Tracker-2800-2800S.pdf - 第25页

25 Figure 2-15 . Selectin g range parameter used while scanning Press the MENU butto n until the power-up menu is reach (the signature will become live at this point). Pressing the SCAN butto n will now scan through the …

100%1 / 86
24
pins to be scanned. Press the PINS button to return to the PACKAGE menu. Press the SINGLE button
to toggle (bottom right image in figure 2-14) between SINGLE (single pass then stop), LOOP
(continuous scan) or STOP (stop on failure).
Figure 2-14. Scan/Package menus
Press the MENU button until the power-up menu is reached (the signature will become live). The pins
can be manually scanned by pressing the PIN button on the LCD. Make sure you have a Common
connection to the PCBs being tested otherwise you will see only open circuit signatures (horizontal
lines) on the Tracker display. The display will change to show the current pin number at the top. Press
the and arrow buttons to step manually through the pins on the connected device with the
signature for each pin will be displayed on the LCD. Press A to view only channel A, B to view only
channel B or press the ALT button to view signatures on channel A and channel B. Comparison of the
A (green signature) and B (red signature) channels is automatic when the Tracker is in ALT mode.
Note the green “check” or red “X” mark on the bottom right portion of the LCD will indicate a passed
or failed comparison.
To scan without having to increment the and arrow buttons on the PIN menu, press the front panel
SCAN button and the 2800S will step through the pins automatically. Press the SCAN button again to
stop the scanning. By default, selecting the SCAN button will test each pin in each 2800S resistance
range before stepping to the next pin. To change which range parameter is scanned, press the front
panel MENU button to display the Main menu (left image in figure 2-15) on the LCD. Press the
SCAN button on the LCD to display the SCAN menu (middle image in figure 2-15) and press the
MODE button on the LCD to display the MODE menu (right image in figure 2-15). Use the and
arrow buttons to change the scan mode between resistance (RES), frequency (FREQ), none (NONE)
and voltage (VOLT).
Press SCAN Press PACKAGE Configure Package
25
Figure 2-15. Selecting range parameter used while scanning
Press the MENU button until the power-up menu is reach (the signature will become live at this point).
Pressing the SCAN button will now scan through the component pins using the range parameter
selected in the Scan MODE menu.
2-13 Auxiliary (AUX) connector description
The rear panel Auxiliary (AUX) 9 pin Din connector pins are shown figure 2-16. This connection is
used for control of the Tracker with the optional footswitch accessory.
Press the SCAN button Press the MODE button Use and to select the Scan mode
Figure 2-16
Auxiliary connector pin out
diagram
26
2-14 Using the Optional Footswitch with the Tracker 2800/2800S
An optional footswitch (Huntron part number 98-0314) is available for controlling certain functions of
the Tracker 2800. It connects to the auxiliary port located on the Tracker back panel.
Figure 2-17
Footswitch accessory for Tracker 2800/2800S
The footswitch can be used for hands-free control of start/stop of scanning, changing the channel and
changing to a pre-selected range.
To configure the footswitch control press the front panel MENU button to display the Main menu on
the LCD (left image in figure 2-18). Press the CONTROL button on the LCD to display the Control
menu (image second from left in figure 2-18). Press the SWITCH button on the LCD to display the
Switch menu (image third from left in figure 2-18). Use the and arrow buttons to select the
footswitch control between SCAN (image third from left in figure 2-18), CHANNEL (middle image
on right in figure 2-18) and RANGE (bottom image on right in figure 2-18).
Figure 2-18 Tracker 2800 Footswitch Control Menus
Press the MODE button to return to the Control menu or press the front panel MENU continuously
until the power-up menu is displayed.
Press the CONTROL button Press the SWITCH button Use and to select the Switch mode