Handbuch für das Achsprüfgerät / Axis T ester Manual SIPLACE Axis T e ster 8 Using the menus Ausgabe 08/2003 Edition 74 Fig. 8 - 34 Stopping the measurement t o identify zero pulse errors The resul t appears in th e next…

SIPLACE Axis Tester Handbuch für das Achsprüfgerät / Axis Tester Manual
Ausgabe 08/2003 Edition 8 Using the menus
73
Fig. 8 - 31 Setting markers T1 and T2 for the plot area
Æ Use and to move marker T1. Use and to move the zero line
up or down.
Æ Press to confirm your settings.
Æ Set marker T2 as described for marker 1.
Æ Select "Quit". The numerical value of the time window is output to the plotter.
8.4.4 “Track signal” menu
This menu is designed to help you analyze and troubleshoot track signals.
Æ Select the “Track signals” option from the “Properties” menu (see Fig. 8 - 23 on page 69).
The following window appears.
Fig. 8 - 32 “Track signals” menu, “Reference error”
8.4.4.1 Checking track signals for zero pulse errors
Æ Select the “Reference error” menu option.
The window for starting the measurement appears.
Fig. 8 - 33 Starting the measurement to identify zero pulse errors
Æ Turn the axis either manually or using the SITEST program. Then return it to the starting point
and stop the measurement.
Menu for testing the zero pulses
Press to start the measuremen
t.

Handbuch für das Achsprüfgerät / Axis Tester Manual SIPLACE Axis Tester
8 Using the menus Ausgabe 08/2003 Edition
74
Fig. 8 - 34 Stopping the measurement to identify zero pulse errors
The result appears in the next window.
Fig. 8 - 35 Result of the zero pulse measurement
Ideally, the counter value should read 0 since the forward and backward movements were
identical. In this case, a 0 will also be output for the “Error per zero impulse” value.
If the counter value is other than 0, then counting errors have occurred. A minus sign before
the counter value means that more zero pulses were measured during the backward move-
ment than during the forward movement. In this case, a value other than 0 will again be output
for “Error per zero impulse”.
8.4.4.2 Checking track signals for errors
Fig. 8 - 36 “Track signals” menu, “Line error”
Æ Select the “Line error” option from the “Track signals” menu.
The window for selecting the display mode appears.
Press to stop the measuremen
t.
Counter value
Number of measured zero pulses
Counting error per zero pulse
Menu for testing the track signals

SIPLACE Axis Tester Handbuch für das Achsprüfgerät / Axis Tester Manual
Ausgabe 08/2003 Edition 8 Using the menus
75
Fig. 8 - 37 “Select mode” menu
Æ Select “Show all”. The window for traversing the axis appears, followed by the window for stop-
ping the measurement.
Fig. 8 - 38 Prompts to traverse the axis and stop the measurement
Æ Press to stop the measurement.
The result window appears.
Fig. 8 - 39 Window for selecting the zero pulse
Æ Use the , , and cursor keys to select the required zero pulse
(the maximum number of zero pulses is 37).
Æ Press
All zero pulses are displayed.
Only zero pulses > 5 msec apart are
displayed.
This filter function should be used for
standard applications.
P
rompt to traverse the axis Prompt to stop the measurement
Duration of the end signal
PLEASE NOTE: The first measurement
is always invalid since there is no refer-
ence point for calculating the difference.8
Signals a counting error
Displays the recorded zero pulse
Signals a timing error
Black background:
The test results for this position are displayed in the next window.