SMW200A_specs_en_3606-8037-22_v2700.pdf - 第56页
Version 27.00, October 2024 56 Rohde & Schwar z R&S ® SMW200A Vec tor Signal Generator Interface Standard DIG I/Q, in line wi th R&S ® Digital I/Q In terface PAD- R 21 , I/Q data and con trol signals, da ta a…

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 55
I/Q data
Resolution
18 bit
Logic format
two’s complement
Physical signal level
Setting range
0 to –60 dBFS
Resolution
0.01 dBFS
Bandwidth (RF)
system configuration mode: advanced
0.8 · sample rate
Control signals
markers
3
Earliest supported R&S
®
SMW200A
firmware version
4.30.046.221
HS DIQ I/Q interface
Interface
Standard
HS DIG I/Q,
in line with R&S
®
Digital I/Q Interface 40G
PAD-R
20
(DIG I/Q 40G),
I/Q data and control signals
Level
LVDS
Connector
QSFP+ / QSFP 28
I/Q sample rate
Sample rate
maximum sample rate depends on connected receiving device and system
configuration mode
system configuration mode: standard
40G
up to 1.05 GHz
50G
up to 1.25 GHz
system configuration mode: advanced
analog and digital (HS)
1000 MHz
digital only (HS)
up to 250 MHz
Resolution
0.001 Hz
Frequency uncertainty
< (1 · 10
−12
+ relative deviation of
reference frequency) · sample rate (nom.)
I/Q data
Resolution
up to 16 bit
Logic format
two’s complement
Physical signal level
Setting range
0 to –60 dBFS
Setting resolution
0.01 dBFS
Bandwidth (RF)
system configuration mode: standard
0.83 · sample rate
system configuration mode: advanced
0.8 · sample rate
Control signals
markers
2
Setup external RF with R&S
®
SMW-B13XT to R&S
®
SMW-B9
Earliest supported R&S
®
SMW200A
firmware version
4.70.128.xx
Notes
If both R&S
®
SMW200A have DACW board revision 4.00 and DACW board revision
5.00, use DACW board revision 5.00 as signal source.
Setup external RF with R&S
®
SMW-B13XT to R&S
®
SMM100A
Earliest supported R&S
®
SMW200A
firmware version
4.90.049.xx
Setup external RF with R&S
®
SMW-B13XT to R&S
®
SMCV100B
Earliest supported R&S
®
SMW200A
firmware version
4.90.049.xx
Input parameters
DIQ I/Q interface
Input level
peak level
Peak level
Setting range
referenced to full scale
–60 dB to +3 dB
Resolution
0.01 dB
Crest factor
Setting range
0 dB to +30 dB
Resolution
0.01 dB
Adjust level function
automatically determines peak level and
crest factor of input signal
20
R&S
®
Digital I/Q Interface 40G PAD-R is a Rohde & Schwarz internal company guideline for the transmission of digital I/Q data. It is supported by a
wide range of signal generators, signal analyzers and radio communication testers.

Version 27.00, October 2024
56 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Interface
Standard
DIG I/Q, in line with
R&S
®
Digital I/Q Interface PAD-R
21
,
I/Q data and control signals, data and
interface clock
Level
LVDS
Connector
26-pin MDR
I/Q sample rate
With source "user-defined", the sample rate must be entered via the parameter
"sample rate". With source "Digital I/Q In", the sample rate will be used based on
information provided by the transmitting device.
Source
user-defined, Digital I/Q In
Sample rate
maximum sample rate depends on
connected receiving device
400 Hz to 250 MHz
Resolution
source: user-defined
0.001 Hz
Frequency uncertainty
source: user-defined
< (1 · 10
−12
+ relative deviation of
reference frequency) · sample rate (nom.)
I/Q data
Resolution
18 bit
Logic format
two’s complement
Bandwidth (RF)
system configuration mode: advanced
0.8 · sample rate
Control signals
markers
3
HS DIQ I/Q interface
Input level
peak level
Setting range
–60 dB to +3 dB, referenced to full scale
Setting resolution
0.01 dB
Crest factor
Setting range
0 dB to +30 dB
Setting resolution
0.01 dB
Adjust level function
automatically determines peak level and
crest factor of input signal
Standard
HS DIG I/Q, in line with
R&S
®
Digital I/Q Interface 40G PAD-R
21
(DIG I/Q 40G), I/Q data and control
signals
Level
LVDS
Connector
QSFP+ / QSFP 28
I/Q sample rate
Source
the sample rate will be used based on
information provided by the transmitting
device
HS digital I/Q In
Sample rate
maximum sample rate depends on connected transmitting device and system
configuration mode
system configuration mode: standard
40G
up to 1.05 GHz
50G
up to 1.25 GHz
system configuration mode: advanced
up to 250 MHz
with R&S
®
SMW-K822 option
up to 500 MHz
with R&S
®
SMW-K823 option
up to 1000 MHz
Resolution
0.001 Hz
Frequency uncertainty
< (1 · 10
−12
+ relative deviation of
reference frequency) · sample rate (nom.)
I/Q data
Resolution
16 bit
Logic format
two’s complement
Bandwidth (RF)
system configuration mode: standard
0.83 · sample rate
Control signals
markers
2
21
R&S
®
Digital I/Q Interface PAD-R is a Rohde & Schwarz internal company guideline for the transmission of digital I/Q data. It is supported by a wide
range of signal generators, signal analyzers and radio communication testers.

