MDO3000 Programmer Manual.pdf - 第1004页
Commands Listed in A lphabetical Order (See page 2-95, Wa veform T ransfer Command Gr oup .) (See page D-1, W aveform T ransfer (WFMOutpr e and CURV e Query) Examples .) Group W aveform T ransfer Syntax WFMOutpre:YZE ro?…

Commands Listed in Alphabetical Order
Group
Waveform Trans
fer
Syntax
WFMOutpre:YOFf?
Related Commands
DATa:SOUrce,
WFMOutpre:BYT_Nr
Examples
WFMOUTPRE:YOFF? might return :WFMOUTPRE:YOFF - 50.0000E+0 indicating
that the position indicator for the waveform was 50 digitizing levels (2 divisions)
below cen
ter screen (for 8-bit waveform data).
WFMOutpre:YUNit? (Query Only)
This query returns the units of data points in the outgoing waveform record
transferred using the
CURVe? query. This can be any of several string values,
depending upon the vertical units of the source waveform (specified by the
DATa:SOUrce command). Typically, this is “V” for volts. If the waveform
speci
fied by DATa:SOUrce is not turned on, nothing is returned and an error
event is generated.
(See p
age 2-95, Waveform Transfer Command Group.)
(See page D-1, Wa veform Transfer (WFMOutpre and CURVe Query) Examples.)
Group
Waveform Transfer
Syntax
WFMOutpre:YUNit?
Related Commands
DATa:SOUrce
Examples
WFMOUTPRE:YUNIT? might return :WFMOUTPRE:YUNIT "dB" indicating that
the vertical units for the waveform are measured in decibels.
WFMOutpre:YZEro? (Query Only)
This query returns the vertical offset of the source waveform. You can
determine the units using the WFMOutpre:YUNit? query. See the description of
WFMInpre:YZEro to see how this offset is used to convert waveform sample
values to volts. If the waveform specified by DATa:SOUrce is not turned on,
nothing is returned and an error event is generated.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-973

Commands Listed in Alphabetical Order
(See page 2-95, Wa
veform Transfer Command Group.)
(See page D-1, Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
Group
Waveform Transfer
Syntax
WFMOutpre:YZEro?
Related Commands
DATa:SOUrce,
WFMOutpre:YUNit?
Examples
WFMOUTPRE:YZERO? might return :WFMOUTPRE:YZERO -100.0000E-3
indicating th
at vertical offset is set to -100 mV.
ZOOm? (Query Only)
Returns the current vertical and horizontal positioning and scaling of thedisplay.
Group
Zoom
Syntax
ZOOm?
Examples
ZOOM? might return :ZOOM:MODE 1;GRATICULE :SIZE
80;SPLIT EIGHTYTWENTY; :ZOOM:ZOOM1:STATE
1;SCALE 400. 0000E-12;POSITION 46.8986;FACTOR
50.0000E+3;HORIZONTAL:POSITION 46.8986;SCALE 40 0.0000E-12
ZOOm{:MODe|:STATE}
Turns Zoom mode on or off. The Zoom query returns the current state of Zoom
mode. This command is equivalent to pressing the zoom button located on the
front panel.
Group
Zoom
Syntax
ZOOm{:MODe|:STA
TE} {ON |OFF|<NR1>}
ZOOm{:MODe|:STATE}
2-974 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Arguments
ON turns on Zoom
mode.
OFF turns off Zoom mode.
<NR1> is an integer. 0 turns off Zoom mode; any other value turns on Zoom mode.
Examples
ZOOM:MODE OFF turns off Zoom mode.
ZOOM:MODE? might return :ZOOM:MODE 1 indicating that Zoom mode is
currently turned on.
ZOOm:ZOOM<x>? (Query Only)
Returns the current vertical and horizontal positioning and scaling of thedisplay.
<x> can only be 1.
Group
Zoom
Syntax
ZOOm
:ZOOM<x>?
Examples
ZOOM:ZOOM1? might return :ZOOM:ZOOM1:STATE 1;SCALE
400.0000E-12;POSITION 46.8986;FACTOR 50.0000E+3;HORIZO
NTAL:POSITION 46.8986;S CALE 400.0000E-12
.
ZOOm:ZOOM<x>:FACtor? (Query Only)
Returns the zoom factor of a particular zoom box. <x> can only be 1.
Group
Zoom
Syntax
ZOOm:ZOOM<x>:FACtor?
Returns
<NR1> is an integer that specifies the zoom factor of a zoom box.
ZOOm:ZOOM<x>:POSition
Sets the horizontal position of the zoom box, in terms of 0 to 100.0% of upper
window. <x> can only be 1.