SMW200A_specs_en_3606-8037-22_v2700.pdf - 第47页
Version 27.00, October 2024 Rohde & Schwar z R&S ® SMW200A Vec tor Signal Generator 47 External trigger input selectable from U SER 1, 2, 3 on front panel or T/M/C 1, T/M 2, T/M 3 of respec tive baseband gene rat…

Version 27.00, October 2024
46 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Standard baseband generator (R&S
®
SMW-B10 option) –
arbitrary waveform mode
One or two R&S
®
SMW-B10 can be installed. Their I/Q signals can be assigned a frequency offset and/or be added in the digital
domain with settable level ratio.
Prerequisite: Either R&S
®
SMW-B13 or R&S
®
SMW-B13T must be installed.
Waveform length
1 sample to 64 Msample in one-sample
steps
with R&S
®
SMW-K511 option
(memory extension)
1 sample to 512 Msample in one-sample
steps
with R&S
®
SMW-K512 option
(memory extension)
1 sample to 1 Gsample in one-sample steps
Supported file formats
.wv, .mat, .csv, .iq.tar
Nonvolatile memory
hard disk
Sample resolution
equivalent to D/A converter
16 bit
Sample rate
400 Hz to 150 MHz
with R&S
®
SMW-K522 option
400 Hz to 200 MHz
Sample frequency error
internal clock
< (5 · 10
−14
+ relative deviation of reference
frequency) · sample rate (nom.)
Sample clock source
internal, external
Bandwidth (RF)
using the maximum sample rate,
rolloff to –0.1 dB
120 MHz
using a reduced sample rate,
rolloff to –0.1 dB
(The waveform is automatically
interpolated to the internal sample rate
of 150 MHz.)
0.8 · sample rate
Bandwidth (RF) with R&S
®
SMW-K522
option
using the maximum sample rate,
rolloff to –0.1 dB
160 MHz
using a reduced sample rate,
rolloff to –0.1 dB
(The waveform is automatically
interpolated to the internal sample rate
of 200 MHz.)
0.8 · sample rate
Frequency offset
The frequency offset can be used to shift the center frequency of the wanted baseband
signal. The restrictions caused by the modulation bandwidth still apply.
Frequency offset setting range
–60 MHz to +60 MHz
with R&S
®
SMW-K522 option
–80 MHz to +80 MHz
Frequency offset setting resolution
0.01 Hz
Frequency offset error
< 7 · 10
–7
Hz + relative deviation of
reference frequency · frequency offset
(nom.)
Triggering
A trigger event restarts I/Q generation. The I/Q signal is then synchronous with the
trigger (with a specific timing jitter).
Trigger source
event triggered via GUI or remote
command
internal
event triggered by other baseband
generator
internal (baseband A/B)
event triggered by external trigger signal
external
Trigger modes
The signal is generated continuously.
auto
The signal is generated continuously.
A trigger event causes a restart.
retrig
The signal is started only when a trigger
event occurs. Subsequent trigger
events are ignored.
armed auto
The signal is started only when a trigger
event occurs. Every subsequent trigger
event causes a restart.
armed retrig
The signal is started only when a trigger
event occurs. The signal is generated
once.
single