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 57
Wideband baseband generator (R&S
®
SMW-B9 option) – arbitrary waveform
mode
One or two R&S
®
SMW-B9 can be installed. Their I/Q signals can be assigned a frequency offset.
Prerequisite: R&S
®
SMW-B13XT must be installed.
Waveform length
1 sample to 256 Msample in one-sample
steps
with R&S
®
SMW-K515 option
(memory extension)
1 sample to 2 Gsample in one-sample
steps
Supported file formats
.wv, .mat, .csv, .iq.tar
Nonvolatile memory
hard disk
Sample resolution
equivalent to D/A converter
14 bit
Sample rate
400 Hz to 600 MHz
with R&S
®
SMW-K525 option
400 Hz to 1200 MHz
with R&S
®
SMW-K527 option
400 Hz to 2400 MHz
Sample frequency error
internal clock
< (1 · 10
−12
+ relative deviation of
reference frequency) · sample rate (nom.)
Sample clock source
internal
Bandwidth (RF)
at maximum sample rate,
rolloff to –0.1 dB
500 MHz
at reduced sample rate,
rolloff to –0.1 dB
(The waveform is automatically
interpolated to the internal sample rate of
600 MHz.)
0.833 · sample rate
Bandwidth (RF) with R&S
®
SMW-K525
option
at maximum sample rate,
rolloff to –0.1 dB
1000 MHz
at reduced sample rate,
rolloff to –0.1 dB
(The waveform is automatically
interpolated to the internal sample rate of
1200 MHz.)
0.833 · sample rate
Bandwidth (RF) with R&S
®
SMW-K527
option
at maximum sample rate,
rolloff to –0.1 dB
2000 MHz
at reduced sample rate,
rolloff to –0.1 dB
(The waveform is automatically
interpolated to the internal sample rate of
2400 MHz.)
0.833 · sample rate
Frequency offset
Using the frequency offset, the center frequency of the wanted baseband signal can be
shifted. The restrictions caused by the modulation bandwidth still apply.
Frequency offset setting range
–250 MHz to +250 MHz
with R&S
®
SMW-K525 option
–500 MHz to +500 MHz
with R&S
®
SMW-K527 option
–1000 MHz to +1000 MHz
Frequency offset setting resolution
0.01 Hz
Frequency offset error
< 9 · 10
–6
Hz + relative deviation of
reference frequency · frequency offset
(nom.)
Triggering
A trigger event restarts I/Q generation. The I/Q signal is then synchronous with the
trigger (with a specific timing jitter).
Trigger source
event triggered via GUI or remote
command
internal
event triggered by other baseband
generator
internal (baseband A/B)
event triggered by external trigger signal
external
Trigger modes
The signal is generated continuously.
auto
The signal is generated continuously.
A trigger event causes a restart.
retrig
The signal is started only when a trigger
event occurs. Subsequent trigger events
are ignored.
armed auto
The signal is started only when a trigger
event occurs. Every subsequent trigger
event causes a restart.
armed retrig