Wideband differential analog I/Q outputs (R&S
®
SMW-K17 option)
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13XT option. Differential analog I/Q outputs can
be used on signal path A only. If the differential output mode is activated, analog I/Q outputs for signal path B are not available.
Output impedance
Single-ended
50 Ω
Differential
100 Ω
Output voltage (V
out
)
output voltage depends on set modulation signal
Single-ended
EMF
0.02 V to 1 V (V
p
)
Resolution
0.1 mV
Differential
EMF
0.04 V to 2 V (V
pp
)
Resolution
0.1 mV
Bias voltage (single-ended and differential)
EMF
–0.2 V to +2.5 V
15
Resolution
0.1 mV
Uncertainty
1 % + 2 mV
Offset voltage
Differential
EMF
(–2 V + V
out
) to (+2 V – V
out
)
RF envelope: on
(R&S
®
SMW-K540 required), EMF
–2 V to +2 V
Resolution
0.1 mV
Uncertainty
1 % + 1 mV
Differential signal balance
at R
L
= 50 Ω, output voltage > 0.5 V (V
p
)
Magnitude
up to 100 MHz
0.1 dB (meas.)
up to 500 MHz
0.15 dB (meas.)
up to 1000 MHz
0.2 dB (meas.)
Frequency response
16
at R
L
= 50 Ω, output voltage > 0.5 V
(V
p
)
Magnitude
up to 100 MHz
0.1 dB (meas.)
up to 1000 MHz
0.2 dB (meas.)
Wideband noise
10 MHz sine wave at 1 MHz offset
–160 dBc (typ.)
Measured phase noise of wideband analog I/Q outputs; single-ended sine wave with f = 100 MHz
15
The magnitude of the sum of output voltage and bias voltage must not exceed 4 V.
16
"Optimize internal I/Q impairments for RF output" switched off.
Running ...
R& S FSUP 50 Signal Source A nalyzer LO CKED
Set t ings Residual Noise [T1 w / o spurs] P hase Det ect or +40 dB
Signal F requenc y: 1 00 .0 00 0 0 2 MH z I nt P H N (10 0 .0 .. 1 .0 M ) -8 3 .3 dBc
Signal L ev el: 3 .22 dBm Res idual P M 5 .5 2 9 m°
C ros s C orr M ode H armonic 1 R es idual FM 1 0 .3 6 5 H z
I nternal Ref T uned I nternal P has e Det RM S Jitter 0 .15 3 6 ps
P has e Nois e [dBc /H z]
RF A tten 5 dB
T op - 8 0 dBc /Hz
1 kHz
10 kHz
100 kHz
100 Hz
1 MHz
-170
-160
-150
-140
-130
-120
-110
-100
-90
LoopBW 100 Hz
1 CLRW R
SMTH 1%
2 CLRW R
A
SPR OFF
TH 0dB
Frequency Offset
SMB100A Ser.No 175321
Date: 31.MAR.2003 05:52:13

Version 27.00, October 2024
52 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Measured phase noise of wideband analog I/Q outputs; single-ended sine wave with f = 300 MHz
Measured phase noise of wideband analog I/Q outputs; single-ended sine wave with f = 1 GHz
Running ...
R& S FSUP 50 Signal Source A naly zer LO CKED
Set t ings Residual Noise [ T1 w / o spurs] P hase Det ect or +40 dB
Signal Frequenc y: 1 .00 0 0 0 0 GH z I nt P H N (10 0 .0 .. 1 .0 M ) -6 4 .8 dB c
Signal Level: 4 .33 dB m Res idual P M 4 6 .85 0 m°
C ros s C orr M ode H armonic 1 Res idual FM 3 1 .87 3 H z
I nternal Ref T uned I nternal P has e Det RM S J itter 0 .1 3 0 1 ps
P has e N ois e [dBc /H z]
RF A tten 5 dB
T op -7 0 dB c/H z
1 kHz
10 kHz
100 kHz
100 Hz
1 MHz
-150
-140
-130
-120
-110
-100
-90
-80
LoopBW 300 Hz
1 CLRW R
SMTH 1%
2 CLRW R
A
SPR OFF
TH 0dB
Frequency Offset
SMB100A Ser.No 175321
Date: 31.MAR.2003 05:57:49
Running ...
