SMW200A_specs_en_3606-8037-22_v2700.pdf - 第68页
Version 27.00, October 2024 68 Rohde & Schwar z R&S ® SMW200A Vec tor Signal Generator Crest factor reduction ( R&S ® SMW -K54 8 option) Each R&S ® SMW-K548 op tion require s a standard baseban d generato…

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 67
Digital Doherty (R&S
®
SMW-K546 option)
The Digital Doherty option only applies to instruments equipped with two RF paths and two baseband generators. Two
R&S
®
SMW-K541 options and the R&S
®
SMW-B90 option (phase coherence) must be installed as prerequisite.
State
on/off
Maximum input power (PEP
in
max.)
Setting range
–145.00 dB to +30.00 dB
Setting resolution
0.01 dB
Shaping
polynomial, from table, classic Doherty
User-defined frequency response correction (R&S
®
SMW-K544 option)
This option can be installed once if the instrument is equipped with the R&S
®
SMW-B13 option. If the instrument is equipped with the
R&S
®
SMW-B13T or R&S
®
SMW-B13XT option, user-defined frequency response correction can be used either on signal path A or B
with one R&S
®
SMW-K544 option. For user-defined frequency response correction to be used on signal paths A and B simultaneously,
two R&S
®
SMW-K544 must be installed.
State
on/off
Scattering parameters
File format
*.s<n>p (e.g. *.s2p)
Maximum number of points
16384
Number of cascadable datasets
up to 10
Additional frequency response
File format
*.fres, *.ucor
Number of files
up to 5
Absolute level correction at center
frequency
based on S-parameter data
on/off
Minimum compensation bandwidth
with R&S
®
SMW-B13/-B13T options
8 MHz
with R&S
®
SMW-B13XT option
100 MHz
Automated RF port alignment (R&S
®
SMW-K545 option)
Instruments with wideband baseband (R&S
®
SMW-B13XT):
For each installed RF path, R&S
®
SMW-B9, R&S
®
SMW-K61 and R&S
®
SMW-K544 must be installed as prerequisite. Furthermore, the
instrument must be equipped with the R&S
®
SMW-B90 option.
Instruments with standard baseband (R&S
®
SMW-B13/-B13T):
For each installed RF path, R&S
®
SMW-B10, R&S
®
SMW-K61 and R&S
®
SMW-K544 must be installed as prerequisite. Furthermore, the
instrument must be equipped with the R&S
®
SMW-B90 option.
The option cannot be installed if an R&S
®
SMW-B1044O, R&S
®
SMW-B2044O, R&S
®
SMW-B1056O or R&S
®
SMW-B1067O option
is installed.
To run this option a setup should be defined and generated using the R&S
®
RFPAL software. At least two signal paths should be
provided. In case of a setup with multiple instruments, an instrument is designated as primary instrument and should be used to
control the option.
State
on/off
Align
aligned, not aligned
Setup file
setup file including alignment data is
generated by R&S
®
RFPAL
*.rfsa
Additional S-parameter files
File format
*.s<n>p (e.g. *.s2p)
Maximum number of points
16384
Number of cascadable datasets
recommended ≤ 2
up to 10

Version 27.00, October 2024
68 Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator
Crest factor reduction (R&S
®
SMW-K548 option)
Each R&S
®
SMW-K548 option requires a standard baseband generator (R&S
®
SMW-B10 option) or a wideband baseband generator
(R&S
®
SMW-B9 option). If two baseband generators are installed, crest factor reduction can be applied either on path A or B with one
R&S
®
SMW-K548 option. For crest factor reduction to be applied on paths A and B simultaneously, two R&S
®
SMW-K548 must be
installed.
Crest factor reduction can be applied to any waveform loaded in the arbitrary waveform generator.
State
on/off
Algorithm
clipping and filtering
Desired crest factor delta
–20 dB to 0 dB
Maximum iterations
1 to 10
Filter mode “simple”
Signal bandwidth
0 Hz to input file sample rate
Channel spacing
0 Hz to input file sample rate
Filter mode “enhanced”
Passband frequency
0 Hz to ½ of input file sample rate
Stopband frequency
0 Hz to ½ of input file sample rate
Maximum filter order
21 to 300
Slow I/Q (R&S
®
SMW-K551 option)
In slow I/Q mode, the generated signal’s clock rate can be reduced (e.g. a 20 MHz LTE signal is generated with a clock rate of
240 kHz instead of the original 30.72 MHz). This feature can be used to run tests on hardware emulation platforms not yet capable of
full-speed signal processing. The signal and fading characteristics are comparable to those of a system running at full speed. The
actual clock rate of the generated signal is controlled by the device connected to the digital I/Q output connectors of the
R&S
®
SMW200A.
R&S
®
SMW-K551 on instruments with wideband baseband (R&S
®
SMW-B9, R&S
®
SMW-B13XT)
At least one R&S
®
SMW-B9 wideband baseband generator option and one R&S
®
SMW-K19 digital baseband output for wideband
baseband option must be installed.
Note:
Only available for system configuration mode: advanced and signal outputs: digital only (HS).
All digital I/Q outputs need to run at the same clock rate.
The minimum clock rate is limited by the external controlling device only.
The R&S
®
SMW200A can handle varying clock rates.
With activated slow I/Q mode, marker signals are only available via the digital I/Q interface, and not via USER or T/M/C connectors.
With activated slow I/Q mode, no digital baseband inputs are available.
R&S
®
SMW-K551 on instruments with standard baseband (R&S
®
SMW-B10, R&S
®
SMW-B13/-B13T)
At least one R&S
®
SMW-B10 standard baseband generator option and one R&S
®
SMW-K18 digital baseband output option must be
installed.
Note:
All digital I/Q outputs need to run at the same clock rate.
The minimum clock rate is limited by the external controlling device only.
The R&S
®
SMW200A can handle varying clock rates.
In digital only/digital only multiplexed mode, marker signals are only available via the digital I/Q interface, and not via USER or T/M/C
connectors.
In digital only/digital only multiplexed mode with activated slow I/Q, no digital baseband inputs are available.

