Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 65
Impulsive noise simulation
This function allows to add a pulsed AWGN signal to the wanted signal with settable number of pulses per frame and within settable
limits of randomly distributed pulse intervals.
Impulsive noise
AWGN signal data
see R&S
®
SMW-K62 option
C/I
Setting range
Depends on the set RF level.
The PEP of the sum signal (wanted signal
+ noise) must not exceed the maximum
possible PEP of the respective RF path.
–35 dB to +60 dB
Setting resolution
0.01 dB
Frame duration
0.1 ms to 1000.0 ms
Pulse duration
fixed
0.25 µs
Pulses per frame
1 to 40000
Minimum pulse interval
for pulses per frame > 1
Setting range
0.25 µs to 16 ms
Setting resolution
0.25 µs
Maximum pulse interval
for pulses per frame > 1
Setting range
0.25 µs to 16 ms
Setting resolution
0.25 µs
Distribution of pulse intervals
PRBS
Availability of phase noise and impulsive noise for different baseband configurations
Baseband main
module
Fading/baseband configuration
Phase noise
Impulsive noise
R&S
®
SMW-B13
standard
●
●
R&S
®
SMW-B13T
standard
●
●
advanced
up to 4 streams
–
●
more than 4 streams
–
–
R&S
®
SMW-B13XT
standard
●
●
advanced
up to 4 streams
●
●
more than 4 streams
●
●

Version 27.00, October 2024
66 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Envelope tracking (R&S
®
SMW-K540 option)
With this option, the analog I/Q outputs can be used to generate an analog signal corresponding to the envelope of the I/Q signal to
test envelope tracking modulators.
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13 or R&S
®
SMW-B13XT option. If the instrument
is equipped with the R&S
®
SMW-B13T option, envelope tracking can be used either on signal path A or B with one R&S
®
SMW-K540
option. For envelope tracking to be used on signal paths A and B simultaneously, two R&S
®
SMW-K540 and one R&S
®
SMW-B13T
must be installed.
Instruments equipped with the R&S
®
SMW-B13 or R&S
®
SMW-B13T option: For each R&S
®
SMW-K540 option to be installed, an
R&S
®
SMW-K16 option must be installed, and the instrument must be equipped with at least one standard baseband generator
(R&S
®
SMW-B10 option).
Instruments equipped with the R&S
®
SMW-B13XT option: For R&S
®
SMW-K540 option to be installed, the R&S
®
SMW-K17 option must
be installed, and the instrument must be equipped with at least one wideband baseband generator (R&S
®
SMW-B9 option).
General
Envelope voltage adaptation
auto normalized, auto power, manual
Output type
single-ended, differential
Bias voltage
see section Differential analog I/Q outputs or Wideband differential analog I/Q outputs
Offset voltage
see section Differential analog I/Q outputs or Wideband differential analog I/Q outputs
Envelope to RF delay
Setting range
–1 µs to +1 µs
Setting resolution
1 ps
Shaping
off, linear, from table, polynomial,
detroughing
Envelope voltage adaptation modes: auto normalized and auto power
Power amplifier input power P
in
Setting range
–145.00 dB to +30.00 dB
Setting resolution
0.01 dB
Power amplifier supply voltage V
CC
V
CC
= envelope voltage · DC modulator gain + V
CC, Offset
DC modulator gain
–20.00 dB to +20.00 dB
Power amplifier offset voltage V
CC,
Offset
0 V to 30 V
Envelope voltage adaptation mode: manual
Pregain
Setting range
–20.00 dB to 0.00 dB
Setting resolution
0.01 dB
Postgain
Setting range
–3.00 dB to +20.00 dB
Setting resolution
0.01 dB
Clipping level
upper and lower limit can be set
separately
0 % to 100 %
Maximum output voltage
see Output voltage in section Differential analog I/Q outputs
AM/AM, AM/PM predistortion (R&S
®
SMW-K541 option)
Instruments with wideband baseband (R&S
®
SMW-B13XT):
Each R&S
®
SMW-K541 option to be installed requires a wideband baseband generator (R&S
®
SMW-B9 option) and an RF path. If the
instrument is equipped with two baseband generators and two RF paths, predistortion can be used either on signal path A or B with
one R&S
®
SMW-K541 option. To allow AM/AM, AM/PM predistortion to be used on signal paths A and B simultaneously, two
R&S
®
SMW-K541 must be installed; furthermore, the instrument must be equipped with two R&S
®
SMW-B9 options and two RF paths,
i.e. an R&S
®
SMW-B2xx frequency option for path B must be installed.