R& S FSUP 50 Signal Source A nalyzer LOCKED
Set t ings Residual Noise [ T1 w / o spurs] P hase Det ect or +40 dB
Signal Frequenc y: 3 0 0 .0 00 0 0 5 M H z I nt P H N (1 0 0.0 .. 1.0 M ) -7 4 .6 dBc
Signal Level: 2 .7 1 dBm Res idual P M 15 .15 6 m°
C ros s C orr M ode H armonic 1 Res idual FM 1 2 .7 6 6 H z
I nternal Ref T uned Internal P has e D et RM S J itter 0 .14 0 3 ps
P has e N ois e [dBc /H z]
RF A tten 5 dB
T op -8 0 dBc /H z
1 kHz
10 kHz
100 kHz
100 Hz
1 MHz
-170
-160
-150
-140
-130
-120
-110
-100
-90
LoopBW 300 Hz
1 CLRW R
SMTH 1%
2 CLRW R
A
SPR OFF
TH 0dB
Frequency Offset
SMB100A Ser.No 175321
Date: 31.MAR.2003 05:55:42

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 53
Digital baseband inputs/outputs for wideband baseband
Depending on the installed software and hardware options, the R&S
®
SMW200A is able to receive digital baseband signals and output
digital baseband signals. The digital I/Q input/output can be used for the lossless connection of the R&S
®
SMW200A to the digital I/Q
input/output of other Rohde & Schwarz instruments.
Digital baseband outputs: At least one R&S
®
SMW-K19 option must be installed. Digital baseband outputs can be used either on signal
path A or B with one R&S
®
SMW-K19 option. For digital baseband outputs to be used on signal paths A and B simultaneously,
two R&S
®
SMW-K19 must be installed. To enable two or more digital baseband outputs in multichannel or other advanced modes,
two R&S
®
SMW-K19 must be installed.
The R&S
®
SMW-K19 option requires R&S
®
SMW-B13XT with DACW board revision 4.00 or higher.
Signal outputs
system configuration mode: standard
analog only, digital only (HS
17
)
system configuration mode: advanced
18
analog and digital, analog and digital (HS),
digital only (HS)
Digital only (HS)
The streams are output via the digital I/Q outputs only (HS DIG I/Q interface standard).
Analog I/Q outputs are not available. External modulation signals can be output via the
RF outputs (I/Q modulation mode: external wideband I/Q).
with R&S
®
SMW-K551 installed and
system configuration mode: advanced
The instrument runs at reduced speed,
depending on the device connected to the
digital I/Q output (slow I/Q).
Analog and digital
The instrument runs in regular operating mode, both analog and digital outputs
(DIG I/Q interface standard) are available.
Analog and digital (HS)
The instrument runs in regular operating mode, both analog and digital outputs
(HS DIG I/Q interface standard) are available.
Analog only
The instrument runs in regular operating mode, only analog outputs are available.
Number of digital outputs
according to selected system configuration
(see table below)
signal outputs: digital only (HS)
maximum 2 (on R&S
®
SMW-B13XT)
signal outputs: analog and digital
maximum 8 (on R&S
®
SMW-B13XT and
R&S
®
SMW-B15) depending on
entities · RX antennas of MIMO/SIMO
configuration
signal outputs: analog and digital (HS)
maximum 2 (on R&S
®
SMW-B13XT)
Number of streams per output
signal outputs: digital only (HS)
system configuration mode: standard
1 to 2
system configuration mode: advanced
1 to 8
Number of streams per input
system configuration mode: standard;
signal outputs: analog only,
HS DIG I/Q
1 to 2
system configuration mode: advanced; signal outputs: analog and digital,
200 MHz, interface either DIG I/Q or HS DIG I/Q
HS DIG I/Q
1 to 2
DIG I/Q
1 to 2
system configuration mode: advanced;
signal outputs: analog and digital,
400 MHz or 800 MHz, HS DIQ I/Q
1 to 2
17
HS = high-speed.
18
The following functions are not available in advanced system configuration mode: analog modulation, modulation sources for analog modulation,