Version 27.00, October 2024
Rohde & Schwarz R&S
®
SMW200A Vector Signal Generator 69
Bandwidth extension (R&S
®
SMW-K555 option)
The R&S
®
SMW-K555 option requires two R&S
®
SMW-B9 wideband baseband generator options and two R&S
®
SMW-K527 baseband
extension to 2 GHz RF bandwidth options. Single and dual unit operation is supported.
Bandwidth extension enhances the usable clock rate of the arbitrary waveform generator up to 4.8 GHz and can be used with any
waveform loaded int the arbitrary waveform generator. To run this option, an external power combiner and a measurement device is
needed. The measurement device can be either an analyzer or a power meter.
Supported standards and modulation
systems
with R&S
®
SMW-B9 option – arbitrary
waveform mode
ARB
with R&S
®
SMW-K414 option
OFDM
with R&S
®
SMW-K261 option
multicarrier CW
with R&S
®
SMW-K477 option
IEEE 802.11ay
State
on, off
Setup file
setup file including alignment data is
generated by bandwidth extension option
*.bwsa
Clock rate
200 MHz to 4.8 GHz
Bandwidth
up to 4 GHz
Waveform sample length
513 sample to 256 Msample
with R&S
®
SMW-K515 option
513 sample to 2 Gsample
Linearize RF (R&S
®
SMW-K575 option)
The R&S
®
SMW-K575 option requires at least one R&S
®
SMW-B10 standard baseband generator or R&S
®
SMW-B9 wideband
baseband generator.
Linearize RF improves the EVM and ACLR performance for high output powers over the whole RF frequency range of the device.
State
auto/off
Notched signals (R&S
®
SMW-K811 option)
At least one R&S
®
SMW-B10 standard baseband generator option or R&S
®
SMW-B9 wideband baseband generator option must be
installed. If two baseband generators are installed, notched signals can be generated either on path A or B with one R&S
®
SMW-K811
option. For notched signals to be generated on paths A and B simultaneously, two R&S
®
SMW-K811 must be installed.
Up to 25 band-stop filters can be applied to the baseband signal.
Center frequency and bandwidth can be set independently for each band-stop filter.
Supported standards and modulation
systems
with R&S
®
SMW-B9 or R&S
®
SMW-B10
option – arbitrary waveform mode
ARB
with R&S
®
SMW-K55 option
LTE
with R&S
®
SMW-K115 option
cellular IoT
with R&S
®
SMW-K114 option
custom OFDM
with R&S
®
SMW-K130 or
R&S
®
SMW-K355 option
OneWeb
with R&S
®
SMW-K52 option
DVB-H/DVB-T
with R&S
®
SMW-K116 option
DVB-S2/DVB-S2X
Number of notches
1 to 25
Notch width
0 Hz to 0.1 · clock frequency
Notch center frequency
–0.5 · clock frequency to +0.5 · clock
frequency
Customized digital input (R&S
®
SMW-K556 option)
With the R&S
®
SMW-K556 option, I/Q data from an existing hardware can be fed into the BBIN HS DIG I/Q inputs. This option can be
installed once or twice. Each R&S
®
SMW-K556 option to be installed requires a R&S
®
SMW-B9 wideband baseband generator option.
The existing hardware requires a Xilinx Virtex FPGA and the corresponding Rohde & Schwarz IP core.
Interface
Technical specifications
HS DIG I/Q interface input parameters
see section Digital baseband
inputs/outputs for wideband baseband
Interface parameters
Connector
HS DIG I/Q 1 and HS DIG I/Q 2 on rear
panel
QSFP+
Protocol
R&S
®
Digital I/Q Interface HS
Data rate
50 Gbit/s
sample rate up to 1.25 GHz