Instruments with standard baseband (R&S
®
SMW-B13/-B13T):
Each R&S
®
SMW-K541 option to be installed requires a standard baseband generator (R&S
®
SMW-B10 option) and an RF path.
If the instrument is equipped with two baseband generators and two RF paths, predistortion can be used either on signal
path A or B with one R&S
®
SMW-K541 option. To allow AM/AM, AM/PM predistortion to be used on signal paths A and B
simultaneously, two R&S
®
SMW-K541 must be installed; furthermore, the instrument must be equipped with two R&S
®
SMW-B10
options, the R&S
®
SMW-B13T option and two RF paths, i.e. an R&S
®
SMW-B2xx frequency option for path B must be installed.
State
on/off
Maximum input power (PEP
in
max.)
Setting range
–145.00 dB to +30.00 dB
Setting resolution
0.01 dB
Shaping
polynomial, from table

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 67
Digital Doherty (R&S
®
SMW-K546 option)
The Digital Doherty option only applies to instruments equipped with two RF paths and two baseband generators. Two
R&S
®
SMW-K541 options and the R&S
®
SMW-B90 option (phase coherence) must be installed as prerequisite.
State
on/off
Maximum input power (PEP
in
max.)
Setting range
–145.00 dB to +30.00 dB
Setting resolution
0.01 dB
Shaping
polynomial, from table, classic Doherty
User-defined frequency response correction (R&S
®
SMW-K544 option)
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13 option. If the instrument is equipped with the
R&S
®
SMW-B13T or R&S
®
SMW-B13XT option, user-defined frequency response correction can be used either on signal path A or B
with one R&S
®
SMW-K544 option. For user-defined frequency response correction to be used on signal paths A and B simultaneously,
two R&S
®
SMW-K544 must be installed.
State
on/off
Scattering parameters
File format
*.s<n>p (e.g. *.s2p)
Maximum number of points
16384
Number of cascadable datasets
up to 10
Additional frequency response
File format
*.fres, *.ucor
Number of files
up to 5
Absolute level correction at center
frequency
based on S-parameter data
on/off
Minimum compensation bandwidth
with R&S
®
SMW-B13/-B13T options
8 MHz
with R&S
®
SMW-B13XT option
100 MHz
Automated RF port alignment (R&S
®
SMW-K545 option)
Instruments with wideband baseband (R&S
®
SMW-B13XT):
For each installed RF path, R&S
®
SMW-B9, R&S
®
SMW-K61 and R&S
®
SMW-K544 must be installed as prerequisite. Furthermore, the
instrument must be equipped with the R&S
®
SMW-B90 option.
Instruments with standard baseband (R&S
®
SMW-B13/-B13T):
For each installed RF path, R&S
®
SMW-B10, R&S
®
SMW-K61 and R&S
®
SMW-K544 must be installed as prerequisite. Furthermore, the
instrument must be equipped with the R&S
®
SMW-B90 option.
The option cannot be installed if an R&S
®
SMW-B1044O, R&S
®
SMW-B2044O, R&S
®
SMW-B1056O or R&S
®
SMW-B1067O option
is installed.
To run this option a setup should be defined and generated using the R&S
®
RFPAL software. At least two signal paths should be
provided. In case of a setup with multiple instruments, an instrument is designated as primary instrument and should be used to
control the option.
State
on/off
Align
aligned, not aligned
Setup file
setup file including alignment data is
generated by R&S
®
RFPAL
*.rfsa
Additional S-parameter files
File format
*.s<n>p (e.g. *.s2p)
Maximum number of points
16384
Number of cascadable datasets
recommended ≤ 2
up to 10