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 47
External trigger input
selectable from USER 1, 2, 3 on front panel
or T/M/C 1, T/M 2, T/M 3 of respective
baseband generator on rear panel
Connector type
USER 1, 2, 3 on front panel,
T/M/C 1, T/M 2, T/M 3 of respective
baseband generator on rear panel
BNC female
Input level
0 V to 3 V (nom.)
Threshold
USER 1, 2, 3
settable from 0.1 V to 2.0 V
T/M/C 1, T/M 2, T/M 3
settable from 0.3 V to 2.0 V
Input damage voltage
–0.5 V; 3.8 V
Input impedance
selectable
1 kΩ or 50 Ω (nom.)
Trigger jitter
±2.5 ns
External trigger delay
Setting range
0 sample to 2.147 · 10
9
sample
Setting resolution
without R&S
®
SMW-B14 option
5 ns
with R&S
®
SMW-B14 option
1/fading clock rate (= 5 ns or 10 ns)
External trigger inhibit
Setting range
0 sample to (21.47 s · sample rate) sample
Setting resolution
1 sample
External trigger pulse width
> 7.5 ns
Marker signals
Number of marker signals
3
Operating modes
unchanged, restart, pulse, pattern, ratio
Marker outputs
selectable from USER 1, 2, 3 on front panel
or T/M/C 1, T/M 2, T/M 3 of respective
baseband generator on rear panel
Connector type
USER 1, 2, 3 on front panel,
T/M/C 1, T/M 2, T/M 3 of respective
baseband generator on rear panel
BNC female
Level
LVTTL
Marker delay
Setting range
0 sample to (waveform length – 1) sample
Setting resolution
1 sample
Marker duration
Minimum value
1 sample
Multisegment waveform mode
Number of segments
1 to 1024
Changeover modes
GUI, remote control, external trigger
Extended trigger modes
same segment, next segment, next
segment seamless, sequencer
Changeover time
at 50 MHz clock rate, external trigger,
without clock change
20 µs (meas.)
Seamless changeover
output up to end of current segment,
followed by changeover to next segment
Sequencer play list length
max. 1024
Sequencer segment repetitions
max. 1 048 575
Multicarrier waveform mode
Number of carriers
max. 512
Total RF bandwidth
max. 120 MHz
with R&S
®
SMW-K522 option
max. 160 MHz
Carrier spacing
Setting range
depends on number of carriers and signal
RF bandwidth
Setting resolution
0.01 Hz
Crest factor modes
maximize, minimize, off
Signal period modes
longest file, shortest file, user (max. 1 s)
Single carrier gain
Setting range
–80 dB to 0 dB
Setting resolution
0.01 dB
Single carrier start phase
Setting range
0° to 360°
Setting resolution
0.01°
Single carrier delay
Setting range
0 s to 1 s
Setting resolution
1 ns

Version 27.00, October 2024
48 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Extended sequencing (R&S
®
SMW-K501 option)
The R&S
®
SMW-K501 option enables waveform sequencing and real-time signal generation for ultra long playtime. Waveform
variations such as offset frequency, amplitude and phase are calculated in real-time and do not require precalculated waveforms.
The R&S
®
SMW-K501 option offers two different modes:
In user mode, all sequences are based on user-defined XML based lists with up to 5 levels of nested loops. Special list types for
frequency changes over time and amplitude changes over time are also available.
In pulse sequencer mode, the extended sequencing is controlled by the external R&S
®
Pulse Sequencer Software, a powerful software
tool for simulating complex sequencing scenarios.
At least one R&S
®
SMW-B10 option (standard baseband generator) must be installed. If two R&S
®
SMW-B10 options are installed
(signal paths A and B), extended sequencing can be used either on signal path A or B with one R&S
®
SMW-K501 option. For extended
sequencing to be used simultaneously on signal paths A and B, two R&S
®
SMW-K501 options must be installed.
General settings
Modes
sequencing via user-defined XML lists
user
controlled by external
R&S
®
Pulse Sequencer Software
(R&S
®
SMW-K300 required)
pulse sequencer
User mode
List types
Sequencing lists define an arbitrary
number of entries that represent either a
waveform or a sublist with further entries.
sequencing list
Time lists store a list of different off times
between waveform segments. They can
be referenced in sequence entries.
time list
Attenuation lists define the power level of
the output signal over time.
attenuation list
Hopping lists define frequency offsets of
the output signal over time.
hopping list
Sequence
link to a sequencing list XML file
Attenuation over time
link to an attenuation list XML file
Hopping
link to a hopping list XML file
Pulse sequencer mode
see R&S
®
Pulse Sequencer Software specifications (PD 3607.1388.22)
Waveform segments
Segment length
1 sample to 64 Msample
Minimum memory allocation
64 sample
Maximum number of segments
depends on segment lengths and
baseband generator ARB memory size
Waveform sequences
Sequencing
continuously repeating
Maximum number of segments per
sequence
depends on segment lengths and
baseband generator ARB memory size
Maximum number of segment repetitions
2
32
Clock
see section Standard baseband generator
(R&S
®
SMW-B10 option) – arbitrary
waveform mode
Triggering
see section Standard baseband generator
(R&S
®
SMW-B10 option) – arbitrary
waveform mode
Marker signals
Number of marker signals
3
Operating modes
marker at every start of sequence
restart
marker 1 embedded in waveform
unchanged
XML-defined marker for each entry
entry
Marker outputs
see section Standard baseband generator
(R&S
®
SMW-B10 option) – arbitrary
waveform mode
Marker delay
see section Standard baseband generator
(R&S
®
SMW-B10 option) – arbitrary
waveform mode
Marker duration
see section Standard baseband generator
(R&S
®
SMW-B10 option) – arbitrary
waveform mode