ.)
9
"Optimize internal I/Q impairments for RF output" switched off.
10
Value applies after 1 hour warm-up time and recalibration for 4 hours of operation and temperature variations of less than +5 °C.
Version 27.00, October 2024
42 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Differential analog I/Q outputs (R&S
®
SMW-K16 option)
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13 option. If the instrument is equipped with the
R&S
®
SMW-B13T option, differential analog I/Q outputs can be used either on signal path A or B with one R&S
®
SMW-K16 option. For
differential analog I/Q outputs to be used on signal paths A and B simultaneously, two R&S
®
SMW-K16 must be installed.
Output impedance
Single-ended
50 Ω
Differential
100 Ω
Output voltage (V
out
)
output voltage depends on set modulation signal
Single-ended
EMF
0.02 V to 2 V (V
p
)
Resolution
1 mV
Differential
EMF
0.04 V to 4 V (V
pp
)
Resolution
2 mV
Bias voltage (V
bias
)
Single-ended
EMF
4 V to (+4 V V
out
)
Differential
EMF
(4 V + V
out
/ 2 + V
offset
/ 2) to
(+4 V V
out
/ 2 V
offset
/ 2)
Resolution
2 mV
Uncertainty
1 % + 4 mV
Offset voltage (V
offset
)
Differential
EMF
(4 V + V
out
/ 2 + V
bias
/ 2) to
(+4 V V
out
/ 2 V
bias
/ 2)
Resolution
0.1 mV
Uncertainty
1 % + 0.1 % · bias voltage + 1 mV
Differential signal balance
at R
L
= 50 Ω, output voltage > 0.5 V (V
p
)
Magnitude
up to 10 MHz
< 0.2 dB, 0.05 dB (meas.)
up to 80 MHz
0.2 dB (meas.)
Frequency response
11
at R
L
= 50 Ω, output voltage > 0.5 V
(V
p
)
Magnitude
up to 10 MHz
0.02 dB (meas.)
up to 80 MHz
0.03 dB (meas.)
11
"Optimize internal I/Q impairments for RF output" switched off.
Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 43
Digital baseband inputs/outputs
Depending on the installed software and hardware options, the R&S
®
SMW200A is able to receive digital baseband signals and to
output digital baseband signals. The digital I/Q input/output can be used for the lossless connection of the R&S
®
SMW200A to the
digital I/Q input/output of other Rohde & Schwarz instruments (for example the CMW500 wideband radio communication tester in
fading applications).
Digital baseband outputs: At least one R&S
®
SMW-K18 option must be installed. This option can be installed once if the instrument is
equipped with the R&S
®
SMW-B13 option. If the instrument is equipped with the R&S
®
SMW-B13T option, digital baseband outputs can
be used either on signal path A or B with one R&S
®
SMW-K18 option. For digital baseband outputs to be used on signal paths A and B
simultaneously, two R&S
®
SMW-K18 must be installed. Furthermore, to enable two or more digital baseband outputs in MIMO modes,
two R&S
®
SMW-K18 must be installed.
Signal outputs
analog and digital, digital only
with 2 × R&S
®
SMW-K18 installed
analog and digital, digital only, digital only
multiplexed
Digital only
The streams are output via the digital I/Q outputs only; analog I/Q outputs are not
available. External modulation signals can be output via the RF outputs (I/Q modulation
mode: external wideband I/Q).
Note: System configurations with more than 4 streams are not available in this mode.
with R&S
®
SMW-K551 installed
The instrument runs at reduced speed
depending on the device connected to the
digital I/Q output (slow I/Q).
Digital only multiplexed
The streams are output via BBMM1 and BBMM2 in multiplexed mode, i.e. up to 4
streams are output via a single digital output. Analog I/Q outputs are not available.
External modulation signals can be output via the RF outputs (I/Q modulation mode:
external wideband I/Q).
Note: All system configurations available on the instrument are available in this mode.
with R&S
®
SMW-K551 installed
The instrument runs at reduced speed
depending on the device connected to the
digital I/Q output (slow I/Q).
Analog and digital
The instrument runs in regular operating mode, both analog and digital outputs are
available, slow I/Q is not possible.
Number of digital outputs
according to selected system configuration
(see table below)
Number of streams per digital output
digital only
1
digital only multiplexed
1 to 4
Bandwidth
general
according to selected system configuration
(see section Multichannel, MIMO, fading
and noise, specifications for
R&S
®
SMW-K74, -K75, -K76 options)
4 streams mapped to one digital output
40 MHz
The following table gives an overview of which software and hardware options are required for which digital I/Q connectivity:
Minimum required R&S
®
SMW200A
options
Digital I/Q inputs
Digital I/Q outputs
R&S
®
SMW-B13 + 1 × R&S
®
SMW-K18
1
R&S
®
SMW-B13T + 2 × R&S
®
SMW-K18
2
1 × R&S
®
SMW-B10
1
1 × R&S
®
SMW-B10 + R&S
®
SMW-B13 +
1 × R&S
®
SMW-K18
1
1
1 × R&S
®
SMW-B10 + R&S
®
SMW-B13T +
2 × R&S
®
SMW-K18
1
2
2 × R&S
®
SMW-B10
2
2 × R&S
®
SMW-B10 + R&S
®
SMW-B13 +
1 × R&S
®
SMW-K18
2
1
2 × R&S
®
SMW-B10 + R&S
®
SMW-B13T +
2 × R&S
®
SMW-K18
2
2