satisfy the reference value is searched automatically and it is displayed in the result file. Perform mapping with the <Refine Mapping Result> check box being selected.  S tate display Displays current progressing…

100%1 / 245
<2. Camera> group
Selects the camera for performing lighting brightness test or mapping.
Activated cameras can be designated at the same time.
<3. Tool Setting> group
Performs test or preparation for mapping.
à <Prepare Test / Mapping> Button
Return all nozzles mounted on the head to each corresponding ANC
hole. Before doing so, arrange the LightFly nozzle, a nozzle for
calibration into the No. 1 hole of the ANC.
If the nozzle arrangement is finished, the <Start> button and <Test>
button in the <Test> group and <Mapping>group.
<4. Search Area> Group
Clicking the button corresponding to the light for which mapping is to be
performed will display the following dialog box.
Click on the button corresponding to the illumination for mapping, set the
test area (Area 1 ~ 4) to 16 illumination levels for each lighting, and setup
the average illumination level for each pixel of the test area to the
maximum value.
<Test> group
Designates the file to save the test results and starts testing.
<Mapping> group
à
button
Select the Map File to be used for mapping. Since the file name is set
automatically, do not change it.
à <Start> button
Start the Mapping
à <Test> button
Perform test for the mapping result. The light level that does not
satisfy the reference value is searched automatically and it is displayed
in the result file.
Perform mapping with the <Refine Mapping Result> check box being
selected.
State display
Displays current progressing state.
Camera (Number of camera for which test or maping is completed) /
(Total number of selected camera)
(I: Inner Level, O: Outer Level, S: Side Level, Back: Back Level)
<Close> button
Closes the dialog box..
<Import> button
First click this button and select the map file on which the result is to be
reflected.
Change the light level after importing the data from the file in which the
mapping result is saved.
Click the <Import> button with the <Import All> check box being selected.
<Update> button
Transmits the set data to the equipment and closes the dialog box.
<Cancel> button
Ignores the set data and closes the dialog box.
1.2.11. Board Position Calibration
Teach the stopper position in the conveyor installation area and the X position of
the sensor. The following is the board position calibration procedure.
1. Select the edit box for the < Sensor X> using the mouse and move the
downward camera (Fiducial Camera) to the position of Sensor X by clicking
the <Move> button.
2. Perform teaching of the exact position using the teaching box. Enter the precise
coordinates for ‘Sensor X’ by pressing the <Get> button of the teaching box.
3. Select the <Stopper X> combo box using the mouse and then select the stopper
to teach and click the <Move> button, move fiducial Camera to position of
Stopper X.
4. Click the <Stopper Up/Down> button in the left of the <Stopper X> edit box
first to move the stopper up and then teach the stopper position correctly using
the teaching box.
5. Click the <Get> button to enter the coordinates in the <Stopper X> edit box.
6. move the stopper down by clicking the <Stopper Up/Down> button.
1
2
3